Analytic study on arsenic content in plastic materials  

Lee, Jae-In (Test & Standardization Center, Korea Institute of Ceramic Eng. & Tech.)
Choi, Zel-Ho (Test & Standardization Center, Korea Institute of Ceramic Eng. & Tech.)
Choi, Ki-In (Test & Standardization Center, Korea Institute of Ceramic Eng. & Tech.)
Ko, Byung-Rae (Korea Conformity Laboratories)
Choi, In-Suk (Korea Conformity Laboratories)
Cho, Hun-Sik (Korea Conformity Laboratories)
Publication Information
Analytical Science and Technology / v.25, no.6, 2012 , pp. 402-409 More about this Journal
Abstract
Quantitative analytical condition for arsenic in plastic materials by the flow injection-hydride generation-inductively coupled plasma-atomic emission spectrometry was studied. The plastic samples were dissolved by wet-acid digestion method and microwave-acid digestion method. The reproducibility and accuracy in this method was verified using of certified reference materials(CRMs) CRM-EC680k and CRM-EC681k. The analyical results agrees with certified value within the range of uncertainty. The results of CRM-EC680k wer 4.1~4.3 mg/kg (certified value $4.1{\pm}0.5$ mg/kg), and the results of CRM-EC681k were 28.9~30.6 mg/kg (certified value $29.1{\pm}1.8$ mg/kg).
Keywords
arsenic; HG-ICP-AES; HG-AAS; wet-acid digestion method; microwave digestion method;
Citations & Related Records
연도 인용수 순위
  • Reference
1 J. A. C. Broekaert, T. Graule, H. Janett, G. Tolg and P. Tschopel, Fresenius Z. Anal. Chem., 322, 825 (1989).
2 T. Ishizuka, Y. Uwamino and A. Tange, Anal. Chim. Acta, 161, 285 (1984).   DOI   ScienceOn
3 R. A. Sutherland and F. M. G. Taek, Advances in Environmental Research, 8, 37 (2003).   DOI   ScienceOn
4 J. Wang, T. Nakazako, K. Sakanishi, O. Yamada, H. Tao and I. Saito, Talanta, 68, 1584 (2006).   DOI   ScienceOn
5 V. Sandroni and C. M. M. Smith, Anal. Chim. Acta, 468, 335 (2002).   DOI   ScienceOn
6 M.Thomson, B. Pahlavapour and S. J. Walton, Analyst, 103, 568 (1978).   DOI
7 N. J. Agtertdenbos and S. J. Santosa, Anal. At. spectrom., 5, 681 (1990).   DOI
8 C. C. Y. Chan and R. S. Sandana, Analytica Chim. Acta, 270, 231 (1992).   DOI   ScienceOn
9 L. Ebdon, L. Pitts and R. Cornelis 'Trace Element Speciation for Environment Food and Health, ed. Royal Society of Chemistry, Cambridge, UK, 2002.
10 J. O. Nriagu, 'Arsenic in the Environment Part II: Human Health and Ecosystem Effects, ed. John Wiley & Sons Inc., New York, 1994.
11 KS M 3210 플라스틱-카드뮴 정량 방법-습식분해법.
12 KS M 3211 플라스틱중의 납 정량 방법.
13 IEC 62321 Procedures for theDetermination of Levels of Six Regulated Substances (Lead, Mercury, Cadmium, Hexavalent Chromium, Polybrominated Biphenyls, Polybrominated Diphenyl Ethers) in Electrotechnical Products (2008).