Browse > Article

Some considerations for the analytical approaches to measure atmospheric ammonia: tests on a relative performance of a portable colorimeter against UV/VIS spectrophotometer  

Kim, K.H. (Dept. of Earth & Environmental Sciences, Sejong University)
Park, S.Y. (Dept. of Earth & Environmental Sciences, Sejong University)
Kim, S.T. (Dept. of Environmental and Life Engineering, Daejeon University)
Publication Information
Analytical Science and Technology / v.18, no.5, 2005 , pp. 425-430 More about this Journal
Abstract
In this study, we investigated the performance of a portable colorimeter used for the measurements of atmospheric ammonia against the more reliable method like UV/VIS spectrophotometer. For the purpose of this study, we tested a portable colorimeter and UV/VIS spectrophotometer for such basic analytical parameters as detectibility, reproducibility, linearity, and accuracy. According to our study, the performance of the colorimeter was fairly good to show a good agreement with the UV/VIS system. However, the performance of colorimeter suffered from both low and high concentration ranges: it was found to have very poor reproducibility at lower concentration (below $10\mu}g$), while its linear dynamic range was limited at the upper end (e.g., near $100{\mu}g$). If we assume that air samples are collected up to 30 liter, it implies that the colorimeter can be used to measure samples containing NH3 in the range of 500 ppb to 5 ppm (without the aid of diluting technique). Consequently, we recommend the use of such instrument with the awareness of its basic properties pertaining to the fundamental performance (such as its detectable range).
Keywords
atmospheric; ammonia; colorimeter; UV/VIS spectrophotometer; performance test;
Citations & Related Records
Times Cited By KSCI : 3  (Citation Analysis)
연도 인용수 순위
1 K.-H. Kim, Y.-J. Choi, E.-C. Jeon and Y. Sunwoo, 'Characterization of malodorous sulfur compounds in landfill gas' Atmospheric Environment, 39(6), 1103- 1112(2005)   DOI   ScienceOn
2 이동수, 정주영, 권은정, 박영순, 최건형, '시화 반월공단 악취물질에 대한 실시간 측정분석 (보고서)'., 국립환경 연구원(2001).
3 박상진, '공기희석관능법을 이용한 국내하수처리장처리공정별 악취농도에 대한 조사연구' 한국냄새환경학회지, 2(1), 32-37(2003)
4 양성봉 등, '안산 반월시역 악취배출원 조사 및 저감 방안 연구 (보고서)' 국립환경연구원(2003)
5 김강석, '왕겨를 활용한 수중 암모니아성 질소의 제거에 관한 연구' 공주대 교육대학원 석사논문 (2003)
6 정형근, 김범식, '인도페놀법을 이용한 수용액 중 암모니아 정량에 관한 연구: 평형법, 반응속도법, 흐름 주입분석법의 비교와 평가' 한국환경과학회지, 4(1) 92(1994)
7 홍윤정, 김기현, '대기 중 카보닐 계열 성분의 분석 기법의 연구: 포름알데하이드와 DNPH의 반응 특성을 중심으로' 한국분석과학회지, 18(1), 42-49(2005)
8 K. H. Kim, 'Performance characterization of the GC/ PFPD for H2S, CH3SH, DMS, and DMDS in air' Atmospheric Environment, 39(12), 2235-2242(2005)   DOI   ScienceOn
9 김기현, 주도원, 최여진, 홍윤정, 사재환, 박종호, 전의찬, 최청렬, '구윤서 안산시 주거지역을 중심으로 한 환경대기 중 휘발성 유기 화합물과 황계열 성분의 온라인 연속 측정 연구' 한국대기환경학회, 21(2), 215-226(2005a).
10 정구회, 김선태, 박민수, 정재호, '하수처리장에서의 악취 배출 성상 조사' 한국냄새환경학회지, 3(4), 225-233(2004)
11 김기현, 최여진, 홍윤정, 사재환, 박종호, 전의찬, 최청렬, 구윤서, '반월공단내 주요 산업 시설물들의 대기배출시설을 중심으로 한 주요 악취성분의 배출특성 및 배출원별 악취인자 선별 방식에 대한 예비 연구' 한국대기환경학회, 21(2), 215-226(2005b)