SiPM PET/CT에서 3D 프린팅 기반 자체제작한 팬텀을 이용한 iMAR 알고리즘 유용성 평가에 관한 연구

The feasibility of algorithm for iterative metal artifact reduction (iMAR) using customized 3D printing phantom based on the SiPM PET/CT scanner

  • 이민규 (의정부을지대학교병원 핵의학과) ;
  • 박찬록 (을지대학교 방사선학과)
  • Min-Gyu Lee (Dept. of nuclear medicine, Uijeongbu Eulji Medical Center) ;
  • Chanrok Park (Dept. of Radiological Science, Eulji University)
  • 투고 : 2023.11.21
  • 심사 : 2023.12.14
  • 발행 : 2024.05.30

초록

Purpose: To improve the image quality in positron emission tomography (PET), the attenuation correction technique based on the computed tomography (CT) data is important process. However, the artifact is caused by metal material during PET/CT scan, and the image quality is degraded. Therefore, the purpose of this study was to evaluate image quality according to with and without iterative metal artifact reduction (iMAR) algorithm using customized 3D printing phantom. Materials and Methods: The Hoffman and Derenzo phantoms were designed. To protect the gamma ray transmission and express the metal portion, lead substance was located to the surface. The SiPM based PET/CT was used for acquisition of PET images according to application with and without iMAR algorithm. The quantitative methods were used by signal to noise ratio (SNR), coefficient of variation (COV), and contrast to noise ratio (CNR). Results and Discussion: The results shows that the image quality applying iMAR algorithm was higher 1.15, 1.19, and 1.11 times than image quality without iMAR algorithm for SNR, COV, and CNR. Conclusion: In conclusion, the iMAR algorithm was useful for improvement of image quality by reducing the metal artifact lesion.

키워드

참고문헌

  1. Endo K, Oriuchi N, Higuchi T, Lida Y, Hanaoka H, Miyakubo M, Ishikita T, Koyama K, PET and PET/CT using 18F-FDG in the diagnosis and management of cancer patients. Int J Clin Onclo. 2006;11:286-96.
  2. Zaharchuk G, Next generation research applications for hybrid PET/MR and PET/CT imaging using deep learning. Eur J Nucl Med Mol Imaging. 2019;46:2700-7.
  3. Sureshbabu W, Mawlawi O, PET/CT imaging artifacts. J Nucl Med Technol. 2005;33:156-161.
  4. van der Vos CS, Arens AIJ, Hamill JJ, Hofmann C, Panin VY, Meeuwis APW, Bisser EP, de Geus-Oei LF, Metal artifact reduction of CT scans to improve PET/CT. J Nucl Med 2017;58:1867-72.
  5. Bebbington N, Holdgaard P, iMAR dramatically reduces CT artefacts for hip prostheses and increases accuracy of SUVs when used for PET attenuation correction with low-dose CT. J Nucl Med. 2017;58(supplement 1):1345.
  6. Schabel C, Gatidis S, Bongers M, Huttig F, Bier G, Kupferschlaeger J, Bamberg F, La Fougere C, Nikolaou K, Pfannenberg C, Improving CT-Based PET Attenuation Correction in the Vicinity of Metal Implants by an Iterative Metal Artifact Reduction Algorithm of CT Data and Its Comparison to Dual-Energy-Based Strategies: A Phantom Study. Invest Radiol. 2017;52:61-5.
  7. Kang SH, Lee S, Lee Y, Experimental study of optimization of noise level in brain SPECT images using non-local means approach with various reconstruction methods. Nucl Eng Technol. 2023;55:1527-32.
  8. Jo JH, Kim JS, Choi J, Yoon MS, Park C, Design and analysis of a dedicated quality control phantom in nuclear medicine imaging based on the 3D printing technique using various printing materials. Optik. 2022;270:169909.