DOI QR코드

DOI QR Code

방사선치료에서 환자맞춤형 선량품질보증 장치의 비교

A Comparison of Patient-specific Delivery Quality Assurance (DQA) Devices in Radiation Therapy

  • 장경환 (극동대학교 방사선학과)
  • 투고 : 2023.02.24
  • 심사 : 2023.04.25
  • 발행 : 2023.06.30

초록

This study aimed to compare the results of delivery quality assurance (DQA) using MapCHECK and OCTAVIUS for radiation therapy. Thirty patients who passed the DQA results were retrospectively included in this study. The point dose difference (DD) and gamma passing rate (GPR) were analyzed to evaluate the agreement between the measured and planned data for all cases, Plan complexity was evaluated to analyze dosimetric accuracy by quantifying the degree of modulation according to each plan. We analyzed the monitor units (MUs) and total MUs for each plan to evaluate the correlation between the MUs and plan complexity. We used a paired t-test to compare the DD and GPRs that were obtained using the two devices. The DDs and GPRs were within the tolerance range for all cases. The average GPRs difference between the two devices was statistically significant for the brain, and head and neck for gamma criteria of 3%/3 mm and 2%/2 mm. There was no significant correlation between the modulation index and total MUs for any of the cases. These DQA devices can be used interchangeably for routine patient-specific QA in radiation therapy.

키워드

참고문헌

  1. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM task group No. 218. Med Phys. 2018;45(4):e53-83. https://doi.org/10.1002/mp.12810
  2. Intensity Modulated Radiation Therapy Collaborative Working Group. Intensity-modulated radiotherapy: Current status and issues of interest. Int J Radiat Oncol Biol Phys. 2001;51(4):880914.
  3. Cho B. Intensity-modulated radiation therapy: A review with a physics perspective. Radiat Oncol J. 2018;36(1):1-10. https://doi.org/10.3857/roj.2018.00122
  4. Chang KH. Treatment planning guideline of EBT film-Based delivery quality assurance using statistical process control in helical tomotherapy. Journal of Radiological Science and Technology. 2022;45(5):439-48. https://doi.org/10.17946/JRST.2022.45.5.439
  5. Chang KH, Lee YH, Park BH Han MC, Kim J, et al. Statistical analysis of treatment planning parameters for prediction of delivery quality assurance failure for helical tomotherapy. Technol Cancer Res Treat. 2020;19:1-9. https://doi.org/10.1177/1533033820979692
  6. Chang KH. Feasibility on Statistical process control analysis of delivery quality assurance in helical tomotherapy. Journal of Radiological Science and Technology. 2022;45(6):491-502. https://doi.org/10.17946/JRST.2022.45.6.491
  7. Jursinic PA, Sharma R, Reuter J. MapCHECK used for rotational IMRT measurements: Step-and-shoot, Tomotherapy. RapidArc. Med Phys. 2010;37:2837-46. https://doi.org/10.1118/1.3431994
  8. Keeling VP, Ahmad S, Jin H. A comprehensive comparison study of three different planar IMRT QA techniques using MapCHECK2. J Appl Clin Med Phys. 2013;14(6):4398.
  9. Li JG, Yan G, Liu C. Comparison of two commercial detector arrays for IMRT quality assurance. J Appl Clin Med Phys. 2009 Apr 29;10(2):62-74. https://doi.org/10.1120/jacmp.v10i2.2942
  10. Lee YC, Kim Y. A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments. J Radiosurg SBRT. 2021;7(4):295-307.
  11. Spezi E, Angelini AL, Romani F, Ferri A. Characterization of a 2D ion chamber array for the verification of radiotherapy treatments. Phys Med Biol. 2005;50:3361-73. https://doi.org/10.1088/0031-9155/50/14/012
  12. Thiyagarajan R, Nambiraj A, Sinha SN, Yadav G, Kumar A, Subramani V, et al. Analyzing the performance of ArcCHECK diode array detector for VMAT plan. Rep Pract Oncol Radiother. 2016;21(1): 50-6. https://doi.org/10.1016/j.rpor.2015.10.004
  13. Chang KH, Kim DW, Choi JH, Shin HB, Hong CS, Jung DM, et al. Dosimetric comparison of four commercial patient-specific quality assurance devices for helical tomotherapy. J Korean Phys Soc. 2020;76(3):257-63. https://doi.org/10.3938/jkps.76.257
  14. Myers P, Stathakis S, GutiGrrez AN, Esquivel C, Mavroidis P, Papanikolaou N. Evaluation of PTW Seven29 for tomotherapy patient-specific quality assurance and comparison with ScandiDos Delta(4). J Med Phys. 2012;37(2):72-80. https://doi.org/10.4103/0971-6203.94741
  15. Chung E, Kwon D, Park T, Kang H, Chung Y. Clinical implementation of Dosimetry CheckTM for TomoTherapy® delivery quality assurance. J Appl Clin Med Phys. 2018 Nov;19(6):193-9. https://doi.org/10.1002/acm2.12480
  16. Ghafarian M, Price M, Morales-Paliza M. Comparison of pretreatment VMAT quality assurance with the integral quality monitor (IQM) and electronic portal imaging device (EPID). J Appl Clin Med Phys. 2021;22(3):166-75. https://doi.org/10.1002/acm2.13201
  17. Chong NS, Jason JS, Kung WH, Chen CA, et al. Patient delivery quality assurance for linac-based IMRT and helical tomotherapy using solid state detectors. Radiation Measurements. 2011;46(12):1993-5. https://doi.org/10.1016/j.radmeas.2011.08.004
  18. Son J, Baek T, Lee B, Shin D, Park SY, Park J, et al. A comparison of the quality assurance of four dosimetric tools for intensity modulated radiation therapy. Radiol Oncol. 2015;49(3):307-13. https://doi.org/10.1515/raon-2015-0021
  19. MapCHECKTM Reference Guide. Document 1175011, Rev T, 23 September 2011. Sun Nuclear Corporation.
  20. Low DA, Moran JM, Dempsey JF, Dong L, Oldham M. Dosimetry tools and techniques for IMRT. Med Phys. 2011;38:1313-38. https://doi.org/10.1118/1.3514120
  21. Webb S. Use of a quantitative index of beam modulation to characterize dose conformality: Illustration by a comparison of full beamlet IMRT, few-segment IMRT (fsIMRT) and conformal unmodulated radiotherapy. Phys Med Biol. 2003;48:2051-62. https://doi.org/10.1088/0031-9155/48/14/301
  22. Chang KH, Chang JS, Park K, Chung SY, Kim SY, Park RH, et al. A retrospective dosimetric analysis of the new ESTRO-ACROP target volume delineation guidelines for postmastectomy volumetric modulated arc therapy after implant-Based immediate breast reconstruction. Front Oncol. 2020;10: 578921.
  23. Knill C, Snyder M, Rakowski JT, Zhuang L, Matuszak M, Burmeister J. Investigating ion recombination effects in a liquid-filled ionization chamber array used for IMRT QA measurements. Med Phys. 2016;43(5):2476.
  24. Markovic M, Narayanasamy G, Stathakis S, Mavroidis P, Jurkovic IA, Saenz D, et al. Clinical evaluation of a two-dimensional liquid-filled ion chamber detector array for verification of high modulation small fields in radiotherapy. J Med Phys. 2019;44(2):91-8.
  25. McCulloch J, Pawlowski J, Kirby N, Rasmussen K, Shi Z, Myers P, Stathakis S, Papanikolaou N, Saenz DL. Patient-specific dose quality assurance of single-isocenter multiple brain metastasis stereotactic radiosurgery using PTW Octavius 4D. J Appl Clin Med Phys. 2020;21(9):107-15. https://doi.org/10.1002/acm2.12979
  26. Du W, Cho SH, Zhang X, Hoffman KE, Kudchadker RJ. Quantification of beam complexity in intensity-modulated radiation therapy treatment plans. Med Phys. 2014;41:021716.
  27. Park JM, Kim JI, Park SY. Modulation indices and plan delivery accuracy of volumetric modulated arc therapy. J Appl Clin Med Phys. 2019;20(6):12-22.  https://doi.org/10.1002/acm2.12589