Automatic Exposure Control Performance Evaluation of Digital Radiographic Imaging System by Manufacturer Using Coins

동전을 이용한 제조사 별 디지털 방사선 영상 시스템의 자동노출제어 성능 평가

  • Lim, Se-Hun (Department of Radiology, Chungbuk National University Hospital) ;
  • Seoung, Youl-Hun (Department of Radiological Science, College of Health Medical Science, Cheongju University)
  • 임세훈 (충북대학교병원 영상의학과) ;
  • 성열훈 (청주대학교 보건의료대학원 방사선학과)
  • Received : 2022.02.01
  • Accepted : 2022.02.17
  • Published : 2022.02.28

Abstract

In this study, we proposed an image quality control for an automatic exposure control (AEC) of digital radiographic imaging system and tried to analyze the performance of the AEC by various manufacturer. The subjects of the experiment were analyzed for the AEC image quality evaluation using digital radiation generators from four manufacturer such as PHILIPS, GE Healthcare, SAMSUNG Healthcare, DK Medical Solution. We used as materials for the implementation of the image quality evaluation by coins (500 won, KOMSCO, Korea). This study evaluated the performance evaluation of the AEC as image quality and exposure dose (Milliampere-seconds; mAs). The image quality evaluation was tried visual assessment by two radiologic technologists and contrast to noise (CNR) by ImageJ. The exposure dose investigated mAs on digital radiation generators. The radiographic coin images acquired 360 images based on change in the control factors of the AEC, which were kVp, the consistency of field configuration and dominant zone, sensitivity and density. As a result, there was a significant difference in the AEC performance between manufacturer. The CNR by the AEC for each manufacturer showed a difference of up to about 1.9 times. The exposed tube current by the AEC for each manufacturer showed a difference of up to about 5.8 times. It is expected that our proposed evaluation method using coins could be applied as the AEC performance evaluation method in the future.

Keywords

References

  1. Bushberg JT, Seibert JA, Leidholdt Jr. EM, Boone JM. The Essential Physics of Medical Imaging. 3rd ed. Philadelphia: Lippincott Williams & Wilkins; 2012.
  2. Seibert JA. The AAPM/RSNA Physics Tutorial for Residentents. Imaging & Therapeutic Technology. 1997;17(6):1533-57.
  3. Bushong SC. Radiologic science for technologists: Physics, biology, and protection. 10th ed. St. Louis, MO. Elsevier-Mosby; 2013.
  4. Jeong MG, Seoung YH. Effects of Field Configuration Shielding Area and Changing of Density and Sensitivity on Tube Current and Image Quality in Automatic Exposure Control System. The Korean Society of Radiology. 2020;14(5):635-42.
  5. Jung JS, Choi BW, Kim SH, Kim YM, Shim JN, Ahn HS, et al. Effect of Automacit Exposure Control Marker with Chest Radiography in Radiation Reduction. Korean Society of Radiological Science. 2014;37(3):170-85.
  6. Jeong MG, Seoung YH. Effects of Exposure Dose Reduction by Optimization of Automatic Exposure Control Factors in Digital Radiographic Examination of Paranasal Sinus. Journal of the Korean Society of Radiological Science. 2021;44(3):173-81.
  7. Choi SS, Lim CH, Jeong SH. A Study on the Using of Automatic Exposure Control in the Chest Radiography. Journal of the Korean Society of Radiological Science. 2019;42(1):19-24.
  8. Yoon SH, Choi JG, Han DK. The Influence of the Change of Patient Radiation Exposure Dose Distribution on the Grid Condition and Detector Acquisition Dose on the Exposure Distance in the Use of Amorphous Silicon Thin Film Transistor Detector with AEC. Journal of the Korean Society for Digital Imaging in Medicine. 2007;9(2):23-30.
  9. Park HS, Kim MS, Jeong HM, Lee JW. A Study on the Optimization of Image Quality and Dose in Chest PA Digital Radiography. The Korean Society of Radiology. 2017;11(1):55-61. https://doi.org/10.7742/jksr.2017.11.1.55
  10. Choi KK, Lee CY, Shin DS, Kim CN, Choi KY, Huh J. Visual Evaluation of Rib Shading and Lung Marking in High-Tube-Voltage of Chest Radiography. Journal of the Korean Society of Radiological Technology. 1992;15(1):99-105.
  11. Lim SH, Seoung YH. Correlation Analysis of Control Factors in Automatic Exposure Control of Digital Radiography System Based on Fine Contrast Images. Journal of the Korean Society of Radiological Science. 2021;44(1):1-8.
  12. Majoor JEA, Vermeer KA, Andrinopoulou ER, Lemij HG. Contrast-to-Noise Ratios for Assessing the Detection of Progression in the Various Stages of Glaucoma. Transl Vis Sci Tech. 2019;8(3):1-12.
  13. Rossi RP, Lin PJP, Rauch PL, Strauss KJ. Report of the Diagnostic X-Ray Imaging Committee Task Group on Performance Specifications and Acceptance Testing for X-Ray Generators and Automatic Exposure Control Devices. AAPM REPORT No. 14; 1985.
  14. Dobbins JT, Samei E, Chotas HG. Chest radiography: Optimization of x-ray spectrum for cesium iodide-amorphous silicon flat-panel detector. Radiology. 2003;226(1):221-30. https://doi.org/10.1148/radiol.2261012023
  15. Seoung YH, Lim JD. A Study on Reducing of Entrance Surface Dose with the Eye in the Computed Radiography by Using High Kilo Voltage Peak Technique. Journal of the Korea Safety Management & Science. 2011;13(2):91-6. https://doi.org/10.12812/KSMS.2011.13.2.091