DOI QR코드

DOI QR Code

Iatrogenic Spark Burn Injury to the Chest From a Transcutaneous Pacing Patch

경피적 심장 충격 장치의 사용으로 발생한 의인성 전기 화상의 증례 보고

  • Choi, Jong Yun (Department of Plastic and Reconstructive Surgery, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Cha, Won Jin (Department of Plastic and Reconstructive Surgery, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Jung, Ee Room (Department of Plastic and Reconstructive Surgery, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Seo, Bommie Florence (Department of Plastic and Reconstructive Surgery, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Jung, Sung-No (Department of Plastic and Reconstructive Surgery, Uijeongbu St. Mary's Hospital, College of Medicine, The Catholic University of Korea)
  • 최종윤 (가톨릭대학교 의과대학 의정부성모병원 성형외과) ;
  • 차원진 (가톨릭대학교 의과대학 의정부성모병원 성형외과) ;
  • 정이룸 (가톨릭대학교 의과대학 의정부성모병원 성형외과) ;
  • 서보미 (가톨릭대학교 의과대학 의정부성모병원 성형외과) ;
  • 정성노 (가톨릭대학교 의과대학 의정부성모병원 성형외과)
  • Received : 2021.09.07
  • Accepted : 2021.09.16
  • Published : 2021.12.31

Abstract

Iatrogenic electrical burns that occur from the use of a defibrillator, a paddle-type cardiac shock device, have been reported in various forms. Electrical burns are usually conducted directly through the skin and are more damaging than scald burns or contact burns. A transcutaneous cardiac pacing device is a patch-type cardiac shock device that automatically delivers a shock when an abnormal heart rhythm is detected. We introduce a unique case of iatrogenic electrical burns caused by the transcutaneous pacing patch of a cardiac shock device. Electrical energy was converted into a spark due to foreign bodies deposited around the patch, resulting in damage to the peripheral area of the skin.

Keywords

References

  1. Zoll PM. Resuscitation of the heart in ventricular standstill by external electric stimulation. New England Journal of Medicine 1952;247:768-71. https://doi.org/10.1056/NEJM195211132472005
  2. Madsen JK, Meibom J, Videbak R, Pedersen F, Grande P. Transcutaneous pacing: experience with the zoll noninvasive temporary pacemaker. American Heart Journal 1988;116(1, Part 1):7-10. https://doi.org/10.1016/0002-8703(88)90242-6
  3. Reisin L, Baruchin AM. Iatrogenic defibrillator burns. Burns 1990;16:128. https://doi.org/10.1016/0305-4179(90)90171-r
  4. Singleton MJ, Schoenfeld MH, Bhave PD, Beaty EH, Whalen SP. Third-degree burns caused by transcutaneous pacing for third-degree heart block. HeartRhythm Case Rep. 2020;6:495-8. https://doi.org/10.1016/j.hrcr.2020.05.006
  5. PRIDE HB, McKINLEY DF. Third-degree burns from the use of an external cardiac pacing device. Critical Care Medicine 1990;18:572. https://doi.org/10.1097/00003246-199005000-00020
  6. Chang HC, Lin JF, Yeh KH. Paddle position and contact force: an important step to check when troubleshooting for refractory ventricular fibrillation. Hong Kong Journal of Emergency Medicine 2017;23:289-92. https://doi.org/10.1177/102490791602300506
  7. Saaiq M, Zaib S, Ahmad S. Electrocautery burns: experience with three cases and review of literature. Annals of burns and fire disasters 2012;25:203.
  8. Pagan-Carlo LA, Stone MS, Kerber RE. Nature and determinants of skin "Burns" after transthoracic cardioversion. The American Journal of Cardiology 1997;79:689-91. https://doi.org/10.1016/S0002-9149(96)00845-4
  9. Payne JE, Morgan JL, Weachter RR, Alpert MA. Third-degree burns associated with transcutaneous pacing. BMJ Case Reports 2018:bcr-2018-226769.