Variation on Estimated Values of Radioactivity Concentration according to the Change of the Acquisition Time of SPECT/CT

SPECT/CT의 획득시간 증감에 따른 방사능농도 추정치의 변화

  • Kim, Ji-Hyeon (Department of Nuclear Medicine, Seoul Medical Center) ;
  • Park, Hoon-Hee (Department of Radiological Technology, Shingu College) ;
  • Lee, Joo-Young (Department of Radiological Technology, Songho University)
  • 김지현 (서울특별시 서울의료원 핵의학과) ;
  • 박훈희 (신구대학교 방사선과) ;
  • 이주영 (송호대학교 방사선과)
  • Received : 2021.10.26
  • Accepted : 2021.12.10
  • Published : 2021.12.31

Abstract

With the recent development of precision medicine(Theranostics), interest and utilization of the quantitative function of SPECT/CT are increasing. This study aims to investigate the effect on the radioactivity concentration estimate by the increase or decrease in the total time of SPECT/CT imaging conditions. A standard image was obtained by the conditions of a total acquisition time of 600 sec(10 sec/f × 120 frames) by diluting 99mTc 91.76 MBq in a cylindrical phantom filled with sterile water, and a comparative image was obtained by increasing the total acquisition time by -90%, -75%, -50%, -25%, +50%, +100%. The CNR, radioactive concentration estimate(cps/ml), and the variation rate(%) of the recovery coefficient(RC) were analyzed by measuring the overall coefficient of interest in each image. The results[CNR, Radiation Concentration, RC] by the change in the number of projections for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.5%, +3.90%, 1.04] at -90%, [-77.9%, +2.71%, 1.03] at -75%, [-55.6%, +1.85%, 1.02] at -50%, [-33.6%, +1.37%, 1.01] at -25%, [-33.7%, +0.71%, 1.01] at +50%, [+93.2%, +0.32%, 1.00] at +100%. and also The results[CNR, Radiation Concentration, RC] by the acquisition time change for each increase or decrease rate(-90%, -75%, -50%, -25%, +50%, +100%) of total acquisition time are as follows. [-89.3%, -3.55%, 0.96] at -90%, [-73.4%, -0.17%, 1.00] at -75%, [-49.6%, -0.34%, 1.00] at -50%, [-24.9%, 0.03%, 1.00] at -25%, [+49.3%, -0.04%, 1.00] at +50%, [+99.0%, +0.11%, 1.00] at +100%. Image quality(CNR) showed a pattern of change in proportion to the increase or decrease in the total acquisition time of SPECT/CT, but the result at quantitative evaluation showed a change of less than 5% in all experimental conditions, maintaining quantitative accuracy(RC less than 0.05) without much influence.

Keywords

Acknowledgement

This work was supported by the research fund of Songho University(SH-2020-A)

References

  1. ICRP. Radiological Protection in Therapy with Radiopharmaceuticals. ICRP Publication 140; 2018.11.
  2. Arda K, Joseph AO, Richard LW, Michael MG, Van Abbeele AD. Theranostics: The Role of Quantitative Nuclear. Semin Radiat Oncol. 2021;31(1):28-36. https://doi.org/10.1016/j.semradonc.2020.07.003
  3. Bin H, Eric CF. Effects of shortened acquisition time on accuracy and precision of quantitative estimates of organ activity. Am Assoc Phys Med. 2010;37(4):1807-15.
  4. Park HH, Kim TH, Shin JY, Lee TS, Lyu KY. Usefulness of CT based SPECT Fusion Image in the lung Disease: Preliminary Study. Journal of Radiological Science and Technology. 2012;35(1):59-64.
  5. NEMA. Performance Measurements of Gamma Cameras. NEMA Standards Publication NU 1-2012. 2013:44-47.
  6. IAEA. Quantitative Nuclear Medicine Imaging: Concepts, Requirements and Methods. IAEA Human Health Reports No.9. 2014;9:36.
  7. Nakamura Y, Tomiguchi S, Katsuda N. Usefulness of the collimator detector response(CDR) recovery and the scatter correction by the effective scatter source estimation(ESSE) method in myocardial SPECT study. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2010;66(6):609-17. https://doi.org/10.6009/jjrt.66.609
  8. Johannes Z, Alexander HV, Amos Y, Joachim H, Torsten K. Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction. J Nucl Med. 2010;51:921-8. https://doi.org/10.2967/jnumed.109.071571
  9. Kim JG, Ham JC, Oh SH, Kang CK, Kim JS. A Study on the Tendency of Dose value According to Dose calibrator Measurement Depth and Volume. J Nucl Med Technol. 2020;24(1):20-6.
  10. Alfonso RM, Ole Marius HR, Jan DH, Svein-Erik M, Andreas A, Lediju Bell MA. The Generalized Contrast-to-Noise Ratio: A Formal Definition for Lesion Detectability. IEEE Trans Ultrason Ferroelectr Freq Control. 2020;67(4):745-59. https://doi.org/10.1109/tuffc.2019.2956855
  11. Anil A, Jonathan R, Anthony SP, Michael LO. Improving Spatial Resolution Using Incoherent Subtraction of Receive Beams Having Different Apodizations. IEEE Trans Ultrason Ferroelectr Freq Control. 2019;66(1):5-17. https://doi.org/10.1109/tuffc.2018.2876285
  12. Lee JY, Lee EK, Kim KW, Jeong HW, Lyu KY, Park HH, Son JH, et al. Evaluation and Comparison of Contrast to Noise Ratio and Signal to Noise Ratio According to Change of Reconstruction on Breast PET/CT. Journal of Radiological Science and Technology. 2017;40(1):79-85. https://doi.org/10.17946/JRST.2017.40.1.12
  13. Kim KW, Min JW, Lyu KY, Kim JM, Jeong HW, Lee JA, Jung JH, et al. Comparison Study on CNR and SNR of Thoracic Spine Lateral Radiography. Journal of Radiological Science and Technology. 2013;36(4):273-80.
  14. Marin I, Ryden T, Van Essen M, Svensson J, Gracheva N, Koster U, et al. Establishment of a clinical SPECT/CT protocol for imaging of 161Tb. EJNMMI Physics. 2020;7(45):6-7. https://doi.org/10.1186/s40658-020-0274-7
  15. NIH Health & Human Services. Help Me Understand Genetics 'Precision Medicine'. U.S.National Library of Medicine 'MedlinePlus' Web site. Available from: https://medlineplus.gov/download/genetics/understanding/precisionmedicine.pdf.