DOI QR코드

DOI QR Code

Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique

  • Hur, Beong Ik (Department of Neurosurgery, Pusan National University Hospital) ;
  • Jin, Seong Jin (Gamma-knife Center, Haeundae Paik Hospital, Inje University) ;
  • Kim, Gyeong Rip (Gamma-knife Center, Pusan National University Yangsan Hospital) ;
  • Kwak, Jong Hyeok (Department of Radiology, Pusan National University Yangsan Hospital) ;
  • Kim, Young Ha (Department of Neurosurgery, Pusan National University Yangsan Hospital) ;
  • Lee, Sang Weon (Gamma-knife Center, Pusan National University Yangsan Hospital) ;
  • Sung, Soon Ki (Gamma-knife Center, Pusan National University Yangsan Hospital)
  • Received : 2020.06.26
  • Accepted : 2020.12.15
  • Published : 2021.01.01

Abstract

Objective : High precision and accuracy are expected in gamma knife radiosurgery treatment. Because of the requirement of clinically applying complex radiation and dose gradients together with a rapid radiation decline, a dedicated quality assurance program is required to maintain the radiation dosimetry and geometric accuracy and to reduce all associated risk factors. This study investigates the validity of Leksell Gamma plan (LGP)10.1.1 system of 5th generation Gamma Knife Perfexion as modified variable ellipsoid modeling technique (VEMT) method. Methods : To verify LGP10.1.1 system, we compare the treatment plan program system of the Gamma Knife Perfexion, that is, the LGP, with the calculated value of the proposed modified VEMT program. To verify a modified VEMT method, we compare the distributions of the dose of Gamma Knife Perfexion measured by Gafchromic EBT3 and EBT-XD films. For verification, the center of an 80 mm radius solid water phantom is placed in the center of all sectors positioned at 16 mm, 4 mm and 8 mm; that is, the dose distribution is similar to the method used in the x, y, and z directions by the VEMT. The dose distribution in the axial direction is compared and analyzed based on Full-Width-of-Half-Maximum (FWHM) evaluation. Results : The dose profile distribution was evaluated by FWHM, and it showed an average difference of 0.104 mm for the LGP value and 0.130 mm for the EBT-XD film. Conclusion : The modified VEMT yielded consistent results in the two processes. The use of the modified VEMT as a verification tool can enable the system to stably test and operate the Gamma Knife Perfexion treatment planning system.

Keywords

Acknowledgement

This research was supported by the Korea Research Foundation (NRF) funded by the government (Ministry of Science, Technology and Information) in 2017 (NRF-2017R1C1B5077022).

References

  1. Baghani HR, Aghamiri SM, Mahdavi SR, Robatjazi M, Zadeh AR, Akbari ME, et al. : Dosimetric evaluation of gafchromic EBT2 film for breast intraoperative electron radiotherapy verification. Phys Med 31 : 37-42, 2015 https://doi.org/10.1016/j.ejmp.2014.08.005
  2. Beck J, Berndt A : An algorithm for independent verification of Gamma Knife treatment plans. Med Phys 31 : 2780-2784, 2004 https://doi.org/10.1118/1.1786532
  3. Casanova Borca V, Pasquino M, Russo G, Grosso P, Cante D, Sciacero P, et al. : Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J Appl Clin Med Phys 14 : 4111, 2013
  4. Chung JP, Oh SW, Seong YM, Chun KJ, Chung HT : An effective calibration technique for radiochromic films using a single-shot dose distribution in Gamma Knife®. Phys Med 32 : 368-378, 2016
  5. Elekta : Leksell Gamma Knife Perfexion - Application Introductory Training, (Dose calculation for Leksell Gamma Knife Perfexion). Available at : https://www.elekta.com/radiosurgery/leksellgamma-knife-perfexion
  6. Hur BI, Choi BK, Sung SK, Cho WH, Cha SH, Choi CH : The variable ellipsoid modeling technique as a verification method for the treatment planning system of gamma knife radiosurgery. J Korean Neurosurg Soc 47 : 128-133, 2010 https://doi.org/10.3340/jkns.2010.47.2.128
  7. Hur BI : Comparisons between the two dose profiles extracted from Leksell GammaPlan and calculated by variable ellipsoid modeling technique. J Korean Soc Radiol 11 : 9-17, 2017 https://doi.org/10.7742/jksr.2017.11.1.9
  8. Jin JY, Drzymala R, Li Z : A simple method of independent treatment time verification in gamma knife radiosurgery using integral dose. Med Phys 31 : 3337-3344, 2004 https://doi.org/10.1118/1.1824192
  9. Khachonkham S, Dreindl R, Heilemann G, Lechner W, Fuchs H, Palmans H, et al. : Characteristic of EBT-XD and EBT3 radiochromic film dosimetry for photon and proton beams. Phys Med Biol 63 : 065007, 2018 https://doi.org/10.1088/1361-6560/aab1ee
  10. Kondziolka D, Maitz AH, Niranjan A, Flickinger JC, Lunsford LD : An evaluation of the Model C gamma knife with automatic patient positioning. Neurosurgery 50 : 429-431; discussion 431-432, 2002 https://doi.org/10.1227/00006123-200202000-00041
  11. Maitz AH, Wu A, Lunsford LD, Flickinger JC, Kondziolka D, Bloomer WD : Quality assurance for Gamma Knife stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 32 : 1465-1471, 1995 https://doi.org/10.1016/0360-3016(95)00577-L
  12. Marcu SM, Wu QJ, Pillai K, Weinhous MS : GammaPlan-Leksell Gamma Knife radiosurgery treatment planning verification method. Med Phys 27 : 2146-2149, 2000 https://doi.org/10.1118/1.1287051
  13. Najafi M, Geraily G, Shirazi A, Esfahani M, Teimouri J : Analysis of Gafchromic EBT3 film calibration irradiated with gamma rays from different systems: Gamma Knife and Cobalt-60 unit. Med Dosim 42 : 159-168, 2017 https://doi.org/10.1016/j.meddos.2017.01.003
  14. Novotny J, Bhatnagar JP, Niranjan A, Quader MA, Huq Ms, Bednarz G, et al. : Dosimetric comparison of the Leksell Gamma Knife Perfexion and 4C. J Neurosurg 109 Suppl : 8-14, 2008 https://doi.org/10.3171/JNS/2008/109/12/S3
  15. Tsai JS, Engler MJ, Rivard MJ, Mahajan A, Borden JA, Zheng Z : A formalism for independent checking of Gamma Knife dose calculations. Med Phys 28 : 1842-1849, 2001 https://doi.org/10.1118/1.1398038
  16. Wu A, Lindner G, Maitz AH, Kalend AM, Lunsford LD, Flickinger JC, et al. : Physics of Gamma Knife approach on convergent beams in stereotactic radiosurgery. Int J Radiat Oncol Biol Phys 18 : 941-949, 1990 https://doi.org/10.1016/0360-3016(90)90421-f
  17. Zhang P, Dean D, Wu QJ, Sibata C : Fast verification of Gamma Knifetrade mark treatment plans. J Appl Clin Med Phys 1 : 158-164, 2000 https://doi.org/10.1120/jacmp.v1i4.2638