DOI QR코드

DOI QR Code

Design of Capacitive Sensors for Blood Vessel Condition Using FEA Simulation; For Developing of an Implantable Telemetry System to Monitoring the Arterial Change

FEA 시뮬레이션을 이용한 혈관 상태 측정용 커패시티브 센서 설계; 체내 동맥 혈관 변화 모니터링이 가능한 이식형 텔레메트리 시스템 개발을 위한

  • Kang, So Myoung (Dept of Biomedical Eng, Graduate School of Medicine, Keimyung University) ;
  • Lee, Jae Ho (Dept of Anatomy, School of Medicine, Keimyung University) ;
  • Wei, Qun (Dept of Biomedical Eng, School of Medicine, Keimyung University)
  • Received : 2019.09.18
  • Accepted : 2019.10.23
  • Published : 2019.11.30

Abstract

For developing a wireless implantable device to monitoring the artery variation in real-time. The concept of a special vessel variation measurement capacitive sensor is presented in this paper. The sensor consists of two part; main sensor to measuring the arterial variation, and reference sensor is used to improve the accuracy of the capacitance value variation. Before sensor manufacture, a model of the sensor attached on the artery was designed in 3D to conduct in the FEA simulation to validate the validity and feasibility of the idea. The artery model was designed as layered structures and made of collagenous soft tissues with intima inside, followed by the media and the adventitia. Also, a grease layer was designed in the inner of the arterial wall to imitate the clogged arteries. The simulation was divided into two parts; sensor performance test by changing the diameter of the grease layer, and arterial wall tension test by changing the blood pressure. As the simulation results, the capacitance value measured by the proposed sensor is decreased follow the diameter of the grease increased. Also, large elastic deformation of the arterial wall since changing the blood pressure has been observed.

Keywords

References

  1. Statistics Korea, 2017 Cause of Death Statistics, 2018.
  2. J.H. Choe, “Coronary Artery Disease (Angina and Myocardial Infarction),” Korea Environmental Preservation Association, Vol. 29, No. 369, pp. 36-37, 2007.
  3. H.W. Roh, J.S. Kim, S.H. S, and H.M. Kwon, "Changes of Blood Flow Characteristics due to Catheter Insertion in the Stenotic Coronary Artery," Proceeding of the Fall Conference of the Korea Society of Mechanical Engineers, pp. 1878-1883, 2002.
  4. J.J. Im, J.H. Kim, M.C. Hwang, C.J. Kim, and C.S. Lessard, "Development of a Noninvasive/Continuous Blood Pressure Measuring System using Piezoelectric Sensor," The Transactions of the Korean Institute of Electrical Engineers, Vol. 46, No .3 pp. 440-444, 1997.
  5. S.J. Lee, M.N. Kim, J.H. Lee, H.G Lim, and J.H. Cho, "Implementation of Implantable Bluetooth Bio-telemetry System for Transmitting Acoustic Signals in the Body with Wireless Recharging Function," Journal of Korea Multimedia Society, Vol. 18, No. 5, pp. 652-662, 2015. https://doi.org/10.9717/kmms.2015.18.5.652
  6. S.H. Min, C.Z. Jin, M.H. Lee, C.A. Lee, and D.G. Paeng, "Time Synchronized Measurement & Analysis of Ultrasound Imaging from Blood with Blood Pressure in the Mock Pulsatile Blood Circulation System," Proceeding of Joint Conference by KSNVE, ASK and KSME(DC), pp. 216-216, 2017.
  7. M.J. Kim, H.J. Park, and Q. Wei, "Development of Water Level Sensor Using Capacitive Sensor to Encourage People to Drink Plenty of Water," Journal of Rehabilitation Welfare Engineering and Assistive Technology, Vol. 12, No. 2 pp. 70-77, 2018. https://doi.org/10.21288/resko.2018.12.2.70
  8. T.J.R Hughes, The Finite Element Method: Linear Static and Dynamic Finite Element Analysis, United States of America: Courier Corporation, 2012.
  9. H.Y. Park, W.H. Kim, and Y.W. Kim, "Measurement of Abdominal Aortic Diameter in Korean Adults," Journal of Korean Surgical Society, Vol. 47, No. 5 pp. 680-689, 1994.
  10. J. Graza, "Heat-resistant Dispersion-strengthened Copper Alloys," Journal of Materials Engineering and Performance, Vol. 1, No. 1, pp 113-121, 1992. https://doi.org/10.1007/BF02650042
  11. T.G. Osimitz, M.L. Eldridge, E. Sloter, W. Welsh, N. Ai, G.S. Sayler, F. Menn, and C. Toole, "Lack of Androgenicity and Estrogenicity of the Three Monomers sued in Eastman' s Tritan$^{TM}$ Copolyesters," Food and Chemical Toxicology, Vol. 50, No. 6 pp. 2196-2205, 2012. https://doi.org/10.1016/j.fct.2012.02.010
  12. G. Holzapfel, Nonlinear Solid Mechanics: A Continuum Approach for Engineering, John Wiley and Sons, Chichester, 2000.
  13. G. Holzapfel, T. Gasser, and R. Ogden, “A New Constitutive Framework for Arterial Wall Mechanics and a Comparative Study of Material Models,” Journal of Elasticity, Vol. 61, No. 1-3, pp. 148, 2000.
  14. I.K. Kang and W.S. Kim, "Blood Compatible Polymers," Polymer, Vol. 13, No. 2, pp. 96-104, 1989.
  15. S.U.S. Choi and J.A. Eastman, "Enhancing Thermal Conductivity of Fluids with Nanoparticles," Procedding of Enhancing Thermal Conductivity of Fluids with Nanoparticles Conference, pp. 12-17, 1995.
  16. V. Pockevicius, V. Markevicius, M. Cepenas, D. Andriukaitis, and D. Navikas, "Blood Glucose Level Estimation Using Interdigital Electrodes," Elektronika IR Elektraotechnika, ISSN 1392-1215, Vol. 19, No. 6, 2013.