DOI QR코드

DOI QR Code

Cementless Total Hip Arthroplasty Using Ceramic Femoral Head on Cross-Linked Ultra-High-Molecular Weight Polyethylene Liner in Patients Older than 65 Years: Minimum Five-Year Follow-Up Results

세라믹 대퇴 골두 및 교차결합 초고분자량 폴리에틸렌 라이너를 이용한 65세 이상 무시멘트형 인공 고관절 전치환술: 최소 5년 중기 추시 결과

  • Yun, Ho Hyun (Department of Orthopedic Surgery, VHS Medical Center) ;
  • Cheong, Ji Young (Department of Orthopedic Surgery, VHS Medical Center) ;
  • Sim, Hyun Bo (Department of Orthopedic Surgery, VHS Medical Center) ;
  • Park, Jae Hong (Department of Orthopedic Surgery, VHS Medical Center)
  • 윤호현 (중앙보훈병원 정형외과) ;
  • 정지영 (중앙보훈병원 정형외과) ;
  • 심현보 (중앙보훈병원 정형외과) ;
  • 박재홍 (중앙보훈병원 정형외과)
  • Received : 2017.11.04
  • Accepted : 2017.12.07
  • Published : 2018.12.30

Abstract

Purpose: To evaluate the utility of ceramic-on-polyethylene articular bearing surface when cementless total hip arthroplasty is performed in patients older than 65 years through an analysis of the minimum five-year follow-up results using the ceramic femoral head and cross-linked polyethylene liner. Materials and Methods: From March 2010 to September 2012, 51 patients (56 hips) who were older than 65 years were enrolled in this retrospective study. The mean age at surgery was $70.9{\pm}5.1years$ old. A clinical assessment was analyzed using the Harris hip score. For the radiographic assessment, the cup inclination and anteversion, stem alignment, and wear amount were measured. The postoperative complications were also determined. Results: The mean Harris hip score was improved from preoperative 48 points to postoperative 87 points (p<0.05). The mean cup inclination was $40.9^{\circ}{\pm}6.4^{\circ}$ and the mean cup anteversion was $20.3^{\circ}{\pm}8.1^{\circ}$. The mean cup anteversion of the elevated liner-used group (16 cases) was $14.3^{\circ}{\pm}7.9^{\circ}$ and the mean cup anteversion of the neutral liner used group (40 cases) was $22.4^{\circ}{\pm}9.1^{\circ}$ (p<0.05). The mean stem alignment angle was $0^{\circ}$ (range, varus $4^{\circ}$-valgus $4^{\circ}$). The mean linear wear amount was $0.458{\pm}0.041mm$ and the average annual linear wear rate was $0.079{\pm}0.032mm/yr$. Six cases (10.7%) of intraoperative periprosthetic femoral fractures were encountered. Conclusion: Based on these results, the use of a ceramic-on-polyethylene articular bearing surface in elderly patients with cementless total hip arthroplasty is beneficial. On the other hand, careful effort is needed to prevent intraoperative periprosthetic femoral fractures.

목적: 65세 이상 노인 환자들을 대상으로 세라믹 대퇴 골두 및 교차결합 폴리에틸렌 라이너를 인공 관절면으로 사용한 무시멘트형 인공 고관절 전치환술의 최소 5년 이상 임상 및 방사선 추시 결과들을 분석하여 세라믹-폴리에틸렌 관절면의 유용성 여부를 알아보고자 하였다. 대상 및 방법: 2010년 3월부터 2012년 9월까지 65세 이상 노인 환자 51명(56 고관절)을 대상으로 후향적 연구를 진행하였다. 평균 연령은 $70.9{\pm}5.1$세였다. 임상 평가는 Harris hip score를 이용하였고, 방사선 평가는 비구 컵 경사각 및 전염각, 대퇴 스템 정렬 상태, 마모를 측정하였다. 수술 후 합병증 발생 유무도 조사하였다. 결과: 임상 평가상 Harris hip score는 수술 전 평균 48점에서 최종 추시 시 평균 87점으로 향상되었다(p<0.05). 방사선 평가상 평균 비구 컵 경사각은 $40.9^{\circ}{\pm}6.4^{\circ}$, 평균 비구 컵 전염각은 $20.3^{\circ}{\pm}8.1^{\circ}$였다. 거상 라이너를 사용한 16예에서 평균 비구 컵 전염각은 $14.3^{\circ}{\pm}7.9^{\circ}$, 중립 라이너를 사용한 40예에서 평균 비구 컵 전염각은 $22.4^{\circ}{\pm}9.1^{\circ}$였다(p<0.05). 평균 대퇴 스템 정렬각은 $0^{\circ}$도(범위, 내반 $4^{\circ}$-외반 $4^{\circ}$)였다. 평균 선상 마모량은 $0.458{\pm}0.041mm$였으며 평균 연간 선상 마모율은 $0.079{\pm}0.032mm/yr$였다. 수술 후 합병증으로 대퇴 스템 주위 골절 6예(10.7%)가 발생하였다. 결론: 본 연구 결과 분석상 탈구나 세라믹-폴리에틸렌 관절면에서 기인한 합병증이 없으므로 65세 이상 노인 환자에서 무시멘트형 인공 고관절 전치환술 시행 시 세라믹-폴리에틸렌 관절면을 유용하게 사용할 수 있으나 수술 중 대퇴 스템 주위 골절 발생을 방지하기 위한 좀 더 세심한 노력이 필요하다.

Keywords

References

  1. Clarke IC, Good V, Anissian L, Gustafson A. Charnley wear model for validation of hip simulators: ball diameter versus polytetrafluoroethylene and polyethylene wear. Proc Inst Mech Eng H. 1997;211:25-36. https://doi.org/10.1243/0954411971534656
  2. Sochart DH, Porter ML. The long-term results of Charnley low-friction arthroplasty in young patients who have congenital dislocation, degenerative osteoarthrosis, or rheumatoid arthritis. J Bone Joint Surg Am. 1997;79:1599-617. https://doi.org/10.2106/00004623-199711000-00001
  3. Endo M, Tipper JL, Barton DC, Stone MH, Ingham E, Fisher J. Comparison of wear, wear debris and functional biological activity of moderately crosslinked and non-crosslinked polyethylenes in hip prostheses. Proc Inst Mech Eng H. 2002;216:111-22. https://doi.org/10.1243/0954411021536333
  4. Galvin A, Kang L, Tipper J, et al. Wear of crosslinked polyethylene under different tribological conditions. J Mater Sci Mater Med. 2006;17:235-43. https://doi.org/10.1007/s10856-006-7309-z
  5. McKellop H, Shen FW, Lu B, Campbell P, Salovey R. Development of an extremely wear-resistant ultra high molecular weight polyethylene for total hip replacements. J Orthop Res. 1999;17:157-67. https://doi.org/10.1002/jor.1100170203
  6. Semlitsch M, Lehmann M, Weber H, Doerre E, Willert HG. New prospects for a prolonged functional life-span of artificial hip joints by using the material combination polyethylene/aluminium oxide ceramin/metal. J Biomed Mater Res. 1977;11:537-52. https://doi.org/10.1002/jbm.820110409
  7. Clarke IC, Gustafson A. Clinical and hip simulator comparisons of ceramic-on-polyethylene and metal-on-polyethylene wear. Clin Orthop Relat Res. 2000;379:34-40. https://doi.org/10.1097/00003086-200010000-00006
  8. Abdel MP, Heyse TJ, Elpers ME, Mayman DJ, Su EP, Pellicci PM, et al. Ceramic liner fractures presenting as squeaking after primary total hip arthroplasty. J Bone Joint Surg Am. 2014;96:27-31. https://doi.org/10.2106/JBJS.M.00737
  9. Baek SH, Kim WK, Kim JY, Kim SY. Do alumina matrix composite bearings decrease hip noises and bearing fractures at a minimum of 5 years after THA? Clin Orthop Relat Res. 2015;473:3796-802. https://doi.org/10.1007/s11999-015-4428-1
  10. Davies AP, Willert HG, Campbell PA, Learmonth ID, Case CP. An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements. J Bone Joint Surg Am. 2005;87:18-27. https://doi.org/10.2106/JBJS.C.00949
  11. Yoon PW, Yoo JJ, Kim Y, Yoo S, Lee S, Kim HJ. The epidemiology and national trends of bearing surface usage in primary total hip arthroplasty in Korea. Clin Orthop Surg. 2016;8:29-37. https://doi.org/10.4055/cios.2016.8.1.29
  12. Lehil MS, Bozic KJ. Trends in total hip arthroplasty implant utilization in the United States. J Arthroplasty. 2014;29:1915-8. https://doi.org/10.1016/j.arth.2014.05.017
  13. Rajaee SS, Trofa D, Matzkin E, Smith E. National trends in primary total hip arthroplasty in extremely young patients: a focus on bearing surface usage. J Arthroplasty. 2012;27:1870-8. https://doi.org/10.1016/j.arth.2012.04.006
  14. Capello WN, D'Antonio JA, Naughton M. Periprosthetic fractures around a cementless hydroxyapatite-coated implant: a new fracture pattern is described. Clin Orthop Relat Res. 2014;472:604-10. https://doi.org/10.1007/s11999-013-3137-x
  15. Bozic KJ, Ong K, Lau E, et al. Risk of complication and revision total hip arthroplasty among Medicare patients with different bearing surfaces. Clin Orthop Relat Res. 2010;468:2357-62. https://doi.org/10.1007/s11999-010-1262-3
  16. Yun HH, Yoon JR, Yang JH, Song SY, Park SB, Lee JW. A validation study for estimation of femoral anteversion using the posterior lesser trochanter line: an analysis of computed tomography measurement. J Arthroplasty. 2013;28:1776-80. https://doi.org/10.1016/j.arth.2012.10.023
  17. Khan M, Della Valle CJ, Jacofsky DJ, Meneghini RM, Haddad FS. Early postoperative complications after total hip arthroplasty: current strategies for prevention and treatment. Instr Course Lect. 2015;64:337-46.
  18. Widmer KH. A simplified method to determine acetabular cup anteversion from plain radiographs. J Arthroplasty. 2004;19:387-90. https://doi.org/10.1016/j.arth.2003.10.016
  19. Yun HH, Shon WY, Yoon JR, Yang JH, Lim DS. Reliability of a PowerPoint method for wear measurement after total hip arthroplasty: a retrieval study using 3-dimensional laser scanning. J Arthroplasty. 2012;27:1530-7. https://doi.org/10.1016/j.arth.2011.12.025
  20. Yun HH, Jajodia NK, Myung JS, Oh JK, Park SW, Shon WY. Use of slide presentation software as a tool to measure hip arthroplasty wear. J Arthroplasty. 2009;24:1210-5. https://doi.org/10.1016/j.arth.2009.06.012
  21. Danoff JR, Bobman JT, Cunn G, et al. Redefining the acetabular component safe zone for posterior approach total hip arthroplasty. J Arthroplasty. 2016;31:506-11. https://doi.org/10.1016/j.arth.2015.09.010
  22. Sculco PK, Austin MS, Lavernia CJ, Rosenberg AG, Sierra RJ. Preventing leg length discrepancy and instability after total hip arthroplasty. Instr Course Lect. 2016;65:225-41.
  23. Cobb TK, Morrey BF, Ilstrup DM. The elevated-rim acetabular liner in total hip arthroplasty: relationship to postoperative dislocation. J Bone Joint Surg Am. 1996;78:80-6. https://doi.org/10.2106/00004623-199601000-00011
  24. Cobb TK, Morrey BF, Ilstrup DM. Effect of the elevated-rim acetabular liner on loosening after total hip arthroplasty. J Bone Joint Surg Am. 1997;79:1361-4. https://doi.org/10.2106/00004623-199709000-00010
  25. Insull PJ, Cobbett H, Frampton CM, Munro JT. The use of a lipped acetabular liner decreases the rate of revision for instability after total hip replacement: a study using data from the New Zealand Joint Registry. Bone Joint J. 2014;96:884-8.
  26. Kim YH, Park JW, Kim JS. Alumina delta-on-highly crosslinked-remelted polyethylene bearing in cementless total hip arthroplasty in patients younger than 50 years. J Arthroplasty. 2016;31:2800-4. https://doi.org/10.1016/j.arth.2016.05.020
  27. Orradre Burusco I, Romero R, Brun M, Lopez Blasco JJ. Cross-linked ultra-high-molecular weight polyethylene liner and ceramic femoral head in total hip arthroplasty: a prospective study at 5 years follow-up. Arch Orthop Trauma Surg. 2011;131:1711-6. https://doi.org/10.1007/s00402-011-1340-3
  28. Wang S, Zhang S, Zhao Y. A comparison of polyethylene wear between cobalt-chrome ball heads and alumina ball heads after total hip arthroplasty: a 10-year follow-up. J Orthop Surg Res. 2013;8:20. https://doi.org/10.1186/1749-799X-8-20
  29. Rajpura A, Kendoff D, Board TN. The current state of bearing surfaces in total hip replacement. Bone Joint J. 2014;96:147-56. https://doi.org/10.2106/JBJS.L.01690
  30. Berry DJ. Epidemiology: hip and knee. Orthop Clin North Am. 1999;30:183-90. https://doi.org/10.1016/S0030-5898(05)70073-0