Experimental Measurement and Monte Carlo Simulation the Correction Factor for the Medium-Energy X-ray Free-air Ionization Chamber

  • Received : 2017.12.18
  • Accepted : 2018.03.16
  • Published : 2018.11.30

Abstract

A key comparison has been made between the air-kerma standards of the National Institute of Metrology (NIM), China, and other Asia Pacific Metrology Programme (APMP) members in the medium-energy X-ray. This paper reviews the primary standard Free-air ionization chamber correction factor experimental method and Monte Carlo simulation method in the NIM. The experimental method and the Monte Carlo simulation method are adopted to obtain the correction factor for the medium-energy X-ray primary standard free-air ionization chamber at 100 kV, 135 kV, 180 kV, 250 kV four CCRI reference qualities. The correction factor has already been submitted to the APMP as key comparison data and the results are in good agreement with those obtained in previous studies. This study shows that the experimental method and the EGSnrc simulation method are usually used in the measurement of the correction factor. In particular, the application of the simulation methods is more common.

Keywords

References

  1. D. T. Burns, C. Kessler and J. P. McCaffrey, Metrologia Technical Supplement 48, 06002 (2011).
  2. D. T. Burns, T. Zhongqing, L. Yazhu, W. Guoqing and F. Cheng, Metrologia Technical Supplement 44, 06008 (2007).
  3. J. J. Wu, J. N. Sun, Y. D. Yang and Y. L. Zhang, J. Acta Metrologica Sinica, 31, 555 (2010).
  4. M. Boutillon, Physics in Medicine & Biology 43, 2061 (1998). https://doi.org/10.1088/0031-9155/43/8/005
  5. T. Kurosawa and N. Takata. CCRI(I)/05-27 Correction factors for free-air ionization chambers for X-rays transmitted through a diaphragm edge and scattered from the surface of the diaphragm aperture.
  6. D. W. O. Rogers et al., BEAMnrc users manual. Technical Rep. PIRS-898, (2005).
  7. I. Kawrakow, EGSnrc C++ Class Library, Technical Report No. PIRS-898, 2005.
  8. E. Mainegra-Hing, N. Reynaert and I. Kawrakow, Medical Physics 35, 3650 (2008). https://doi.org/10.1118/1.2955551
  9. ICRU. Fundamental Quantities and Units for Ionizing Radiation [R]. ICRU Report 33, 1980.

Cited by

  1. Development of the primary standard for low-energy and W/Mo mammography x-rays at OAP, Thailand vol.58, pp.5, 2018, https://doi.org/10.1088/1681-7575/ac08e3
  2. CALIBRATION METHOD AND ABSOLUTE MEASUREMENTS OF AIR KERMA OF DIAGNOSTIC X-RAY REFERENCE QUALITIES USING A FREE-AIR IONISATION CHAMBER vol.197, pp.1, 2018, https://doi.org/10.1093/rpd/ncab161