Delayed Lateral Row Anchor Failure in Suture Bridge Rotator Cuff Repair: A Report of 3 Cases

  • Jeong, Jae-Jung (Department of Orthopedic Surgery, Daejoen St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Ji, Jong-Hun (Department of Orthopedic Surgery, Daejoen St. Mary's Hospital, College of Medicine, The Catholic University of Korea) ;
  • Park, Seok-Jae (Department of Orthopedic Surgery, Daejoen St. Mary's Hospital, College of Medicine, The Catholic University of Korea)
  • Received : 2018.09.05
  • Accepted : 2018.10.10
  • Published : 2018.12.01

Abstract

Compared to single row repair, use of lateral row anchors in suture bridge rotator cuff repair enhances repair strength and increases footprint contact area. If a lateral knotless anchor (push-in design) is inserted into osteoporotic bone, pull-out of the lateral row anchor can developed. However, failures of lateral row anchors have been reported at several months after surgery. In our cases, even though complete cuff healing occurred, delayed pull-out of the lateral row anchor in the suture bridge repair occurred. In comparison to a conventional medial anchor, further biomechanical evaluation of the pull-out force, design, and insertion angle of the lateral anchor is needed in future studies. We report three cases with delayed pull-out of lateral row anchor in suture bridge rotator cuff repair with a literature review.

Keywords

GJGJB6_2018_v21n4_246_f0001.png 이미지

Fig. 3. Postoperative follow-up magnetic resonance imaging of right shoulder showed a suspected suture anchor (arrows) as a dark signal intensity in the glenohumeral joint (A), and arthroscopic examination confirmed that the anchor was in the glenohumeral joint (B).

GJGJB6_2018_v21n4_246_f0002.png 이미지

Fig. 1. (A, B) Immediate postoperative X-ray of right shoulder showed a good arrangement (medial 1 anchor, lateral 2 anchor) of the suture anchor in the suture bridge repair and postoperative 3-month X-ray revealed pull-out of the lateral row anchor from the greater tuberosity. (C, D) In the lateral decubitus position, revision arthroscopy using routine posterior portal was performed. Arthroscopic findings showed pull-out of the lateral anchor and complete healing of the rotator cuff; the protruding anchor was removed using the grasper.

GJGJB6_2018_v21n4_246_f0003.png 이미지

Fig. 2. Postoperative follow-up magnetic resonance imagings of left shoulder showed a suture anchor (arrows) as a dark signal intensity in the subacromial space (A, B) and arthroscopic removal was performed (C).

Table 1. Clinical Outcomes at Last Follow-up

GJGJB6_2018_v21n4_246_t0001.png 이미지

References

  1. Nho SJ, Slabaugh MA, Seroyer ST, et al. Does the literature support double-row suture anchor fixation for arthroscopic rotator cuff repair? A systematic review comparing doublerow and single-row suture anchor configuration. Arthroscopy. 2009;25(11):1319-28. https://doi.org/10.1016/j.arthro.2009.02.005
  2. Kulwicki KJ, Kwon YW, Kummer FJ. Suture anchor loading after rotator cuff repair: effects of an additional lateral row. J Shoulder Elbow Surg. 2010;19(1):81-5. https://doi.org/10.1016/j.jse.2009.05.002
  3. Tingart MJ, Apreleva M, Lehtinen J, Zurakowski D, Warner JJ. Anchor design and bone mineral density affect the pull-out strength of suture anchors in rotator cuff repair: which anchors are best to use in patients with low bone quality? Am J Sports Med. 2004;32(6):1466-73. https://doi.org/10.1177/0363546503262644
  4. Milano G, Grasso A, Zarelli D, Deriu L, Cillo M, Fabbriciani C. Comparison between single-row and double-row rotator cuff repair: a biomechanical study. Knee Surg Sports Traumatol Arthrosc. 2008;16(1):75-80. https://doi.org/10.1007/s00167-007-0382-0
  5. Meier SW, Meier JD. The effect of double-row fixation on initial repair strength in rotator cuff repair: a biomechanical study. Arthroscopy. 2006;22(11):1168-73. https://doi.org/10.1016/j.arthro.2006.07.004
  6. Park MC, ElAttrache NS, Tibone JE, Ahmad CS, Jun BJ, Lee TQ. Part I: footprint contact characteristics for a transosseousequivalent rotator cuff repair technique compared with a double-row repair technique. J Shoulder Elbow Surg. 2007;16(4): 461-8. https://doi.org/10.1016/j.jse.2006.09.010
  7. Sugaya H, Maeda K, Matsuki K, Moriishi J. Repair integrity and functional outcome after arthroscopic double-row rotator cuff repair. a prospective outcome study. J Bone Joint Surg Am. 2007;89(5):953-60. https://doi.org/10.2106/JBJS.F.00512
  8. Tsiouri C, Mok DH. Early pullout of lateral row knotless anchor in rotator cuff repair. Int J Shoulder Surg. 2009;3(3):63-5. https://doi.org/10.4103/0973-6042.59972
  9. Burkhart SS. Suture anchor insertion angle and the deadman theory. Arthroscopy. 2009;25(12):1365.
  10. Mahar A, Allred DW, Wedemeyer M, Abbi G, Pedowitz R. A biomechanical and radiographic analysis of standard and intracortical suture anchors for arthroscopic rotator cuff repair. Arthroscopy. 2006;22(2):130-5. https://doi.org/10.1016/j.arthro.2005.08.042

Cited by

  1. Suture Slippage After Arthroscopic Cuff Repair: Medial Displacement of Suture Knots on Follow-up Ultrasonography vol.9, pp.8, 2018, https://doi.org/10.1177/23259671211021820