DOI QR코드

DOI QR Code

사용 편의성과 소형화 기반의 맥파 측정장치 개발

Development of Pulse Analysis System Based on Convenience and Compactness

  • 투고 : 2017.07.03
  • 심사 : 2017.08.12
  • 발행 : 2017.08.31

초록

To improve and downsize the KIOM-Pulse Analysis System(KIOM-PAS), which has been traditionally used in clinical trials, a miniaturized PAS(M-PAS) with a pneumatic pump has been developed. M-PAS is composed of a measurement module and an arm-mounting module, on which an arm can be placed. The measurement module is equipped with a pressing component and sensor, which is a wearable wristband. The arm-mounting module includes a pneumatic motor, data acquisition board and valves. In addition, the measurement module is divided into a fixing module of band type for attachment to a pulsation site and a sensing module, which includes a sensor and a tube. The fixing module and sensing module are constructed independently, and the detachable fastening method improves the posture convenience of the subject during measurement. M-PAS has been reduced to 1/6 the size of KIOM-PAS, and the measurement time is shortened by 22%. Using a simulator, the difference between the waveforms measured by the two devices exhibited a high degree of similarity of within 3.65%. M-PAS represents improvements in size and convenience compared with KIOM-PAS, and it is expected to be widely used in clinics in the future because it improves the attachment method of the fixing module.

키워드

참고문헌

  1. K.C. KIM, H.J. Kang, Pulse Biomedical Research Methodology, Ansan, Korea: DAEYOMEDI, 2008, pp. 37-40.
  2. H.H. Ryu, S.W. Lee, J. Lee, Y.J. Lee, and J.Y. Kim, "Analysis of Physical Quantification of Pulse Types by Pulse Diagnosis Literatures," Journal of Physiology & Pathology in Korean Medicine, vol. 21, no. 6, pp. 1381-1387, 2007.
  3. H. Kim, J.Y. Kim, Y.J. Park, and Y.B. Park, "Development of pulse diagnostic devices in Korea," Integrative Medicine Research, vol. 2, no. 1, pp. 7-17, 2013. https://doi.org/10.1016/j.imr.2013.01.003
  4. Y.J. Lee, J. Lee, and J.Y. Kim, "Suggestion on an Innovative Pulse Diagnosis System based on Technical Trend Analysis," Journal of Physiology & Pathology in Korean Medicine, vol. 23, no. 1, pp. 174-179, 2009.
  5. J. Lee, Y.J. Lee, Y.J. Jeon, H.J. Lee, H.H. Ryu, and J.Y. Kim, "An Example of Test on Differences of Pulse Waveform Characteristics at Cun, Guan and Chi," Korean journal of oriental medicine, vol. 14, no. 2, pp. 107-112, 2008.
  6. Y.Z. Yoon, J.H. Joh, H.M. Johng, H.S. Shin, and K.S. Soh, "Contrivance of a Radial Pulse Measuring System with Variable Contact Pressure," J. Biomed. Eng. Res., vol. 20, no. 5, pp. 567-572, 1999.
  7. J.U. Kim, B.C. Ku, Y.M. Kim, J.H. Do, J.S. Jang, E.S. Jang, Y.J. Jeon, K.H. Kim, and J.Y. Kim, "The Concept of Sasang Health Index and Constitution-Based Health Assessment: An Integrative Model with Computerized Four Diagnosis Methods," Evidence-Based Complementary and Alternative Medicine, vol. 2013, pp. 13, 2013.
  8. J.Y. Shin, J.H. Lee, B.C. Ku, J.H. Bae, M.H. Jun, J.U. Kim, and T.H. Kim, "Effects of Acupuncture Stimulation on the Radial artery's Pressure Pulse Wave in Healthy Young Participants: Protocol for a prospective, single-Arm, Exploratory, Clinical Study," Journal of pharmacopuncture, vol. 19, no. 3, pp. 197, 2016. https://doi.org/10.3831/KPI.2016.19.020
  9. T.H. Kim, B.C. Ku, J.H. Bae, J.Y. Shin, M.H. Jun, J.W. Kang, J.H. Kim, J.H. Lee, and J.U. Kim, "Hemodynamic changes caused by acupuncture in healthy volunteers: a prospective, single-arm exploratory clinical study," BMC Complementary and Alternative Medicine, vol. 17, no. 1, pp. 274, 2017. https://doi.org/10.1186/s12906-017-1787-z
  10. C.J. Jung, J.H. Cho, M.H. Jun, Y.J. Jeon, and Y.M. Kim, "Variation Factor Assessment of Radial Artery Pulse by the Tonometry Angle of the Pulse Pressure Sensor," Journal of Sensor Science and Technology, vol. 25, no. 2, pp. 138-142, 2016. https://doi.org/10.5369/JSST.2016.25.2.138
  11. Y.J. Jeon, J.Y. Kim, J.H. Bae, H.J. Lee, K.H. Kim, and J.Y. Kim, "Linearity of pressure-response of a pulse diagnostic sensor composed of independently packaged 7 sensor modules," The Korean Institute of Electrical Engineers Bio Medical System Conference, Feb. 2011, pp. 84-86.
  12. J.H. Bae, Y.J. Jeon, J.Y. Kim, and J.U. Kim, "Novel Detection Algorithm of The Upstroke of Pulse Waveform for Continuously Varying Contact Pressure Method," Journal of the Institute of Electronics Engineers of Korea SC, vol. 49, no. 2, pp. 46-54, March 2012.
  13. L. Xu, Y. Yao, H. Wang, D. He, L. Wang, and Y. Jiang, "Morphology variability of radial pulse wave during exercise," Bio-Medical Materials and Engineering, vol. 24, pp. 3605-3611, 2014.