Study on the Fatigue Test and the Accelerated Life Test for Dental Implant using Universal-Joint Test Type

유니버설조인트 시험방식을 이용한 치과용 임플란트의 피로시험 및 가속수명시험에 관한 연구

  • Received : 2016.12.13
  • Accepted : 2017.01.10
  • Published : 2017.03.25

Abstract

Purpose : This paper is a comparative analysis results of the fatigue test for dental implants and accelerated life test by using a static type loading device commonly used in Korea and a dynamic type loading device (universal-joint) recommended by FDA. Methods : Fatigue tests of dental implant is based on ISO 14801 and classified into static load test and dynamic load test. The tests were carried out on three test specimens by four load stress steps under each loading device. For analysis on failure mode such as crack, fracture and permanent deformation of test specimens, we used X-ray three-dimensional computed tomography on test specimens before and after the fatigue tests. The design of the accelerated life test was based on the analysis results of the fatigue life data obtained from the dynamic load test and the statistical analysis software (Minitab ver.15) was used to analyze the appropriate life distribution. Results : As a result of the fatigue tests and the accelerated life tests at same acceleration condition under each test method, the fatigue life under the dynamic type loading device (universal-joint) was shorter than when static type loading device was applied. Conclusion : This paper can be used as a reference when the universal-joint type loading device for implants fatigue test is applied as ISO 14801.

Keywords

References

  1. Won, H.-Y. (2010). "Effect of Implant Types and Bone Resorption on the Fatigue Life and Fracture Characteristics of Dental Implants". Original Article, Vol. 26, No. 2, p. 122.
  2. Han, I.-S. (2006). "Fatigue Life Estimation of Implant using a Finite Element Method". The Korean Academy of Prosthodontics, Vol. 44, No. 4, pp. 414.
  3. Ministry of Food and Drug Safety, Rep. of Korea (2015). "Performance and safety evaluation of dental implants".
  4. ISO 14801 (2016). "Dentistry-Implants-Dynamic loading test for endosseous dental implants".
  5. McMullen, A. (2016). "Endosseous Dental Implant Loading Fixture". Patent Application Publication, US20160076983A1.
  6. IEC 60068-1 (2013). "Environmental testing-Part 1 : General and guidance".
  7. Wiskott, H. W., Nicholls, J. I., and Belser, U. C. (1995). "Stress Fatigue: basic principles and prosthodontic Implications". International Journal of Prosthodontics, Vol. 8, No. 2, pp. 105-116.
  8. Gratton, D. G., Aquilino, S. A., and Stanford, C. M. (2001). "Micromotion and dynamic fatigue properties of the dental implant-abutment interface", Journal of Prosthetic Dentistry, Vol. 85, No. 1, pp. 47-52. https://doi.org/10.1067/mpr.2001.112796
  9. Kim, Y.-G., Kim, T.-H., and Kang, S.-G. (2014). "Lifetime Prediction of RF SAW Duplexer Using Accelerated Life Testing". The Journal of Korea Information and Communications Society, Vol. 39A, No. 10, pp. 616-618. https://doi.org/10.7840/kics.2014.39A.10.616
  10. Nelson, W. (1990). "Accelerated Testing: Statistical Models, Test Plans and Data Analyses". John Wiley and Sons, New York.