A Study on Changes in Body Surface Temperature and Thermal Effect According to Ultrasound Mode

초음파 진단모드에 따른 체표면 온도변화와 열효과에 관한 고찰

  • Yang, Sung-Hee (Department of Radiology, Ilsin Christian Hospital) ;
  • Lee, Jin-Soo (Department of Radiology, University Haeundae Paik Hospital)
  • 양성희 (일신기독병원 영상의학과) ;
  • 이진수 (인제대학교 해운대백병원 영상의학과)
  • Received : 2017.04.26
  • Accepted : 2017.06.12
  • Published : 2017.06.30

Abstract

Recently, as the number of high-risk pregnancies increases, the use of new techniques such as Doppler, which have higher acoustic power than in the past, has been increasingly used in prenatal diagnosis and guidelines have been set up by various organizations to prevent excessive exposure. Therefore, in this study, we tried to investigate the temperature change of the body surface for each test mode according to the long time ultrasound examination and to examine the exposure time which is not influenced by the thermal effect. B mode, C mode, and PD mode according to time, and the temperature difference between exposed and unexposed sites were compared. As a result, the B mode showed a significant difference in the temperature change from 10 minutes, 50 minutes after exposed, 20 minutes from the C mode, and 30 minutes from the PD mode (p<0.01). In all three modes, the temperature difference was different (p<0.000), and PD mode was the most sensitive to temperature change. Also, it was found that the temperature rise time was shortened with the increase of the ultrasonic exposure time. Therefore, it is recommended that ultrasonography to observe the embryo or fetus should be used only for diagnostic purposes, avoiding excessive test time.

최근 고위험 임신군이 증가함에 따라 과거에 비해 음향파워 세기가 높은 도플러초음파 이용이 산전 진단에서 점차 증가하고 있으며, 여러 기관에서 가이드라인을 설정하여 과도한 노출을 자제하고 있다. 따라서 본 연구에서는 장시간 초음파검사 실시에 따른 검사모드 별 체표면의 온도변화를 알아보고 열효과에 영향을 받지 않는 노출시간을 고찰해 보고자 하였다. 시간경과에 따른 B mode, C mode, PD mode 별 초음파를 조사한 부위와 조사하지 않은 부위의 온도변화를 측정하여 차이를 비교하였으며 B mode는 초음파 조사 후 10분, 50분에서, C mode는 20분부터, PD mode는 30분부터 온도변화에 유의한 차이를 나타내었다(p<0.01). 측정시점에 따라 3개의 mode에서 모두 온도상승에 차이를 보였으며(p<0.000), PD mode가 가장 온도변화에 민감한 반응을 나타내었다. 또한 초음파 노출시간이 증가함에 따라 온도상승 시간이 짧아짐을 알 수 있었다. 따라서 배아나 태아를 관찰하기 위한 초음파검사 시 과도한 검사시간을 피하여 진단목적으로만 신속하게 사용할 것을 권고한다.

Keywords

References

  1. Park JS: Is Diagnostic Ultrasound Harmful to the Fetus?. The Journal of the Korean Medical Association., 51, 823-830, 2008 https://doi.org/10.5124/jkma.2008.51.9.823
  2. Kim SH, Im SJ, Kang HL, Choi EH, Kim YL, Kim AL: A basic research study for strengthening pregnancy and childbirth support. National Health Insurance Service., 30-37, 2015
  3. Guidelines for the safety of ultrasonic medical devices, ministry of food and drug safety in korea, 2004
  4. Ang ES Jr, Gluncic V, Duque A, Schafer ME, Rakic P: Prenatal exposure to ultrasound waves impacts neuronal migration in mice. Proceedings of the National Academy of Sciences of the United States of America. 103, 12903-12910, 2006 https://doi.org/10.1073/pnas.0605294103
  5. Sim HS: Ultrasonography Anatomy & Scan Technique 3rd edition. Hanmi Medical Publishing Co, 85-87, 2016
  6. Abramowicz JS: Prenatal exposure to ultrasound waves: is there a risk?. Ultrasound in Obstetrics & Gynecology., 29, 363-367, 2007 https://doi.org/10.1002/uog.3983
  7. Kim JW, Im HS: A Study On Temperature Changes In Thigh According To Diagnostic Ultrasonography Scan Time. Journal of the Korean Society of Radiology, 6, 201-206, 2012 https://doi.org/10.7742/jksr.2012.6.3.201
  8. Abramowicz JS, Barnett SB, Duck FA, Edmonds PD, Hynynen KH, Ziskin MC: Fetal thermal effects of diagnostic ultrasound. Journal of ultrasound in Medicine, 27, 541-559, 2008 https://doi.org/10.7863/jum.2008.27.4.541
  9. Bly S: Obstetric ultrasound biological effects and safety. Journal of Obstetrics and Gynaecology Canada, 27, 572-580, 2005 https://doi.org/10.1016/S1701-2163(16)30716-2
  10. Merritt CR, Kremkau FW, Hobbins JC: Diagnostic ultrasound bioeffects and safety. Ultrasound Obstetric & Gynecology, 2, 366-374, 1992 https://doi.org/10.1046/j.1469-0705.1992.02050366.x
  11. Sheiner E, Shoham VI, Hussey MJ, et al.: First-trimester sonography: is the fetus exposed to high levels of acoustic energy?. Journal of Clinical Ultrasound, 35, 245-249, 2007 https://doi.org/10.1002/jcu.20321
  12. Yoo SH, Choi WJ, Lee SW: A Study on the Actual Output and Thermal Effect in Tissue Mimicking Phantom by the Material of the Ultrasonic Transducer. Korean Society of Physical Medicine, 10, 91-97, 2015
  13. Jung JH, Kim JS, Ha KL, Kim MJ, Cao YG: Tissue Mimicking Phantom for Visualization of Temperature Elevation Caused by Ultrasound. The Journal of the Acoustical Society of Korea, 33, 291-299, 2014 https://doi.org/10.7776/ASK.2014.33.5.291
  14. Arther RM, Gurumurthy KV: A single-pole model for the propagation of ultrasound in soft tissue. The Journal of the Acoustical Society of America, 77, 1589-1597, 1985 https://doi.org/10.1121/1.392002