Abstract

Objectives: To identify bacterial contamination rates of laryngoscope blades and handles stored in emergency crash carts by hospital and area according to the frequency of intubation attempts. Methods: One hundred forty-eight handles and 71 blades deemed ready for patient use from two tertiary hospitals were sampled with sterile swabs using a standardized rolling technique. Samples were considered negative (not contaminated) if no colonies were present on the blood agar plate after an 18-hour incubation period. Samples were stratified by hospital and according to the frequency of intubation attempts (10 attempts per year) using the ${\chi}^2-test$ and Fisher exact test. Results: One or more species of bacteria were isolated from 4 (5.6%) handle tops, 20 (28.2%) handles with knurled surfaces, and 27 (18.2%) blades. No significant differences were found in microbial contamination levels on the handle tops and blades between the two hospitals and two areas according to the frequency of intubation attempts. However, significant differences were found between the two hospitals and two areas in the level of microbial contamination on the handles with knurled surfaces (p<0.05). Conclusions: Protocols and policies must be reviewed to standardize procedures to clean and disinfect laryngoscope blades and handles; handles should be re-designed to eliminate points of contact with the blade; and single-use, one-piece laryngoscopes should be introduced.

Keywords

References

  1. Hall JR. Blood contamination of anesthesia equipment and monitoring equipment. Anesth Analg 1994;78(6):1136-1139.
  2. Morell RC, Ririe D, James RL, Crews DA, Huffstetler K. A survey of laryngoscope contamination at a university and a community hospital. Anesthesiology 1994;80(4):960. https://doi.org/10.1097/00000542-199404000-00038
  3. Lee SO. A survey on hospital infection. J Nurs Acad Soc 1993;23(3):325-338 (Korean). https://doi.org/10.4040/jnas.1993.23.3.325
  4. Kim JM. Current state of nosocomial infections in Korea. Korean J Med 1999;57(4):572-577 (Korean).
  5. Cullen MM, Trail A, Robinson M, Keaney M, Chadwick PR. Serratia marcescens outbreak in a neonatal intensive care unit prompting review of decontamination of laryngoscopes. J Hosp Infect 2005;59(1):68-70. https://doi.org/10.1016/j.jhin.2004.08.003
  6. Jones BL, Gorman LJ, Simpson J, Curran ET, McNamee S, Lucas C, et al. An outbreak of Serratia marcescens in two neonatal intensive care units. J Hosp Infect 2000;46(4):314-319. https://doi.org/10.1053/jhin.2000.0837
  7. Rutala WA, Weber DJ; Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for disinfection and sterilization in healthcare facilities, 2008 [cited 2017 Apr 28]. Available from: https://www.cdc.gov/hai/pdfs/disinfection_nov_2008.pdf.
  8. Ministry of Government Legislation. Guidelines for medical instruments and material disinfection in healthcare facilities [cited 2014 Aug 13]. Available from: http://www.law.go.kr/admRulLsInfoP.do?admRulSeq=2000000014476 (Korean).
  9. AORN Recommended Practices Committee. Recommended practices for cleaning, handling and processing anesthesia equipment. AORN J 2005;81(4):856-857, 860-870. https://doi.org/10.1016/S0001-2092(06)60365-7
  10. Muscarella LF. Reassessment of the risk of healthcare-acquired infection during rigid laryngoscopy. J Hosp Infect 2008;68(2):101-107. https://doi.org/10.1016/j.jhin.2007.11.004
  11. Kim HY, Lee EJ, Jang EK, Park YA. A study for infection control standards for medical devices in NICU. J Korean Clin Nurs Res 2010;16(2):69-84 (Korean).
  12. Williams D, Dingley J, Jones C, Berry N. Contamination of laryngoscope handles. J Hosp Infect 2010;74(2):123-128. https://doi.org/10.1016/j.jhin.2009.09.015
  13. Negri de Sousa AC, Levy CE, Freitas MI. Laryngoscope blades and handles as sources of cross-infection: an integrative review. J Hosp Infect 2013;83(4):269-275. https://doi.org/10.1016/j.jhin.2012.10.015
  14. Esler MD, Baines LC, Wilkinson DJ, Langford RM. Decontamination of laryngoscopes: a survey of national practice. Anaesthesia 1999;54(6):587-592. https://doi.org/10.1046/j.1365-2044.1999.00755.x
  15. Simmons SA. Laryngoscope handles: a potential for infection. AANA J 2000;68(3):233-236.
  16. Lowman W, Venter L, Scribante J. Bacterial contamination of re-usable laryngoscope blades during the course of daily anaesthetic practice. S Afr Med J 2013;103(6):386-389. https://doi.org/10.7196/SAMJ.6385
  17. Call TR, Auerbach FJ, Riddell SW, Kiska DL, Thongrod SC, Tham SW, et al. Nosocomial contamination of laryngoscope handles: challenging current guidelines. Anesth Analg 2009;109(2):479-483. https://doi.org/10.1213/ane.0b013e3181ac1080
  18. EBIOX. Trionic: advanced environmental hygiene [cited 2014 Mar 2]. Available from: http://www.ebiox.co.uk/products/view/trionic/.
  19. Qureshi T, Barbut F, Pernet P, Neyme D, Maury E, Offenstadt G. Laryngoscope handles in a medical intensive care unit: the level of bacterial and occult blood contamination. J Hosp Infect 2008;68(1):94-95. https://doi.org/10.1016/j.jhin.2007.10.014
  20. Telang R, Patil V, Ranganathan P, Kelkar R. Decontamination of laryngoscope blades: is our practice adequate? J Postgrad Med 2010;56(4):257-261. https://doi.org/10.4103/0022-3859.70930
  21. Howell V, Thoppil A, Young H, Sharma S, Blunt M, Young P. Chlorhexidine to maintain cleanliness of laryngoscope handles: an audit and laboratory study. Eur J Anaesthesiol 2013;30(5):216-221. https://doi.org/10.1097/EJA.0b013e3283607827
  22. Hwang SH, Park DU, Ha KC, Cho HW, Yoon CS. Airborne bacteria concentrations and related factors at university laboratories, hospital diagnostic laboratories and a biowaste site. J Clin Pathol 2011;64(3):261-264. https://doi.org/10.1136/jcp.2010.084764
  23. Machan MD, Monaghan WP, McDonough J, Hogan G. Emerging evidence in infection control: effecting change regarding use of disposable laryngoscope blades. AANA J 2013;81(2):103-108.
  24. Anderson K, Gambhir S, Glavin R, Kinsella J. The use of an anaesthetic simulator to assess single-use laryngoscopy equipment. Int J Qual Health Care 2006;18(1):17-22. https://doi.org/10.1093/intqhc/mzi091

Cited by

  1. Contamination and Disinfection of Rigid Laryngoscopes: A Literature Review vol.110, pp.1, 2017, https://doi.org/10.1002/aorn.12724
  2. A cost comparison between reusable flexible and disposable laryngoscopes vol.43, pp.2, 2022, https://doi.org/10.1016/j.amjoto.2021.103321