Abstract
In diagnostic medical imaging, it is essential to reduce the scattered radiation for the high medical image quality and low patient dose. Therefore, in this study, the influence of the scattered radiation on medical images was analyzed as the tube voltage increases. For this purpose, ANSI chest phantom was used to measure the scattering ratio, and the scattering effect on the image quality was investigated by RMS evaluation, RSD and NPS analysis. It was found that the scattering ratio with increasing x-ray tube voltage gradually increased to 48.8% at 73 kV tube voltage and to 80.1% at 93 kV tube voltage. As a result of RMS analysis for evaluating the image quality, RMS value according to increase of tube voltage was increased, resulting in low image quality. Also, the NPS value at 2.5 lp/mm spatial frequency was increased by 20% when the tube voltage was increased by 93 kV compared to the tube voltage of 73 kV. From this study, it can be seen that the scattering radiation have a significant effect on the image quality according to the increase of x-ray tube voltage. The results of this study can be used as basic data for the improvement of medical imaging quality.
진단목적의 방사선 의료 영상의 활용의 증가에 따라 환자의 피폭 선량 증가와 의료영상의 진단적 가치 저하에 기여하는 산란선의 관리 및 저감을 위한 연구는 필수적이라 할 수 있다. 이에 본 연구에서는 관전압 증가에 따른 산란선의 증감이 영상 화질에 미치는 영향을 분석하였다. 이를 위해 ANSI 흉부 팬텀을 이용하여, 관전압 변화에 따른 산란선 발생 비율을 측정하고, 산란선의 발생에 따른 화질 영향을 RMS(Root Mean Square) 입상성 평가, RSD(Relative Standard Deviation) 및 NPS(Noise Power Spectrum) 분석을 통해 고찰하였다. 관전압 증가에 따른 산란선 발생비율은 73 kV 관전압에서 48.8%, 93 kV 관전압 인가시 80.1%로 점차 증가하는 것으로 확인되었다. 관전압 증가에 따른 산란선 증가의 화질 영향을 분석하기 위한 RMS 분석 결과, 관전압 증가에 따른 RMS 값이 증가하는 것으로 나타나 영상의 입상성이 떨어지는 결과로 도출되었다. 공간주파수 2.5 lp/mm에서의 NPS 값 또한 관전압 73 kV 인가에 비해 93kV 관전압 증가시 20% 정도 영상의 잡음이 증가하는 것으로 나타났다. 본 연구를 통하여 관전압 변화에 따른 산란선 발생이 화질에 미치는 영향을 확인할 수 있었으며, 이러한 연구 결과는 의료영상 품질 개선을 위한 연구의 기초 자료로서 활용 가능할 것으로 판단된다.