An Experimental Comparative Study of Radiography, Ultrasonography and CT Imaging in the IV Catheter Fragment

정맥내 카테터 조각의 엑스선, 초음파 및 CT 영상의 실험적 비교 연구

  • Kweon, Dae Cheol (Department of Radiological Science, College of Health Science, Shinhan University)
  • 권대철 (신한대학교 보건과학대학 방사선학과)
  • Received : 2016.04.30
  • Accepted : 2016.06.08
  • Published : 2016.06.30

Abstract

The objective of this study was to detect the fragments generated during IV (intravenous) catheter injection of contrast medium and drug administration in a clinical setting and removal was performed by experimentally producing a phantom, and to compare the radiography, ultrasonography, and multi-detector computed tomography (MDCT) imaging and radiation dose. A 1 cm fragment of an 18 gage Teflon$^{(R)}$ IV catheter with saline was inserted into the IV control line. Radiography, CT, and ultrasonography were performed and radiography and CT dose were calculated. CT and ultrasonography showed an IV catheter fragment clinically and radiography showed no visible difference in the ability to provide a useful image of an IV catheter fragment modality (p >.05). Radiography of effective dose ($0.2139mSv{\cdot}Gy^{-1}{\cdot}cm^{-2}$) form DAP DAP ($0.93{\mu}Gy{\cdot}m^2 $), and dose length product (DLP) ($201mGy{\cdot}cm$) to effective dose was calculated as 0.483 mSv. IV catheter fragment were detected of radiography, ultrasonography and CT. These results can be obtained by menas of an excellent IV catheter fragment of detection capability CT. However, CT is followed by radiation exposure. IV catheter fragment confirming the position and information recommend an ultrasonography.

조영증강을 위해 조영제를 정맥내 카테터로 주입하는 과정에서 정맥태 카테터 조각을 검출하고 제거하기 위해 식염수와 18 게이지 테프론 IV 카테터의 1cm 조각을 삽입하여 팬텀을 제작하여 엑스선, 초음파 및 CT 영상을 비교하고자 한다. 엑스선, 초음파 및 CT에서 정맥내 카테터 조각을 임상적으로 유용하게 보였고, 조각의 묘출 능력에는 차이가 유의한 차이가 없었다.(p >.05). 방사선 선량은 엑스선에서 DAP ($0.93{\mu}Gy{\cdot}m^2$) 및 유효선량 ($0.2139mSv{\cdot}Gy^{-1}{\cdot}cm^{-2}$), CT의 DLP ($201mGy{\cdot}cm$), 유효선량은 0.483 mSv로 산출되었다. 엑스선, 초음파 및 CT에서 정맥내 카테터 조각을 임상적으로 검출하였고, 조각의 묘출 능력에는 CT가 우수하였다. 그러나 엑스선 및 CT는 방사선 피폭을 수반하여 정맥내 카테터 조각의 위치 및 정보를 확인하는 데는 초음파 검사를 추천한다.

Keywords

References

  1. Turner DP, Sommers SC: Accidental passage of polyethylene catheter from cubital vein to right atrium. N Eng J Med, 251, 744, 1954 https://doi.org/10.1056/NEJM195410282511806
  2. Denny MA, Frank LR: Ventricular tachycardia secondary to port-a-cath fracture and embolization. J Emerg Med, 24(1), 29-34, 2003 https://doi.org/10.1016/S0736-4679(02)00664-9
  3. Kweon, DC, Lee JW, Choi JW, Yang SH, Dong KR, Chung W: An IV catheter fragment from the metacarpal vein during MDCT scanning as a human error: experimental and reproducible microscopic magnification analysis. J Radiat Prot, 36(4), 195-199, 2011
  4. Kweon DC, Yoo BG, Yang SH, Kim JG: Findings of an intravenous catheter fragments in the vein using the 3D image reformations of MDCT. Korean J Med Phys, 17, 167-172, 2006
  5. Benson LS, Sathy MJ, Port RB: Forearm compartment syndrome due to automated injection of computed tomography contrast material: Case report. J Orthop Trauma, 10, 433-436, 1996 https://doi.org/10.1097/00005131-199608000-00012
  6. Kweon DC, Kim TH, Yang SH, Yoo BG, Kim MG, Park P: Subcutaneous Injection contrast media extravasation 3D CT appearance. Korean J Med Phys, 16, 47-51, 2005
  7. Shannon FL, McCroskey BL, Moore EE, Moore FA: Venous bullet embolism: rationale for mandatory extraction. J Trauma, 27, 1118-1122, 1987 https://doi.org/10.1097/00005373-198710000-00004
  8. Hansson B, Karambstsakidou A: Relationships Between Entrance Skin Dose, Effective Dose and Dose Area Product for Patients in Diagnostic and Interventional Cardiac Procedures. Radiat Pro Dosimetry, 90, 141-144, 2000 https://doi.org/10.1093/oxfordjournals.rpd.a033104
  9. European Commission. European guidelines on quality criteria for computed tomography (EUR 16262 EN). Luxembourg, Luxembourg: European Commission, 2000
  10. Gschwind CR. The intravenous foreign body: A report of 2 cases. J Hand Surg. 27A, 350-354, 2002
  11. Leunens G, Verstraete J, Van den Bogaert W, Van Dam J, Dutreix A, Van der Schueren E: Human errors in data transfer during preparation and delivery of radiation treatment affecting the final result: garbage in, garbage out. Radiother Oncol, 23(4), 217-22, 1992 https://doi.org/10.1016/S0167-8140(92)80124-2
  12. Gabelmann A, Kramer S, Gorich J: Percutaneous retrieval of lost or misplaced intravascular objects. Am J Roentgenol, 176, 1509-1513, 2001 https://doi.org/10.2214/ajr.176.6.1761509
  13. Vadlamani P, Dawn B, Perry MC: Catheter fragments and embolization from totally implanted venous access ports. Angiology, 49, 1013-1016, 1998 https://doi.org/10.1177/000331979804901207
  14. Weiss CR, Fishman EK: Multidetector 3D CT of pulmonary embolism of a peripheral intravenous line. Emerg Radiol, 11, 247-249, 2005 https://doi.org/10.1007/s10140-004-0393-9
  15. Jang KJ, Kweon DC: HU threshold for IV catheter fragment in peripheral vein of volume rendering 3D MDCT imaging. J Korea Contents Assoc, 7(4), 206-212, 2007 https://doi.org/10.5392/JKCA.2007.7.4.206
  16. Gulcu N, Karaaslan K, Kandirali E, Kocoglu H: Detection of a retained epidural catheter fragment. Reg Anesth Pain Med, 31, 589-590, 2006 https://doi.org/10.1097/00115550-200611000-00026
  17. Lee SY, Na HS, Kim MH, Kim CS, Jeon YT, Hwang JW, Do SH: A sheared catheter fragment in the wrist after arterial cannulation attempt: a case report. Korean J Crit Care Med, 25(2), 118-121, 2010 https://doi.org/10.4266/kjccm.2010.25.2.118
  18. Aras MH, Miloglu O, Barutcugil C, Kantarci M, Ozcan E, Harorli A: Comparison of the sensitivity for detecting foreign bodies among conventional plain radiography, computed tomography and ultrasonography. Dentomaxillofac Rad, 39, 72-78, 2010 https://doi.org/10.1259/dmfr/68589458