Digital Blood Pressure Estimation with the Differential Value of the Arterial Pulse Waveform

맥파의 차동값에 의한 디지털 방식의 혈압 추정 기법

  • 김보연 (대진대학교 컴퓨터공학과) ;
  • 장윤석 (대진대학교 컴퓨터공학과)
  • Received : 2016.04.04
  • Accepted : 2016.06.08
  • Published : 2016.06.30

Abstract

We proposed the new method to estimate the blood pressure with the differential value of the digital arterial pulse waveform and BP relation equation. To get the digital arterial pulse waveform, we use the arterial pulse waveform measurement system that has digital air-pressure sensor device and smart phone. The acquired digital arterial pulse waveforms are classified as hypertension group, normal group, and hypotension group, and we can derive the average differential value between the highest point and lowest point of a single waveform of individuals along with the group. In this study, we found the functional correlation between the blood pressure and differential value as a form of BP relation equation through the regression process on the average of differential value and blood pressure value from a tonometer. The Experimental results show the BP relation equation can give easy blood pressure estimation method with a high accuracy. Although this estimation method has over 66 % error rate and does not give the high level of the accuracy for the diastolic compares to the commercial tonometer, the estimation results for the systolic show the high accuracy that has less than 10 % error rate.

본 연구에서는 디지털 방식으로 측정된 맥파의 파형으로부터 최대 높이와 최소 높이의 차이, 즉 맥파의 차동값을 이용하여 혈압을 측정하는 측정 기법을 제시한다. 이를 위하여 본 연구에서는 적분 방식의 디지털 압력 센서와 블루투스, 스마트폰으로 구성되는 맥파 측정 시스템을 이용하여 맥파 데이터를 수집하였다. 수집된 맥파 데이터들은 고혈압, 정상, 저혈압의 군으로 분류되고, 각 맥파 파형의 최고 값과 최저 값의 개인별 차동값 평균을 도출한다. 맥파 측정시 동시에 혈압계로 측정한 최고혈압과 최저혈압의 값과 맥파 데이터로부터 도출된 차동값의 평균값들을 회귀 분석하면 맥파 차동값과 혈압과의 상관관계인 혈압상관관계식을 도출할 수 있다. 이 혈압상관관계식을 이용하여 임의의 실험자의 맥파 측정을 통한 혈압 추정 실험을 수행한 결과 실험자들의 혈압값들을 용이하게 추정할 수 있었다. 기존의 혈압계를 사용한 측정치에 대하여 제안된 기법은 최저혈압의 경우는 최대 66 %의 오차를 보이므로 그다지 높은 정확성을 보이지 못하지만, 최고혈압의 경우에는 10 % 이하의 오차로 혈압값을 추정할 수 있었다.

Keywords

References

  1. K. Barbe and W. V. Moer, "An Innovative Oscillometric Blood Pressure Measurement: Getting Rid of the Traditional Envelope," in Proceedings of the 2012 Medical Measurements and Applications (MeMeA), IEEE Conference Publications, 2012.
  2. B. Williams, "Achieving Blood Pressure Control in Clinical Practice," Journal of the American Society of Hypertension, Vol.2, No.4, pp.10-15, 2008. https://doi.org/10.1016/j.jash.2008.03.009
  3. Y. Gen, O. Nariaki, S. Gaku, A. Daisaku, S. Shibata, U. Satoshi, and T. Osamu, "Effects of Perindopril on 24-H Blood Pressure in Pypertensive Patients with Diabetic Nephropathy," American Journal of Hypertension, Vol.17, No.5, 2004.
  4. T. Osamu, K. Yoshihiro, Y. Youji, and T. Ikuo, "Noninvasive Measurement of Baroreflex Sensitivity Index using an Indirect and Continuous Blood-Pressure Recorder," Japanese Heart Journal, Vol.27, No.6, pp.849-857, 2008. https://doi.org/10.1536/ihj.27.849
  5. J. M. Kang, T. Yo, and H. C. Kim, "A Wrist-Worn Integrated Health Monitoring Instrument with a Tele-Reporting Device for Telemedicine and Telecare," IEEE Transactions on Instrument and Measurement, Vol.55, No.5, 2006.
  6. M. C. Ahn, J. G. Choi, I. H. Son, S. S. Lee, and K. H. Kim, "Estimated Blood Pressure Algorithm of Wrist Wearable Pulsimeter using by Hall Device," Journal of the Korean Magnetics Society, Vol.20, No.3, pp.106-113, 2010. https://doi.org/10.4283/JKMS.2010.20.3.106
  7. W. Ryu, E. Kim, K. An, S. Woo, and Y Chang, "A Bluetooth based 5-HD Measurement System for u-Healthcare," International Journal of Control and Automation, Vol.6, No.1, pp.141-150, 2013.
  8. Y. D. Cha and G. Yoon, "Ubiquitous Health Monitoring System for Multiple Users using a Zigbee and WLAN Dual-Network," Telemedicine and e-Health, Vol.15 No.9, pp.891-897, Nov., 2009. https://doi.org/10.1089/tmj.2009.0049
  9. Y. Chang and B. Kim, "A Wireless ECG Measurement System based on Zigbee USN," The Korea Information Processing Society Transactions: Part C, Vol.18-C, No.3, pp.195-198, 2011.
  10. D. Nair, S-Y. Tan, H-W. Gan, S-F. Lim, J. Tan, M. Zhu, H. Gao, N-H Chua, W-L. Peh, and K-H. Mak, "The Use of Ambulatory Tonometric Radial Arterial Wave Capture to Measure Ambulatory Blood Pressure: The Validation of a Novel Wrist-Bound Device in Adults," Journal of Human Hypertension, Vol.22, No.3, pp.220-222, 2008. https://doi.org/10.1038/sj.jhh.1002306
  11. Y. Yoon, J. H. Choi, and G. Yoon, "Non-Constrained Blood Pressure Monitoring using ECG and PPG for Personal Healthcare," Journal of Medical Systems, Vol.33, No.4, pp. 261-266, 2009. https://doi.org/10.1007/s10916-008-9186-0
  12. W. Shin, Y. Cha, and G. Yoon, "ECG/PPG Integer Signal Processing for a Ubiquitous Health Monitoring System," Journal of Medical Systems, Vol.34, No.5, pp.891-898, 2010. https://doi.org/10.1007/s10916-009-9304-7
  13. G. Ray, J. Nawarskas, and J. Anderson, "Blood Pressure Monitoring Technique Impacts Hypertension Treatment," Journal of General Internal Medicine, Vol.27, No.6, pp. 623-629, 2012. https://doi.org/10.1007/s11606-011-1937-9
  14. B. Kim, W. Jin, S. H. Woo, and Y. Chang, "A New Approach on Digital Blood Pressure Measurement Method for u-Healthcare Systems," International Journal of Bio-Science and Bio-Technology, Vol.7, No.1, pp.169-178, 2015. https://doi.org/10.14257/ijbsbt.2015.7.1.17