비인두암 토모테라피 시 체중 감소에 따른 피부선량 변화 분석

Analysis of Changes in Skin Dose During Weight Loss when Tomotherapyof Nasopharynx Cancer

  • 장준영 (한림국제대학원대학교 국제방사선학과) ;
  • 김대현 (강동경희대학교병원 흉부외과) ;
  • 최천웅 (강동경희대학교병원 호흡기내과) ;
  • 김보희 (한양대학교 간호대학 간호학과) ;
  • 박철수 (한림성심대학교 방사선학과)
  • Jang, Joon-Young (Department of International Radiology, Hallym University of Graduate Studies) ;
  • Kim, Dae Hyun (Department of Thoracic and Cardiovascular Surgery, Kyunghee University Hospital at Gangdong) ;
  • Choi, Cheon Woong (Department of Respiratory and Critical Care Medicine, Kyunghee University Hospital at Gangdong) ;
  • Kim, Bo-Hui (Department of Nursing, College of Nursing Hanyang University) ;
  • Park, Cheol-Soo (Department of Radiological Science, Hallym Polytechnic University)
  • 투고 : 2016.05.12
  • 심사 : 2016.06.02
  • 발행 : 2016.06.30

초록

항암요법과 방사선치료를 병행하여 두경부암을 치료하는 환자의 경우 항암요법과 방사선치료를 거듭할수록 구토, 메스꺼움, 식욕부진 등의 이유로 환자의 체중 감소가 생기게 된다. 체중 감소는 목의 두께 변화로 나타날 수 있으며 이로 인해 치료하고자 하는 t 타겟과 주위 정상조직의 선량 전달에도 영향을 미치며 방사선이 전달되는 방향에 위치한 피부선량의 변화가 나타날 것으로 사료된다. 특히 비인두암 환자의 경우 비인두 구조가 다른 장기보다 복잡한 구조를 지니고 있고 치료방법인 토모테라피는 수 mm 차이로 급격한 선량 변화가 발생하므로 이러한 체중 감소의 변화에 주의 깊은 관찰이 필요하다. 현재 토모테라피의 경우 매 회 치료 전 정확한 자세 재현을 위한 영상 검증을 통해 이러한 움직임을 확인하고 그 값을 보정하여 치료를 수행한다. 그러나 환자의 체중 감소로 인하여 피부선량에 전달되는 선량 변화의 검증은 되지 않고 있다. 환자의 피부선량이 증가함으로써 나타나는 부작용은 환자의 삶의 질에 영향을 미칠 뿐 만 아니라 2차 암 발생률을 높일 수 있다. 이에 환자의 체중 감소로 인하여 나타나는 환자의 목 두께를 휴먼 팬톰에 볼루스를 올리고 두께를 조절해가며 변화시켜 필름을 이용하여 피부선량의 변화가 어떻게 나타나는지 분석해보고자 한다.

For patients receiving chemotherapy and radiation therapy treatment progresses as vomiting, nausea, weight of the patient because of a loss of appetite it is reduced. The patient's weight and the distance from the skin and the treatment site is expected to be closer, thereby reducing the change in the skin because of this dose. This study tests using a loose see the difference between the volume change appears as the weight of the patient using the same phantom and the phantom body of the patient. To using the same as the position EBT film is attached to the skin of the treatment site and was adjusted to the thickness of the Bolus. And using a computerized treatment planning only tomotherapy equipment was passed under the conditions according to the thickness of the radiation dose. To baseline for accurate reproduction position using the MVCT was applied to treated with verification. By passing a total of three dose reduced the error, it was a measure of the film by using a dedicated scanner, EBT VIDAR scanner. Got an increase in the skin dose is displayed each time the thickness of the bolus reduced, in a bolus was completely removed with the highest value. If the changes appeared dose was greater weight loss patients to chemotherapy and therefore bolus thickness variation considering the weight loss of the patient when applying the tomotherapy of nasopharynx cancer was found that the increase in skin dose be increased. This large patient before treatment due to weight loss over the image verification is considered to be established should consider how to re-create your mask and treatment plan for fixing it.

키워드

참고문헌

  1. Korea Central Cancer Registry, National Cancer Center. Annual report of cancer statistics in Korea in 2012, Ministry of Health and Welfare (2014).
  2. P. Jardel, J. Thariat, P. Blanchard, F. Elloumi, N. Toumi, R. J. Bensadoun, M. Frikha, J. Daoud, and J. Bourhis, Bulletin du Cancer. 5, 1 (2014).
  3. X. Wang, J. Lu, X. Xiong, G. Zhu, H. Ying, S. He, W. Hu, and C. Hu, Med. Dosim. 35, 151 (2010). https://doi.org/10.1016/j.meddos.2009.06.007
  4. S. A. Bhide, M. Davies, and K. Burke, Int. J. Radiat. Oncol. Biol. Phys. 76, 1360 (2010). https://doi.org/10.1016/j.ijrobp.2009.04.005
  5. D. I. Rosenthal, M. S. Chambers, C. D. Fuller, N. C. Rebueno, J. Garcia, and M. S. Kies, Int. J. Radia. Oncol. Biol. Phys. 72, 747 (2008). https://doi.org/10.1016/j.ijrobp.2008.01.012
  6. N. Lee, C. Chuang, J. M. Quivey, T. L. Phillips, P. Akazawa, and L. J. Verhey, Int. J. Radiat. Oncol. Biol. Phys. 53, 630 (2002). https://doi.org/10.1016/S0360-3016(02)02756-6
  7. M. J. Butson, T. Cheung, and P. K. Yu, Appl. Radiat. Isot. 64, 60 (2006). https://doi.org/10.1016/j.apradiso.2005.07.002
  8. K. Tournel, D. Verellen, M. Duchateau, Y. Fierens, N. Linthout, and T. Reynders, Radiother Oncol. 84, 34 (2007). https://doi.org/10.1016/j.radonc.2007.06.003