DOI QR코드

DOI QR Code

Delayed Foreign Body Reaction Caused by Bioabsorbable Plates Used for Maxillofacial Fractures

  • Jeon, Hong Bae (Department of Plastic and Reconstructive Surgery, Dankook University Hospital, Dankook University College of Medicine) ;
  • Kang, Dong Hee (Department of Plastic and Reconstructive Surgery, Dankook University Hospital, Dankook University College of Medicine) ;
  • Gu, Ja Hea (Department of Plastic and Reconstructive Surgery, Dankook University Hospital, Dankook University College of Medicine) ;
  • Oh, Sang Ah (Department of Plastic and Reconstructive Surgery, Dankook University Hospital, Dankook University College of Medicine)
  • Received : 2015.03.31
  • Accepted : 2015.06.01
  • Published : 2016.01.15

Abstract

Background Bioabsorbable plates and screws are commonly used to reduce maxillofacial bones, particularly in pediatric patients because they degrade completely without complications after bone healing. In this study, we encountered eight cases of a delayed foreign body reaction after surgical fixation with bioabsorbable plates and screws. Methods A total of 234 patients with a maxillofacial fracture underwent surgical treatment from March 2006 to October 2013, in which rigid fixation was achieved with the Inion CPS (Inion, Tampere, Finland) plating system in 173 patients and Rapidsorb (Synthes, West Chester, PA, USA) in 61 patients. Their mean age was 35.2 years (range, 15-84 years). Most patients were stabilized with two- or three-point fixation at the frontozygomatic suture, infraorbital rim, and anterior wall of the maxilla. Results Complications occurred in eight (3.4%) of 234 patients, including palpable, fixed masses in six patients and focal swelling in two patients. The period from surgical fixation to the onset of symptoms was 9-23 months. Six patients with a mass underwent secondary surgery for mass removal. The masses contained fibrous tissue with a yellow, grainy, cloudy fluid and remnants of an incompletely degraded bioabsorbable plate and screws. Their histological findings demonstrated a foreign body reaction. Conclusions Inadequate degradation of bioabsorbable plates caused a delayed inflammatory foreign body reaction requiring secondary surgery. Therefore, it is prudent to consider the possibility of delayed complications when using bioabsorbable plates and surgeons must conduct longer and closer follow-up observations.

Keywords

References

  1. Bergsma JE, de Bruijn WC, Rozema FR, et al. Late degradation tissue response to poly(L-lactide) bone plates and screws. Biomaterials 1995;16:25-31. https://doi.org/10.1016/0142-9612(95)91092-D
  2. Rubin JP, Yaremchuk MJ. Complications and toxicities of implantable biomaterials used in facial reconstructive and aesthetic surgery: a comprehensive review of the literature. Plast Reconstr Surg 1997;100:1336-53. https://doi.org/10.1097/00006534-199710000-00043
  3. Orringer JS, Barcelona V, Buchman SR. Reasons for removal of rigid internal fixation devices in craniofacial surgery. J Craniofac Surg 1998;9:40-4. https://doi.org/10.1097/00001665-199801000-00009
  4. Wiltfang J, Merten HA, Schultze-Mosgau S, et al. Biodegradable miniplates (LactoSorb): long-term results in infant minipigs and clinical results. J Craniofac Surg 2000;11:239-43. https://doi.org/10.1097/00001665-200011030-00006
  5. Imola MJ, Hamlar DD, Shao W, et al. Resorbable plate fixation in pediatric craniofacial surgery: long-term outcome. Arch Facial Plast Surg 2001;3:79-90. https://doi.org/10.1001/archfaci.3.2.79
  6. Mackool R, Yim J, McCarthy JG. Delayed degradation in a resorbable plating system. J Craniofac Surg 2006;17:194-7. https://doi.org/10.1097/01.scs.0000194167.50546.7e
  7. Katsuragi YT, Gomi A, Sunaga A, et al. Intracerebral foreign body granuloma caused by a resorbable plate with passive intraosseous translocation after cranioplasty. J Neurosurg Pediatr 2013;12:622-5. https://doi.org/10.3171/2013.9.PEDS13187
  8. Xue AS, Koshy JC, Weathers WM, et al. Local foreign-body reaction to commercial biodegradable implants: an in vivo animal study. Craniomaxillofac Trauma Reconstr 2014;7:27-34. https://doi.org/10.1055/s-0033-1364199
  9. Bell RB, Kindsfater CS. The use of biodegradable plates and screws to stabilize facial fractures. J Oral Maxillofac Surg 2006;64:31-9.
  10. Pietrzak WS, Sarver DR, Verstynen ML. Bioabsorbable polymer science for the practicing surgeon. J Craniofac Surg 1997;8:87-91. https://doi.org/10.1097/00001665-199703000-00004
  11. Eppley BL, Morales L, Wood R, et al. Resorbable PLLA-PGA plate and screw fixation in pediatric craniofacial surgery: clinical experience in 1883 patients. Plast Reconstr Surg 2004;114:850-6. https://doi.org/10.1097/01.PRS.0000132856.69391.43
  12. Kwon H, Kim SW, Jung SN, et al. Cellulitis related to bioabsorbable plate and screws in infraorbital rim fracture. J Craniofac Surg 2011;22:625-7. https://doi.org/10.1097/SCS.0b013e3182085541
  13. Kumar AV, Staffenberg DA, Petronio JA, et al. Bioabsorbable plates and screws in pediatric craniofacial surgery: a review of 22 cases. J Craniofac Surg 1997;8:97-9. https://doi.org/10.1097/00001665-199703000-00006
  14. Bergsma EJ, Rozema FR, Bos RR, et al. Foreign body reactions to resorbable poly(L-lactide) bone plates and screws used for the fixation of unstable zygomatic fractures. J Oral Maxillofac Surg 1993;51:666-70. https://doi.org/10.1016/S0278-2391(10)80267-8
  15. Miller RA, Brady JM, Cutright DE. Degradation rates of oral resorbable implants (polylactates and polyglycolates): rate modification with changes in PLA/PGA copolymer ratios. J Biomed Mater Res 1977;11:711-9. https://doi.org/10.1002/jbm.820110507
  16. Laughlin RM, Block MS, Wilk R, et al. Resorbable plates for the fixation of mandibular fractures: a prospective study. J Oral Maxillofac Surg 2007;65:89-96.
  17. Rha EY, Paik H, Byeon JH. Bioabsorbable plates and screws fixation in mandible fractures: clinical retrospective research during a 10-year period. Ann Plast Surg 2015;74:432-6. https://doi.org/10.1097/SAP.0b013e3182a60693
  18. Sevastjanova NA, Mansurova LA, Dombrovska LE, et al. Biochemical characterization of connective tissue reaction to synthetic polymer implants. Biomaterials 1987;8:242-7. https://doi.org/10.1016/0142-9612(87)90110-4

Cited by

  1. Inflammatory foreign body reaction caused by resorbable materials used for orbital fractures repair : A case report vol.96, pp.49, 2016, https://doi.org/10.1097/md.0000000000008822
  2. Comparison of the long-term clinical performance of a biodegradable and a titanium fixation system in maxillofacial surgery: A multicenter randomized controlled trial vol.12, pp.5, 2017, https://doi.org/10.1371/journal.pone.0177152
  3. Aesthetic treatment of frontal sinus fractures and their complications vol.61, pp.12, 2016, https://doi.org/10.5124/jkma.2018.61.12.732
  4. Delayed formation of sterile abscess after zygomaticomaxillary complex fracture treatment with bioabsorbable plates vol.19, pp.2, 2016, https://doi.org/10.7181/acfs.2018.00038
  5. Surface‐modified halloysite nanotubes reinforced poly(lactic acid) for use in biodegradable coronary stents vol.135, pp.30, 2016, https://doi.org/10.1002/app.46521
  6. Design of biodegradable, implantable devices towards clinical translation vol.5, pp.1, 2020, https://doi.org/10.1038/s41578-019-0150-z
  7. A Novel Screw Drive for Allogenic Headless Position Screws for Use in Osteosynthesis-A Finite-Element Analysis vol.8, pp.10, 2021, https://doi.org/10.3390/bioengineering8100136