The study of Breast Specific Gamma Imaging Protocol using Self-development Phantom

자체 제작된 팬텀을 적용한 Breast Specific Gamma Imaging 검사 프로토콜에 대한 고찰

  • Lee, Hae-Jung (Department of Nuclear Medicine, Konkuk University Medical Center) ;
  • Lee, Juyoung (Graduate school of Public Health, Yonsei University) ;
  • Lim, Kuen-Kyo (Department of Nuclear Medicine, Konkuk University Medical Center) ;
  • Park, Hoon-Hee (Department of Radiological Technology, Shingu College)
  • 이해정 (건국대학교병원 핵의학과) ;
  • 이주영 (연세대학교 보건대학원) ;
  • 임근교 (건국대학교병원 핵의학과) ;
  • 박훈희 (신구대학교 방사선과)
  • Received : 2014.09.23
  • Accepted : 2014.10.22
  • Published : 2014.11.08

Abstract

Purpose As breast cancer patients continue to increase every year, cases of BSGI are on the rise with a heavier reliance on it. However, BSGI protocol in hospitals was not studied enough despite it was covered by hospital's condition and recommendation of manufacturers. The objective of the study was an examination of methods to be applicable to BSGI protocols, putting the self-development phantom to use in quality assessment of the images. Materials and Methods Dilon 6800 (Dilon Technologies Inc, Newport News, USA) was used in the study and five different sizes of sphere were distinctively produced in the phantom. The study used $^{99m}TcO_4$. The cases were classified in to three categories that background radioactivity to region of interest as ratio of 2: 4: 8, They were acquired images for 5, 7, 10mins. The acquired image was set region of interest according to the size of sphere, and We analyzed quantitative and qualitative analysis. The acquired data statistically analyzed with SPSS ver.18.0. Results As the result of quantitative and qualitative analysis, count rate of each sphere in accordance with difference of injection dose showed that higher count rate as injection dose and sphere size increased (P<0.005). Count rate of each sphere in accordance with difference of acquisition time showed that higher count rate as acquisition time and sphere size increased (P<0.005). Contrast noise ratio of each sphere in accordance with difference of injection dose showed that higher contrast noise ratio as injection dose increased. Particularly, Contrast noise ratio of eight times ratio images was the highest among. Contrast noise ratio of each sphere in accordance with difference of acquisition time showed that higher contrast noise ratio as acquisition time increased. And, Contrast noise ratio of seven minute image was the highest among (P<0.005). Conclusion There was significant change of Contrast noise ratio through quantitative and qualitative analysis. Moreover, We found usefulness of phantom. If Institutions identified image through the phantom study and they made BSGI protocol, We expected to help the improvement of diagnostic value of the images.

유방암 환자가 매년 증가함에 따라 BSGI 검사에 대한 의존도가 높아지고 검사 건수가 증가하고 있는 추세이다. 하지만 국내에서 BSGI 검사는 각 기관의 상황 및 업체권고에 따라 검사 프로토콜을 적용하고 있으며 관련 연구가 미비한 실정이다. 그러므로 본 연구에서는 영상의 질을 평가할 수 있는 팬텀을 자체적으로 제작하여 검사 프로토콜에 대해 고찰하고자 하였다. Dilon 6800 BSGI 장비를 이용하여 팬텀 내 구의 크기는 5단계로 구분하여 제작하여 실험하였다. 방사성동위원소는 $^{99m}TcO_4$를 사용하여 배후방사능과 관심영역의 비율을 2 : 4 : 8로 구분하였으며 5, 7, 10분 영상을 획득하였다. 획득된 영상은 각 구의 크기에 따라 관심영역을 설정하고 정량적, 정성적 분석을 시행하였다. 획득된 데이터는 SPSS ver.18.0을 통해 통계 분석하였다. 정량적, 정성적 분석 결과, 방사성동위원소의 주입량에 따른 각 구의 계수율은 주입량이 많을수록 계수율이 증가하였고, 구의 크기가 클수록 계수율이 증가하는 성향을 나타냈다(p<0.005). 영상획득시간에 따른 각 구의 계수율은 획득시간이 길수록 계수율이 증가하였고, 구의 크기가 클수록 증가하는 성향을 나타냈다(p<0.005). 방사성동위원소의 주입량에 따른 각 구의 대조도 잡음비는 주입량이 많을수록 증가하는 성향을 나타냈고, 8배 비율의 획득영상에서 대조도 잡음비가 가장 높았다. 또한 영상획득시간에 따른 각 구의 대조도 잡음비는 획득시간이 길수록 증가하는 성향을 나타냈고, 7분 획득영상에서 대조도 잡음비가 가장 높았다(p<0.005). 자체적으로 제작한 팬텀을 통한 정량적, 정성적 평가에서 통계적으로 유의한 차이가 있었으며, 팬텀의 유용성을 확인할 수 있었다. 또한 본 연구를 통해 장비 특성과 환경을 고려하여 BSGI 검사 프로토콜을 확인하고 활용한다면 보다 임상적 가치가 높을 것으로 사료된다.