Relationship Between Insertion Torque, and Pullout Strength Depending on the Size of the Pilot Hole and Biodegradable Suture Anchor in Osteoporotic Humeral Head

골다공증이 있는 상완골 골두의 파일럿 홀 (Pilot Hole)과 흡수성 나사못의 크기에 따른 토크 (Torque)와 뽑힘 강도 (Pullout Strength) 간의 관계

  • Chun, Yong-Min (The Department of Orthopaedic Surgery, Arthroscopy and Joint Research Institute, Severance Hospital, Yonsei University College of Medicine) ;
  • Lee, Young-Han (The Department of Radiology, Severance Hospital, Yonsei University College of Medicine) ;
  • Kim, Sung-Hwan (The Department of Orthopaedic Surgery, Arthroscopy and Joint Research Institute, Severance Hospital, Yonsei University College of Medicine) ;
  • Park, Yoo-Jung (The Department of Orthopaedic Surgery, Arthroscopy and Joint Research Institute, Severance Hospital, Yonsei University College of Medicine) ;
  • Kim, Sung-Jae (The Department of Orthopaedic Surgery, Arthroscopy and Joint Research Institute, Severance Hospital, Yonsei University College of Medicine)
  • 천용민 (연세대학교 의과대학 정형외과학교실) ;
  • 이영한 (연세대학교 의과대학 영상의학과학교실) ;
  • 김성환 (연세대학교 의과대학 정형외과학교실) ;
  • 박유정 (연세대학교 의과대학 정형외과학교실) ;
  • 김성재 (연세대학교 의과대학 정형외과학교실)
  • Received : 2012.02.27
  • Accepted : 2012.06.06
  • Published : 2012.06.30

Abstract

Purpose: The object of this study was to investigate the difference in torque and pullout strength between the standard anchor insertion (5.0 mm) with a small awl (3.7 mm) and larger anchor insertion (6.5 mm), with a standard awl (5.0 mm) in osteoporotic humeral head. Materials and Methods: The embalmed 24 paired cadaveric shoulders were assigned to either Group A or B. After measuring the bone mineral density (BMD) of the ROI (region of interest) in the humeral head, 5.0 mm suture anchors were inserted using a 3.7 mm awl in Group A1, and the same 5.0 mm anchors were inserted using a 5.0 mm awl in Group A2. The 5.0 mm anchors were inserted using a 5.0 mm awl in Group B1, and 6.5 mm anchors were inserted using a 5.0 mm awl in Group B2. We measured the torques at the time of the anchor insertion and pullout strengths. Results: There was no significant difference in the BMD between the groups. The torque of A1 (20.6 $cN{\cdot}m$) was significantly higher than that of A2 (13.2 $cN{\cdot}m$), and the torque of B2 (20.8 $cN{\cdot}m$) was significantly higher than that of B1(12.1 $cN{\cdot}m$). However, the difference in the increased torque between group A and B was not significant. The pullout strength of A1 (204.2 N) was significantly higher than that of A2 (152.9 N), and the pullout strength of B2 (210.9 N) was significantly higher than that of B1 (149.5 N). However, the difference in the increased pullout strength between Group A and B was not significant. Conclusion: In severe osteoporosis, the use of a larger suture anchor with a standard awl increased the torque and pullout strength significantly, in comparison to the use of the same sized suture anchor and awl. If there is an inadequate interval between the anchors on the greater tuberosity, the use of a 3.7 mm awl and 5.0 mm anchor will be beneficial compared to that of a 5.0 mm awl and 6.5 mm anchor, considering that an increase in the pullout strength does not depend on the awl size.

목적: 본 연구의 목적은 골다공증이 있는 회전근개 파열 환자에서 상완골 골두에 표준 크기 (5.0 mm) 보다 작은 송곳 (awl) (3.7 mm 나사못용)을 사용해서 5.0 mm 나사못을 삽입하는 것이 표준 크기의 송곳과 6.5 mm 나사못을 삽입하는 것에 비해 토크 (torque)나 뽑힘 강도(pullout strength)에서 차이가 있는지를 알아보고자 하였다. 대상 및 방법: 방부 처리된 12명, 24개의 짝지은 사체 견관절을 A군와 B군, 두 군으로 나누어 관심 부위의 골밀도를 측정한 후에 A군에서는 5.0 mm 나사못 삽입을 위해 3.7 mm 송곳 (awl)을 이용하는 경우를 A1군, 5.0 mm 나사못 삽입을 위해 5.0 mm 송곳을 이용하는 경우를 A2군으로 하였다. B군에서는 5.0 mm 송곳을 이용하여 5.0 mm 나사못을 삽입하는 경우를 B1군, 6.5 mm 나사못을 삽입하는 경우를 B2군으로 하여 나사못 삽입 시의 토크와 뽑힘 강도를 측정하였다. 결과: 골밀도는 A군과 B군 간이나 군내에 차이가 없었다. 토크는 A1군이 A2군보다 각각 20.6 $cN{\cdot}m$와 13.2 $cN{\cdot}m$로 유의한 차이를 보였고 (p<0.001), B1군과 B2군에서도 각각 12.1 $cN{\cdot}m$와 20.8 $cN{\cdot}m$로 유의한 차이를 보였다 (p<0.001). 그러나 증가된 정도의 차이는 유의하지 않았다. A1군과 A2군의 뽑힘 강도는 각각 204.2 N과 152.9 N으로 유의한 차이를 보였고 (p<0.001), B1군과 B2군에서도 149.5 N과 210.9 N으로 유의한 차이를 보였다 (p<0.001). 그러나 증가된 정도의 차이는 유의하지 않았다. 결론: 송곳 크기보다 큰 치수의 나사못을 사용하면, 같은 크기의 송곳과 나사못을 사용하는 경우 토크나 뽑힘 강도가 유의하게 증가하였다. 그러나 증가된 정도는 큰 송곳과 작은 송곳 간에 차이가 유의하지 않았다. 따라서 대결절의 footprint 내에 나사못과 나사못 사이의 거리가 충분하지 않은 경우에 5.0 mm 송곳을 이용하여 6.5 mm 나사못을 삽입하는 대신 3.7 mm 송곳을 이용하여 5.0 mm 나사못을 삽입해도 증가되는 뽑힘 강도는 차이가 없을 것으로 생각한다.

Keywords

References

  1. Lehman C, Cuomo F, Kummer FJ, Zuckerman JD. The incidence of full thickness rotator cuff tears in a large cadaveric population. Bull Hosp Jt Dis. 1995;54: 30-1.
  2. Milgrom C, Schaffler M, Gilbert S, van Holsbeeck M. Rotator-cuff changes in asymptomatic adults. The effect of age, hand dominance and gender. J Bone Joint Surg Br. 1995;77:296-8.
  3. Sher JS, Uribe JW, Posada A, Murphy BJ, Zlatkin MB. Abnormal findings on magnetic resonance images of asymptomatic shoulders. J Bone Joint Surg Am. 1995;77:10-5. https://doi.org/10.2106/00004623-199501000-00002
  4. Barber FA, Feder SM, Burkhart SS, Ahrens J. The relationship of suture anchor failure and bone density to proximal humerus location: a cadaveric study. Arthroscopy. 1997;13:340-5. https://doi.org/10.1016/S0749-8063(97)90031-1
  5. Tingart MJ, Apreleva M, Zurakowski D, Warner JJ. Pullout strength of suture anchors used in rotator cuff repair. J Bone Joint Surg Am. 2003;85-A:2190-8.
  6. Tingart MJ, Apreleva M, Lehtinen J, Zurakowski D, Warner JJ. Anchor design and bone mineral density affect the pull-out strength of suture anchors in rotator cuff repair: which anchors are best to use in patients with low bone quality? Am J Sports Med. 2004;32: 1466-73. https://doi.org/10.1177/0363546503262644
  7. Kaar TK, Schenck RC, Jr., Wirth MA, Rockwood CA, Jr. Complications of metallic suture anchors in shoulder surgery: A report of 8 cases. Arthroscopy. 2001;17:31-7. https://doi.org/10.1053/jars.2001.18246
  8. Benson EC, MacDermid JC, Drosdowech DS, Athwal GS. The incidence of early metallic suture anchor pullout after arthroscopic rotator cuff repair. Arthroscopy. 2010;26:310-5. https://doi.org/10.1016/j.arthro.2009.08.015
  9. Kirchhoff C, Braunstein V, Milz S, Sprecher CM, Fischer F, Tami A, Ahrens P, Imhoff AB, Hinter-wimmer S. Assessment of bone quality within the tuberosities of the osteoporotic humeral head: relevance for anchor positioning in rotator cuff repair. Am J Sports Med. 2010;38:564-9. https://doi.org/10.1177/0363546509354989
  10. Mahar A, Allred DW, Wedemeyer M, Abbi G, Pedowitz R. A biomechanical and radiographic analysis of standard and intracortical suture anchors for arthroscopic rotator cuff repair. Arthroscopy. 2006;22: 130-5. https://doi.org/10.1016/j.arthro.2005.08.042
  11. Schneeberger AG, von Roll A, Kalberer F, Jacob HA, Gerber C. Mechanical strength of arthroscopic rotator cuff repair techniques: an in vitro study. J Bone Joint Surg Am. 2002;84-A:2152-60.
  12. Burkhart SS. The deadman theory of suture anchors: observations along a south Texas fence line. Arthroscopy. 1995;11:119-23. https://doi.org/10.1016/0749-8063(95)90100-0
  13. Petit CJ, Boswell R, Mahar A, Tasto J, Pedowitz RA. Biomechanical evaluation of a new technique for rotator cuff repair. Am J Sports Med. 2003;31:849-53. https://doi.org/10.1177/03635465030310062001
  14. Lee S, Mahar A, Bynum K, Pedowitz R. Biomechanical comparison of bioabsorbable sutureless screw anchor versus suture anchor fixation for rotator cuff repair. Arthroscopy. 2005;21:43-7. https://doi.org/10.1016/j.arthro.2004.09.020
  15. Bynum CK, Lee S, Mahar A, Tasto J, Pedowitz R. Failure mode of suture anchors as a function of insertion depth. Am J Sports Med. 2005;33:1030-4. https://doi.org/10.1177/0363546504271746
  16. Gerber C, Fuchs B, Hodler J. The results of repair of massive tears of the rotator cuff. J Bone Joint Surg Am. 2000;82:505-15. https://doi.org/10.2106/00004623-200004000-00006
  17. Djurasovic M, Marra G, Arroyo JS, Pollock RG, Flatow EL, Bigliani LU. Revision rotator cuff repair: factors influencing results. J Bone Joint Surg Am. 2001;83-A:1849-55.
  18. Goradia VK, Mullen DJ, Boucher HR, Parks BG, O'Donnell JB. Cyclic loading of rotator cuff repairs: A comparison of bioabsorbable tacks with metal suture anchors and transosseous sutures. Arthroscopy. 2001;17:360-4. https://doi.org/10.1053/jars.2001.21243
  19. Burkhart SS, Diaz Pagan JL, Wirth MA, Athanasiou KA. Cyclic loading of anchor-based rotator cuff repairs: confirmation of the tension overload phenomenon and comparison of suture anchor fixation with transosseous fixation. Arthroscopy. 1997;13:720-4. https://doi.org/10.1016/S0749-8063(97)90006-2
  20. Yakacki CM, Poukalova M, Guldberg RE, Lin A, Saing M, Gillogly S, Gall K. The effect of the trabecular microstructure on the pullout strength of suture anchors. J Biomech. 2010;43:1953-9. https://doi.org/10.1016/j.jbiomech.2010.03.013