The Effects of a Thyroid Shield Made of a Tissue-Equivalent Material on the Reduction of the Thyroid Exposure Dose in Panoramic Radiography

파노라마촬영 시 조직등가물질을 이용한 갑상선보호대의 갑상선피폭선량 감소효과

  • Lee, Hye-Lim (Department of Dental Radiology, Yonsei University Dental Hospital) ;
  • Kim, Hyun-Yung (Department of Dental Radiology, Yonsei University Dental Hospital) ;
  • Choi, Hyung-Wook (Department of Dental Radiology, Yonsei University Dental Hospital) ;
  • Lee, Hye-Mi (Department of Diagnostic Radiology, Severance Hospital of Yonsei University Health System) ;
  • Lim, Chang-Seon (Department of Radiological Science, Konyang University)
  • 이혜림 (연세대학교치과대학병원 구강악안면방사선과) ;
  • 김현영 (연세대학교치과대학병원 구강악안면방사선과) ;
  • 최형욱 (연세대학교치과대학병원 구강악안면방사선과) ;
  • 이혜미 (세브란스병원 영상의학과) ;
  • 임창선 (건양대학교 방사선학과)
  • Received : 2012.02.10
  • Accepted : 2012.05.10
  • Published : 2012.05.31

Abstract

Exposure-dose reducing effect was measured by using bolus, a tissue-equivalent material as a shield to obtain useful diagnostic images while minimizing the radiation exposure of thyroid which is highly sensitive to radiation during panoramic radiography. The experiment was performed within the period of 1 June 2001 through 30 June 2011 by measuring entrance surface dose and deep dose at the thyroid-corresponding site of a head and neck phantom. As a result, the entrance surface dose in the thyroid for using no shield was 43.84 ${\mu}Gy$ on the average, and the thyroid shield of bolus 10 mm in thickness reduced the dose by 15.45 ${\mu}Gy$(35.24%) to 28.39 ${\mu}Gy$ on the average. The use of a 20 mm thyroid shield resulted in the dose of 25.38 ${\mu}Gy$ on the average, a 18.46 ${\mu}Gy$(42.10%) drop from 43.84 ${\mu}Gy$ for using no shield. On the site 20 mm below the surface, a thyroid shield 10 mm in thickness had no dose-reducing effect, while a 20 mm thyroid shield reduced the dose by 0.06 mSv(20%).

파노라마방사선촬영 시 방사선감수성이 높은 갑상선의 방사선 피폭을 최소화하면서 진단에 유용한 영상을 얻기 위하여 조직등가물질인 보루스(bolus)를 차폐체로 사용하여 피폭선량 감소효과를 측정하였다. 실험은 2011. 6. 1일부터 6. 30일까지 두경부팬톰을 이용하여 갑상선 위치의 표면입사선량과 심부흡수선량을 측정하였다. 그 결과 갑상선 부위 입사표면선량은 평균 43.84 ${\mu}Gy$이었고, 10 mm 두께의 보루스로 만든 갑상선보호대에서는 평균 28.39 ${\mu}Gy$로 15.45 ${\mu}Gy$(35.24%)가 감소되었다. 20 mm 갑상선보호대를 착용했을 경우에는 평균 25.38 ${\mu}Gy$로 미착용 시 평균 43.84 ${\mu}Gy$보다 18.46 ${\mu}Gy$(42.10%)가 감소되었다. 그리고 표면에서 20 mm 깊이에서는 10 mm 두께의 갑상선보호대는 선량 감소효과가 없었고, 20 mm 갑상선보호대를 착용했을 경우에는 0.06 mSv (20%)의 선량감소효과가 있었다.

Keywords

References

  1. S. S. Kim et al, "Radiographic Imaging", p.585, Daihaks publishing co., 2006.
  2. W. G. Chung et al., "Dental Radiography -Principles and Techniques", 3rd ed. p.269, DaehanNarae Publishing Inc., 2009.
  3. T. W. Park et al, "Oral and maxillofacial radiology". 3rd ed. p.138, Narae Publishing Inc., 2001.
  4. M. J. Shin, et al.,"Usefullness of panoramic radiograph for the improvement of periodic oral examination", Korean J Oral Maxillofac Radiol., Vol.40, pp.25-32, 2010.
  5. ICRP, 1990 Recommendations of the International Commission on Radiological Protection, Publication 60, Annals of the ICRP Vol.21, No. 1-3, Pergomon Press, Oxford, 1991.
  6. G. S. Lee et al, "Absorbed and effective dose in direct and indirect digital panoramic radiography", Korean J Oral Maxillofac Radiol, Vol.40, pp.9-14, 2010.
  7. Y. H. Kim, "Method for Measurement of Entrance Surface Dose and Role of Radiation technologist", Journal of Radiological Science and Technology, Vol.28, No.3, pp.173-191, September, 2005.
  8. C. H. Lee, and C. S. Lim, "A Study on Added Filters for Reduction of Radiation Exposure Dose in Skull A-P Projection", Journal of the Korea Academia-Industrial cooperation Society, Vol.12, No.7, pp.3117-3122, July, 2011. https://doi.org/10.5762/KAIS.2011.12.7.3117
  9. Y. H. Kim et al, "Patient exposure doses from medical X-ray examninations in Korea", Journal of Radiological Science and Technology, Vol.28, No.3, pp. 241-248, September, 2005.
  10. C. S. Lim and S. H. Kim, "A Study on the Radiation Dose Managements in the Nuclear Medicine Department", Journal of the Korea Academia-Industrial cooperation Society, Vol.10, No.7, pp.1760-1765, July, 2009. https://doi.org/10.5762/KAIS.2009.10.7.1760
  11. W. K. Choi et al. "Shielding Effect according to the Direction of Control Room Door Opening during Radiography", Journal of the Korea Academia-Industrial cooperation Society, Vol.11, No.9, pp.3347-3352, September, 2010. https://doi.org/10.5762/KAIS.2010.11.9.3347
  12. S. C. Han et al., "Evaluation of Effective Dose in Dental Radiography", Journal of Radiological Science and Technology, Vol.34, No.1, pp.27-33, march, 2011.
  13. D. C. Kweon et al., "A Literature Review and Analysis of Dosimetry in Panoramic Radiography", Journal of Radiological Science and Technology, Vol.33, No.1, pp.1-10, march, 2010.
  14. http://cancer.go.kr/nciapps/fcatalog/ecatalog.jsp?Dir=86
  15. K. R. Dong, "A Study on the effects of scattering dose on eyes and thyroid for panoramagraphy, pp.1118-1123, 2009 Spring International Disign Exhibition, The Korea Contents Association, 2009
  16. E. K. Kim et al., "Development of diagnostic reference level in dental x-ray examination in Korea", p.35, KFDA Rearch Project Report, 2009.
  17. Association of Korean Professors of Oral and Maxillofacial Radiology."Oral and maxillofacial radiology. 4th ed. pp.138-143, Narae Publishing, Inc. 2008.
  18. S. C. Choi, "Difference in radiation absorbed dose according to the panoramic radiographic machines, Korean J Oarl Maxiofacial Radiol, Vol.30, No.1, pp.11-15. 2000.
  19. G. Kaeppler et al., "Influence of the rotation centre in panoramic radiography", Radiat Prot Dosimetry, Vol.128, No.2, pp.239-244. 2008.
  20. H. D. Kang, "A research about guide line of image evaluation in dental diagnostic radiology", p.23, KFDA Rearch Project Report, 2009

Cited by

  1. A Study on the Patient Exposure Doses from the Panoramic Radiography using Dentistry vol.7, pp.1, 2013, https://doi.org/10.7742/jksr.2013.7.1.017
  2. A Study on the Thyroid Dose High-Energy Radiation Therapy of Lung Cancer vol.15, pp.6, 2015, https://doi.org/10.5392/JKCA.2015.15.06.297
  3. A study of beam hardening effect reduction occur in brain CT vol.16, pp.12, 2015, https://doi.org/10.5762/KAIS.2015.16.12.8479
  4. Development of radiation shield providing medical support for radioactive disaster in computer awareness analysis convergence-based pp.1573-7543, 2018, https://doi.org/10.1007/s10586-017-0750-6
  5. Comparative Study of the Effective Dose from Panoramic Radiography in Dentistry Measured Using a Radiophotoluminescent Glass Dosimeter and an Optically Stimulated Luminescence Detector vol.73, pp.9, 2018, https://doi.org/10.3938/jkps.73.1377