ProTaper Universal에 축적된 내부 응력이 피로 파절에 미치는 영향

Effect of Internal Stress on Cyclic Fatigue Failure in ProTaper Universal

  • 조윤성 (강릉원주대학교 치과대학 치의학과 치과보존학교실) ;
  • 김진우 (강릉원주대학교 치과대학 치의학과 치과보존학교실) ;
  • 조경모 (강릉원주대학교 치과대학 치의학과 치과보존학교실) ;
  • 박세희 (강릉원주대학교 치과대학 치의학과 치과보존학교실)
  • Jo, Yun-Sung (Department of Conservative Dentistry, Gangneung-Wonju National University, School of Dentistry) ;
  • Kim, Jin-Woo (Department of Conservative Dentistry, Gangneung-Wonju National University, School of Dentistry) ;
  • Cho, Kyung-Mo (Department of Conservative Dentistry, Gangneung-Wonju National University, School of Dentistry) ;
  • Park, Se-Hee (Department of Conservative Dentistry, Gangneung-Wonju National University, School of Dentistry)
  • 투고 : 2011.12.23
  • 심사 : 2012.03.25
  • 발행 : 2012.03.30

초록

본 연구는 ProTaper Universal에 의도적으로 부여한 내부 응력이 피로 파절에 미치는 영향을 평가하였다. 25 mm 길이의 S1, S2, F1, F2, F3, F4 그리고 F5 파일을 내부 응력을 부여하기 위해서 Endo-Training-Bloc (Dentsply Maillefer)에 auto-stop 될 때까지 속도 300 rpm, 토크 1.0 Ncm의 값으로 밀어 넣었다. 부여한 내부 응력의 횟수에 따라 각각 Stress 0 군(대조군), Stress 1 군, Stress 2 군 그리고 Stress 3 군으로 분류하였다. 경사진 유리판을 사용하였고 파일 파절 시간을 측정하였다. 모든 파일에서는 Stress 3 군은 대조군에 비해 Ni-Ti 파일의 파절 시간이 감소하였고 F2와 F3에서 모든 실험군은 대조군에 비해 파절 시간이 감소하였다. F4와 F5에서 Stress 2 군과 Stress 3 군이 대조군에 비해 파절 시간이 감소하였다. ProTaper Universal의 피로 파절은 파일에 축적된 내부 응력에 의해 영향을 받는다.

The purpose of this study was to evaluate the relation between intentionally induced internal stress and cyclic fatigue failure of ProTaper Universal. ProTaper Universal (Dentsply Maillefer) S1, S2, F1, F2, F3, F4 and F5 (25 mm length) were used in this study. To give the internal stress, the ProTaper Universal were put into the .02 taper Endo-Training-Bloc (Dentsply Maillefer) until auto-stop by torque controlled motor. The rotation speed was 300 rpm and torque value was 1.0 Ncm. ProTaper Universal were grouped by the induced number of internal stress and randomly distributed among one control group and three experimental groups (n=10). The four groups were Stress 0 (control), Stress 1, Stress 2 and Stress 3. These instruments were rotated until separation. For cyclic fatigue measurement, inclined plane was used and time for separation was recorded. It was statistically analyzed using two-way ANOVA and Duncan post-hoc test at 95% confidential level. In all ProTaper Universal, there was statistically significant decrease on time for separation in Stress 3. In F2 and F3, there were statistically significant difference between control group and all experimental groups. And in F4 and F5, Stress 2 and 3 groups showed significantly lower cyclic fatigue resistance from Stress 0 group. In the limitation of this study, cyclic fatigue failure of ProTaper Universal is influenced by accumulated internal stress.

키워드

참고문헌

  1. Walia HM, Brantley WA, Gerstein H. An initial investigation of the bending and torsional properties of Nitinol root canal files. J Endod 1988;14:346-351. https://doi.org/10.1016/S0099-2399(88)80196-1
  2. Mize SB, Clement DJ, Pruett JP, Carnes DL, Jr. Effect of sterilization on cyclic fatigue of rotary nickel-titanium endodontic instruments. J Endod 1998;24:843-847. https://doi.org/10.1016/S0099-2399(98)80015-0
  3. Ayar LR, Love RM. Shaping ability of ProFile and K3 rotary Ni-Ti instruments when used in a variable tip sequence in simulated curved root canals. Int Endod J 2004;37:593-601. https://doi.org/10.1111/j.1365-2591.2004.00851.x
  4. Al-Sudani D, Al-Shahrani S. A comparison of the canal centering ability of ProFile, K3, and RaCe Nickel Titanium rotary systems. J Endod 2006;32:1198-1201. https://doi.org/10.1016/j.joen.2006.07.017
  5. Gambill JM, Alder M, del Rio CE. Comparison of nickel-titanium and stainless steel hand-file instrumentation using computed tomography. J Endod 1996;22:369-375. https://doi.org/10.1016/S0099-2399(96)80221-4
  6. Pruett JP, Clement DJ, Carnes DL, Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod 1997;23:77-85. https://doi.org/10.1016/S0099-2399(97)80250-6
  7. Berutti E, Negro AR, Lendini M, Pasqualini D. Influence of manual preflaring and torque on the failure rate of ProTaper rotary instruments. J Endod 2004;30:228-230. https://doi.org/10.1097/00004770-200404000-00011
  8. Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod 2000;26:161-165. https://doi.org/10.1097/00004770-200003000-00008
  9. Ullmann CJ, Peters OA. Effect of cyclic fatigue on static fracture loads in ProTaper nickel-titanium rotary instruments. J Endod 2005;31:183-186. https://doi.org/10.1097/01.don.0000137641.87125.8f
  10. Peters OA. Current challenges and concepts in the preparation of root canal systems: a review. J Endod 2004;30:559-567. https://doi.org/10.1097/01.DON.0000129039.59003.9D
  11. Parashos P, Gordon I, Messer HH. Factors influencing defects of rotary nickel-titanium endodontic instruments after clinical use. J Endod 2004;30:722-725. https://doi.org/10.1097/01.DON.0000129963.42882.C9
  12. Park SH, Seo HW, CU. H. An evaluation of rotational stability in endodontic electronic motors. J Kor Acad Cons Dent 2010;35:246-256. https://doi.org/10.5395/JKACD.2010.35.4.246
  13. Necchi S, Petrini L, Taschieri S, Migliavacca F. A comparative computational analysis of the mechanical behavior of two nickel-titanium rotary endodontic instruments. J Endod 2010;36:1380-1384. https://doi.org/10.1016/j.joen.2010.03.026
  14. Bahia MG, Martins RC, Gonzalez BM, Buono VT. Physical and mechanical characterization and the influence of cyclic loading on the behaviour of nickel-titanium wires employed in the manufacture of rotary endodontic instruments. Int Endod J 2005;38: 795-801. https://doi.org/10.1111/j.1365-2591.2005.01016.x
  15. Li UM, Lee BS, Shih CT, Lan WH, Lin CP. Cyclic fatigue of endodontic nickel titanium rotary instruments: static and dynamic tests. J Endod 2002;28:448-451. https://doi.org/10.1097/00004770-200206000-00007
  16. Schneider SW. A comparison of canal preparations in straight and curved root canals. Oral Surg Oral Med Oral Pathol 1971;32:271-275. https://doi.org/10.1016/0030-4220(71)90230-1
  17. Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod 2009;35:1469-1476. https://doi.org/10.1016/j.joen.2009.06.015
  18. Parashos P, Messer HH. Rotary NiTi instrument fracture and its consequences. J Endod 2006;32:1031-1043. https://doi.org/10.1016/j.joen.2006.06.008
  19. Grande NM, Plotino G, Pecci R, Bedini R, Malagnino VA, Somma F. Cyclic fatigue resistance and three-dimensional analysis of instruments from two nickel-titanium rotary systems. Int Endod J 2006;39:755-763. https://doi.org/10.1111/j.1365-2591.2006.01143.x
  20. Fife D, Gambarini G, Britto Lr L. Cyclic fatigue testing of ProTaper NiTi rotary instruments after clinical use. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2004;97:251-256. https://doi.org/10.1016/j.tripleo.2003.08.010
  21. Zelada G, Varela P, Martin B, Bahillo JG, Magan F, Ahn S. The effect of rotational speed and the curvature of root canals on the breakage of rotary endodontic instruments. J Endod 2002;28:540-542. https://doi.org/10.1097/00004770-200207000-00014
  22. Haikel Y, Serfaty R, Bateman G, Senger B, Allemann C. Dynamic and cyclic fatigue of engine-driven rotary nickel-titanium endodontic instruments. J Endod 1999;25:434-440. https://doi.org/10.1016/S0099-2399(99)80274-X
  23. Kim JG, Kum KY, ES. K. Comparative study on morphology of cross-section and cyclic fatigue test with different rotary NiTi files and handling methods. J Kor Acad Cons Dent 2006;31:96-102. https://doi.org/10.5395/JKACD.2006.31.2.096
  24. Hwang SY, Oh SR, Lee Y, Lim SM, KY. K. Effect of cross-sectional area of 6 nickel-titanium rotary instruments on the fatigue fracture under cyclic flexural stress: a fractographic analysis. J Kor Acad Cons Dent 2009;34.
  25. Kim HC. Mechanical and geometric features of endodontic instruments and its clinical effect. J Kor Acad Cons Dent 2011;36:1-11. https://doi.org/10.5395/JKACD.2011.36.1.1
  26. Wolcott S, Wolcott J, Ishley D, Kennedy W, Johnson S, Minnich S, Meyers J. Separation incidence of protaper rotary instruments: a large cohort clinical evaluation. J Endod 2006;32:1139-1141. https://doi.org/10.1016/j.joen.2006.05.015
  27. Ounsi HF, Salameh Z, Al-Shalan T, Ferrari M, Grandini S, Pashley DH, Tay FR. Effect of clinical use on the cyclic fatigue resistance of ProTaper nickel-titanium rotary instruments. J Endod 2007;33:737-741. https://doi.org/10.1016/j.joen.2007.03.006
  28. Yum J, Cheung GS, Park JK, Hur B, Kim HC. Torsional strength and toughness of nickel-titanium rotary files. J Endod 2011;37:382-386. https://doi.org/10.1016/j.joen.2010.11.028