DOI QR코드

DOI QR Code

Effect of Bone Cement Volume and Stiffness on Occurrences of Adjacent Vertebral Fractures after Vertebroplasty

  • Kim, Jin-Myung (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology) ;
  • Shin, Dong Ah (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology) ;
  • Byun, Dong-Hak (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology) ;
  • Kim, Hyung-Sun (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology) ;
  • Kim, Sohee (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology) ;
  • Kim, Hyoung-Ihl (Department of Medical System Engineering and Mechatronics, Gwangju Institute of Science and Technology)
  • Received : 2012.04.04
  • Accepted : 2012.11.22
  • Published : 2012.11.28

Abstract

Objective : The purpose of this study is to find the optimal stiffness and volume of bone cement and their biomechanical effects on the adjacent vertebrae to determine a better strategy for conducting vertebroplasty. Methods : A three-dimensional finite-element model of a functional spinal unit was developed using computed tomography scans of a normal motion segment, comprising the T11, T12 and L1 vertebrae. Volumes of bone cement, with appropriate mechanical properties, were inserted into the trabecular core of the T12 vertebra. Parametric studies were done by varying the volume and stiffness of the bone cement. Results : When the bone cement filling volume reached 30% of the volume of a vertebral body, the level of stiffness was restored to that of normal bone, and when higher bone cement exceeded 30% of the volume, the result was stiffness in excess of that of normal bone. When the bone cement volume was varied, local stress in the bony structures (cortical shell, trabecular bone and endplate) of each vertebra monotonically increased. Low-modulus bone cement has the effect of reducing strain in the augmented body, but only in cases of relatively high volumes of bone cement (>50%). Furthermore, varying the stiffness of bone cement has a negligible effect on the stress distribution of vertebral bodies. Conclusion : The volume of cement was considered to be the most important determinant in endplate fracture. Changing the stiffness of bone cement has a negligible effect on the stress distribution of vertebral bodies.

Keywords

References

  1. Adams MA, Bogduk N, Burton K, Dolan P : The Biomechanics of Back Pain. Edinburgh, London, New York : Churchill Livingstone, 2002, pp100
  2. Anselmetti GC, Manca A, Hirsch J, Montemurro F, Isaia G, Osella G, et al. : Percutaneous vertebroplasty in osteoporotic patients : an institutional experience of 1,634 patients with long-term follow-up. J Vasc Interv Radiol 22 : 1714-1720, 2011 https://doi.org/10.1016/j.jvir.2011.09.001
  3. Aquarius R, Homminga J, Verdonschot N, Tanck E : The fracture risk of adjacent vertebrae is increased by the changed loading direction after a wedge fracture. Spine (Phila Pa 1976) 36 : E408-E412, 2011 https://doi.org/10.1097/BRS.0b013e3181f0f726
  4. Bae H, Hatten HP Jr, Linovitz R, Tahernia AD, Schaufele MK, McCollom V, et al. : A prospective randomized FDA-IDE trial comparing Cortoss with PMMA for vertebroplasty : a comparative effectiveness research study with 24-month follow-up. Spine (Phila Pa 1976) 37 : 544-550, 2012 https://doi.org/10.1097/BRS.0b013e31822ba50b
  5. Baroud G, Heini P, Nemes J, Bohner M, Ferguson S, Steffen T : Biomechanical explanation of adjacent fractures following vertebroplasty. Radiology 229 : 606-607; author reply 607-608, 2003 https://doi.org/10.1148/radiol.2292030378
  6. Baroud G, Nemes J, Heini P, Steffen T : Load shift of the intervertebral disc after a vertebroplasty : a finite-element study. Eur Spine J 12 : 421-426, 2003 https://doi.org/10.1007/s00586-002-0512-9
  7. Belkoff SM, Mathis JM, Erbe EM, Fenton DC : Biomechanical evaluation of a new bone cement for use in vertebroplasty. Spine (Phila Pa 1976) 25 : 1061-1064, 2000 https://doi.org/10.1097/00007632-200005010-00004
  8. Berlemann U, Ferguson SJ, Nolte LP, Heini PF : Adjacent vertebral failure after vertebroplasty. A biomechanical investigation. J Bone Joint Surg Br 84 : 748-752, 2002 https://doi.org/10.1302/0301-620X.84B5.11841
  9. Boger A, Heini P, Windolf M, Schneider E : Adjacent vertebral failure after vertebroplasty : a biomechanical study of low-modulus PMMA cement. Eur Spine J 16 : 2118-2125, 2007 https://doi.org/10.1007/s00586-007-0473-0
  10. Boonen S, Wahl DA, Nauroy L, Brandi ML, Bouxsein ML, Goldhahn J, et al. : Balloon kyphoplasty and vertebroplasty in the management of vertebral compression fractures. Osteoporos Int 22 : 2915-2934, 2011 https://doi.org/10.1007/s00198-011-1639-5
  11. Hierholzer J, Fuchs H, Westphalen K, Baumann C, Slotosch C, Schulz R : Incidence of symptomatic vertebral fractures in patients after percutaneous vertebroplasty. Cardiovasc Intervent Radiol 31 : 1178-1183, 2008 https://doi.org/10.1007/s00270-008-9376-7
  12. Kallmes DF, Comstock BA, Heagerty PJ, Turner JA, Wilson DJ, Diamond TH, et al. : A randomized trial of vertebroplasty for osteoporotic spinal fractures. N Engl J Med 361 : 569-579, 2009 https://doi.org/10.1056/NEJMoa0900563
  13. Liebschner MA, Rosenberg WS, Keaveny TM : Effects of bone cement volume and distribution on vertebral stiffness after vertebroplasty. Spine (Phila Pa 1976) 26 : 1547-1554, 2001 https://doi.org/10.1097/00007632-200107150-00009
  14. McArthur N, Kasperk C, Baier M, Tanner M, Gritzbach B, Schoierer O,et al. : 1150 kyphoplasties over 7 years : indications, techniques, and intraoperative complications. Orthopedics 32 : 90, 2009
  15. Pérez-Higueras A, Alvarez L, Rossi RE, Quinones D, Al-Assir I : Percutaneous vertebroplasty : long-term clinical and radiological outcome. Neuroradiology 44 : 950-954, 2002 https://doi.org/10.1007/s00234-002-0856-1
  16. Polikeit A, Nolte LP, Ferguson SJ : The effect of cement augmentation on the load transfer in an osteoporotic functional spinal unit : finite-element analysis. Spine (Phila Pa 1976) 28 : 991-996, 2003
  17. Tanigawa N, Kariya S, Komemushi A, Tokuda T, Nakatani M, Yagi R, et al. : Cement leakage in percutaneous vertebroplasty for osteoporotic compression fractures with or without intravertebral clefts. AJR Am J Roentgenol 193 : W442-W445, 2009 https://doi.org/10.2214/AJR.09.2774
  18. Trout AT, Kallmes DF : Does vertebroplasty cause incident vertebral fractures? A review of available data. AJNR Am J Neuroradiol 27 : 1397-1403, 2006
  19. Trout AT, Kallmes DF, Kaufmann TJ : New fractures after vertebroplasty : adjacent fractures occur significantly sooner. AJNR Am J Neuroradiol 27 : 217-223, 2006
  20. Voormolen MH, Lohle PN, Lampmann LE, van den Wildenberg W, Juttmann JR, Diekerhof CH, et al. : Prospective clinical follow-up after percutaneous vertebroplasty in patients with painful osteoporotic vertebral compression fractures. J Vasc Interv Radiol 17 : 1313-1320, 2006 https://doi.org/10.1097/01.RVI.0000231952.75209.4A

Cited by

  1. Severe kyphotic deformity resulting from collapses of cemented and adjacent vertebrae following percutaneous vertebroplasty using calcium phosphate cement. A case report vol.43, pp.10, 2014, https://doi.org/10.1007/s00256-014-1912-8
  2. Pedicle screw fixation with kyphoplasty decreases the fracture risk of the treated and adjacent non-treated vertebral bodies: a finite element analysis vol.36, pp.6, 2012, https://doi.org/10.1007/s11596-016-1680-x
  3. Appropriate Cement Volume in Vertebroplasty: A Multivariate Analysis with Short-Term Follow-Up vol.12, pp.2, 2012, https://doi.org/10.13004/kjnt.2016.12.2.128
  4. Modification of PMMA vertebroplasty cement for reduced stiffness by addition of normal saline: a material properties evaluation vol.26, pp.12, 2012, https://doi.org/10.1007/s00586-016-4845-1
  5. Bone cement distribution in the vertebral body affects chances of recompression after percutaneous vertebroplasty treatment in elderly patients with osteoporotic vertebral compression fractures vol.12, pp.None, 2012, https://doi.org/10.2147/cia.s113240
  6. Optimizing bone cement stiffness for vertebroplasty through biomechanical effects analysis based on patient-specific three-dimensional finite element modeling vol.56, pp.11, 2012, https://doi.org/10.1007/s11517-018-1844-x
  7. BIOMECHANICAL EVALUATION OF LOW-MODULUS BONE CEMENT FOR ENHANCING APPLICABILITY IN VERTEBROPLASTY - AN EXPERIMENTAL STUDY IN PORCINE MODEL vol.30, pp.1, 2018, https://doi.org/10.4015/s1016237218500023
  8. The effects of vertebral morphology on human lumbar spine with progressive decrease in bone mineral density - An FEM study vol.402, pp.None, 2012, https://doi.org/10.1088/1757-899x/402/1/012091
  9. Biomechanical effects of different vertebral heights after augmentation of osteoporotic vertebral compression fracture: a three-dimensional finite element analysis vol.13, pp.None, 2012, https://doi.org/10.1186/s13018-018-0733-1
  10. Cement injection and postoperative vertebral fractures during vertebroplasty vol.14, pp.None, 2019, https://doi.org/10.1186/s13018-019-1273-z
  11. Application of percutaneous vertebroplasty and percutaneous kyphoplasty in treating Kümmell’s patients with different stages and postural correction status vol.17, pp.4, 2012, https://doi.org/10.1080/17434440.2020.1736039
  12. A COMPARATIVE ANALYSIS OF THE EFFECTS OF TWO AUGMENTERS ON THE SENSITIVITY OF VERTEBRAL BIOMECHANICAL BEHAVIOR IN VERTEBROPLASTY vol.20, pp.4, 2012, https://doi.org/10.1142/s0219519420500074
  13. Influence of cement-augmented pedicle screw instrumentation in an osteoporotic lumbosacral spine over the adjacent segments: a 3D finite element study vol.15, pp.None, 2020, https://doi.org/10.1186/s13018-020-01650-5
  14. Short-term Efficacy of the Percutaneous Vertebroplasty Using a Curved Versus Straight Vertebroplasty Needle in Osteoporotic Vertebral Compression Fractures vol.44, pp.1, 2021, https://doi.org/10.3928/01477447-20201012-03
  15. Functional Properties of Low-Modulus PMMA Bone Cements Containing Linoleic Acid vol.12, pp.1, 2021, https://doi.org/10.3390/jfb12010005
  16. A study on the relationship between the rate of vertebral body height loss before balloon kyphoplasty and early adjacent vertebral fracture vol.34, pp.4, 2012, https://doi.org/10.3233/bmr-200121
  17. Finite element analysis of wedge and biconcave deformity in four different height restoration after augmentation of osteoporotic vertebral compression fractures vol.16, pp.1, 2021, https://doi.org/10.1186/s13018-021-02225-8