한국인 Cadaver Test에 대한 상완골 골수정 개발에 관한 연구

A Study on Development of Humeral Intramedullary Fixation Nail based on Korean Cadaver Tests

  • 전창수 ((주)솔고바이오메디칼 연구소) ;
  • 이재원 ((주)바이오알파) ;
  • 고철웅 (한국생산기술연구원 실버기술개발단) ;
  • 오종건 (고려대학교 구로병원 정형외과) ;
  • 우수현 (인제대학교 의용공학과 생체역학실험실) ;
  • 이성재 (인제대학교 의용공학과 생체역학실험)
  • 투고 : 2010.12.01
  • 심사 : 2011.05.26
  • 발행 : 2011.08.01

초록

The aim of this study is to develop a humeral Intramedullary fixation nail (HIFN) suitable for Korean people. In this study, CT images were obtained from 72 Korean cadaveric humeral bones and 3D Korean humeral bone models were reconstructed based on the CT images to investigate anatomical characteristics. Major design parameters of HIFN were selected using the morphological measurement information of the Korean humeral bone models. Through finite element analysis and mechanical tests, the developed HIFN prototype was compared with the Polarus HIFN ($ACUMED^{(R)}$, USA), and it was found that the HIFN prototype showed similar and/or superior mechanical performance compared to the Polarus HIFN. Also, clinical validation for the HIFN prototype was carried out to check predictable troubles in surgical operations. Finally, optimal design modification was proposed to prevent the possible axillary nerve injury due to the locking screw system of the HIFN prototype.

키워드

참고문헌

  1. Bigliani, L. U., Flatow, E. L. and Pollock, R. G., "Fractures of the Proximal Humerus, 2nd Ed.," The Shoulder, pp. 337-389, 1998.
  2. Bengner, U., Johnell, O. and Redlund-Johnell, I., "Changes in the incidence of fracture of the upper end of the humerus during a 30-year period," Clin. Orthop. Relat. Res., Vol. 231, pp. 179-182, 1988.
  3. Park, S. R., Lee, T. J., Kim, R. S., Moon, K. H. and You, D. S., "Result of Interlocking Intramedullary Nailing for Humeral Shaft Fracture Evaluation of Post-operative Shoulder Function," J. of the Korean Fracture Society, Vol. 20, No. 2, pp. 166-171, 2007. https://doi.org/10.12671/jkfs.2007.20.2.166
  4. Linn, J., "Treatment of Humeral Shaft Fractures with Humeral Locked Nail and Comparison with Plate Fixation," J. Trauma, Vol. 44, No. 5, pp. 859-864, 1998. https://doi.org/10.1097/00005373-199805000-00019
  5. Blum, J., Janzing, H., Gahr, R., Langendorff, H. S. and Rommens, P. M., "Clinical Performance of a New Medullary Humeral Nail: Antegrade versus Retrograde Insertion," J. Orthop. Trauma, Vol. 15, No. 5, pp. 342-349, 2001. https://doi.org/10.1097/00005131-200106000-00007
  6. Fuchtmeier, B., May, R., Hente, R., Maghsudi, M., Volk, M., Hammer, J., Nerlich, M. and Prantl, L., "Proximal Humerus Fractures: A Comparative Biomechanical Analysis of Intra and Extramedullary Implants," Arch. Orthop. Trauma Surgery, Vol. 127, No. 6, pp. 441-447, 2007. https://doi.org/10.1007/s00402-007-0319-6
  7. Oh, J. K., Oh, C. W., Park, S. H., Roh, K. J., Jeong, C. W., "Conformity of the LCP-DF (Locking Compression Plate-Distal Femur) in Korean Adult Femur - A Cadaver Study," J. of the Korean Fracture Society, Vol. 18, No. 4, pp. 399-404, 2005. https://doi.org/10.12671/jkfs.2005.18.4.399
  8. Yoo, K. J., Jung, T. G., Ha, J. H., Hwang, H. Y., Ko, C. W., Lim, D. S., Lee, J. W., Lee, S. J. and Oh, J. G., "Mechanical Study of Humeral Intramedullary Fixation Nail," The 19th JSME Bioengineering Conference, pp. 177-178, 2007.
  9. Yoo, K. J., Lee, J. W., Jung, T. G., Lee, S. J., Ha, J. H., Lim, D. S., Oh, J. G. and Ko, C. W., "A Biomechanical Study on Development of Humeral Intramedullary Fixation Nail based on Korean CT Information," The 20th JSME Bioengineering Conference, pp. 187-188, 2008.
  10. Kamineni, S., Ankem, H. and Sanghavi, S., "Anatomical Considerations for Percutaneous Proximal Humeal Fracture Fixation," Injury, Vol. 35, No. 11, pp. 1133-1136, 2004. https://doi.org/10.1016/j.injury.2003.08.008
  11. ASTM F 1264 Standards, "Standard Specification and Test Methods for Intramedullary Fixation Device," Annual Book of ASTM Standards, Vol. 13.01, pp. 603-620, 2001.