Rat Magnetocardiograms Measured by YBCO SQUID Magnetometer

YBCO SQUID를 이용한 실험쥐의 심자도 측정

  • Ahn, San (Korea Research Institute of Standards and Science) ;
  • Kim, I.S. (Korea Research Institute of Standards and Science) ;
  • Song, J.H. (Chungnam National University)
  • Received : 2011.03.19
  • Accepted : 2011.04.04
  • Published : 2011.04.30

Abstract

We have developed a high-$T_C$ SQUID magnetometer system to measure magnetocardiograms of laboratory rats. White noise of the measurement system was about 50 fT/$Hz^{1/2}$ when measured in a magnetically shielded box. We optimized the measurement position to obtain clear MCG wave from rat's small heart by using grid measurements. With the optimization, the MCG signal was successfully detected with the peak amplitude of about 50 pT. We could observe well defined P-, QRS-, and T-wave from the rat MCG. The results suggest that the developed system has a strong potential to monitor the progress of heart disease model using laboratory rat.

Keywords

References

  1. V. Pizzella, S. D. Penna, C. D. Gratta, and G. L. Romani, "SQUID systems for biomagnetic imaging", Supercond. Sci. Technol., vol. 14, pp. R79-R114, Apr. 2001. https://doi.org/10.1088/0953-2048/14/7/201
  2. I. Tavarozzi, S. Comani, C. D. Gratta, G. L. Romani, S. D. Luzio, D. Brisinda, S. Gallina, M. Zimarino, R. Fenici, R. D. Caterina, "Magnetocardiography: current status and perspective. Part I: Physical principles and instrumentation", Ital. Heart J. vol. 3, pp. 75-85, 2002.
  3. I. Tavarozzi, S. Comani, C. D. Gratta, S. D. Luzio, G. L. Romani, S. Gallina, M. Zimarino, D. Brisinda, R. Fenici, R. D. Caterina, "Magnetocardiography: current status and perspectives. Part II: Clinical applications", Ital. Heart J. vol. 3, PP. 151-165, 2002.
  4. S. Yamada, I. Yamaguchi, "Magnetocardiograms in clinical medicine: Unique information on cardiac ischemia, arrhythmias, and fetal diagnosis", Internal Medicine, vol. 44, pp. 1-99, 2005. https://doi.org/10.2169/internalmedicine.44.1
  5. K. Sternickel, A. I. Braginski, "Biomagnetism using SQUIDs: status and perspective", supercon. Sci. Technol., vol. 44, pp. S160-S171, 2006.
  6. D. Brisinda, M. E. Caisto, and Riccardo Fenici, "Contactless magnetocardiographic mapping in anesthetized Wistar rats: evidence of age-related changes of cardiac electrical activity", Am J Physiol Heart Circ Physiol 291: H368-H378, 2006. https://doi.org/10.1152/ajpheart.01048.2005
  7. R. Fischer, A. Gapelyuk, N. Wessel, K. Gruner, A. Gruner, D. Müller, R. Dietz, A. Schirdewan, "Time course of changes in cardiac magnetic field mapping after myocardial infarction in rats", International Congress Series 1300, pp. 484-487, 2007. https://doi.org/10.1016/j.ics.2006.12.018
  8. Y. Aita, S. Kambe, T. Nagaishi and H. Itozaki, "Magnetocardiogram of a rat observed by a high-TC SQUID sensor system", Physica C, pp. 378-381, 1400-3, 2002. https://doi.org/10.1016/S0921-4534(02)01734-3
  9. S. Kuriki, A. Hayashi, Y. Hirata, T. Naiki and K. Kawahara, "High-TC SQUID magnetometer for measurements of heart-rate signal of small animal", Chin. J. Phys. Vol. 42, pp. 501-507, 2004.
  10. I. S. Kim, K. K. Yoo, Y. H. Lee, K. W. Kim and Y. K. Park, "High-TC SQUID magnetometers for low noise measurements of magnetocardiograms", IEEE Trans. Appl. Supercond. vol. 15, pp. 652-655, 2005. https://doi.org/10.1109/TASC.2005.849985
  11. H. C. Bazett, "An analysis of the time-relations of electrocardiograms", Heart (7): 353-370, 1920.
  12. L. S. Fridericia, "The duration of systole in the electrocardiogram of normal subjects and of patients with heart disease", Acta Medica Scandinavica (53): 469-486, 1920.
  13. A. Sagie, M. G. Larson, R. J. Goldberg, J. R. Bengston, D. Levy, "An improved method for adjusting the QT interval for heart rate (the Framingham Heart Study)", Am J Cardiol 70 (7): 797-801, 1992. https://doi.org/10.1016/0002-9149(92)90562-D