The Usefulness Assessment of Verifying Daily Output by Using CHECKMATE$^{TM}$

CHECKMATE$^{TM}$를 이용한 일일 출력 검증의 유용성 평가

  • Cho, Han-Sang (Department of Radiation Oncology, Ajou University Hospital) ;
  • Nam, Sang-Soo (Department of Radiation Oncology, Ajou University Hospital) ;
  • Park, Hae-Jin (Department of Radiation Oncology, Ajou University Hospital) ;
  • Kim, Mi-Hwa (Department of Radiation Oncology, Ajou University Hospital) ;
  • Park, An-Tae (Department of Radiation Oncology, Ajou University Hospital)
  • 조한상 (아주대학교부속병원 방사선종양학과) ;
  • 남상수 (아주대학교부속병원 방사선종양학과) ;
  • 박혜진 (아주대학교부속병원 방사선종양학과) ;
  • 김미화 (아주대학교부속병원 방사선종양학과) ;
  • 박안태 (아주대학교부속병원 방사선종양학과)
  • Received : 2010.09.28
  • Accepted : 2011.01.21
  • Published : 2011.03.31

Abstract

Purpose: In this study, we tried to check the usefulness of two Linear Accelerators, Clinac IX and 21EX (Varian, Palo Alto, CA), which are equipped in Ajou Medical Center. From 2008 to 2010, we evaluated the error range of Absolute Dose based on the daily output, which was measured by CHECKMATE$^{TM}$ (Sun Nuclear, Melbourne, FL). Materials and Methods: For Daily Q.A, photon beams of two linear accelerators, 21EX and IX (6 MV and 10 MV, respectively) were measured daily by using CHECKMATE$^{TM}$ just before the treatment began, while the absolute dose was measured biweekly by using water phantom. We analyzed the data of measured values from the daily Q.A and the absolute dose from 2008 to 2010 for 21EX, and from 2009 to 2010 for IX. We utilized Excel 2007 (Microsoft, USA) to evaluate Average, Standard deviation and Confidence level of the data. Furthermore, in order to check the measured values of CHECKMATE$^{TM}$ and the significance of absolute dose, each error value was compared and analyzed. Results: During the observation period, the output of two equipment's absolute dose increased in process of time and in both 6 MV and 10 MV, there was a similar increasing trend. In addition, the error rate of the measured value of CHECKMATE$^{TM}$ and the value of absolute dose were under 0.34, which means that there is a similarity relationship between the two measured values. After checking that the measured value of CHECKMATE$^{TM}$ increased, We measured the absolute dose to adjust that. When the error range was close to 2~3%, the number of changing the output was four for 21EX and three for IX. Conclusion: As a result of measuring and analyzing the daily output changes for two years by using CHECKMATE$^{TM}$, we could find that there is a significance between the output which we should obey during Q.A, and the measured value of absolute dose within the error tolerance of 2~3%. Thus, the use of CHECKMATE$^{TM}$ can be positively considered for more efficient and reliable daily output verification of linear accelerator. It can also be a good standard for other medical centers to understand the trends of linear accelerator and to refer to for the correction of each output.

목 적: 본 연구에서는 본원에서 보유하고 있는 선형가속기 Clinac IX과 21EX (Varian, Palo Alto, CA)에 대하여 2008년부터 2010년까지 CHECKMATE$^{TM}$ (Sun Nuclear, Melbourne, FL)를 이용하여 측정한 일일 출력 값을 바탕으로 절대 선량에 대한 오차 범위를 평가함으로써 그 유용성을 확인하고자 하였다. 대상 및 방법: 일일 정도 관리는 Varian사의 21EX, IX 두 대의 선형가속기의 광자선(6 MV, 10 MV)을 대상으로 매일 치료시작 전 CHECKMATE$^{TM}$를 사용하여 측정되었고, 절대선량은 격주간격으로 물 팬텀을 이용하여 측정되었다. 21EX의 경우 2008년부터 2010년, IX는 2009년부터 2010년까지의 일일 정도 관리의 측정치와 절대선량 측정치의 데이터 값을 분석하였다. 각 데이터는 Excel 2007 (Microsoft, USA)을 이용하여 평균, 표준편차, 신뢰수준을 평가하였고, CHECKMATE$^{TM}$의 측정값과 절대선량의 유의성을 확인하기 위하여 각 데이터의 오차 값을 각각 비교 분석하였다. 결 과: 관찰 기간 동안 두 장비의 절대선량의 출력은 시간이 지남에 따라 점차 증가함을 보였고, 이 증가는 6 MV와 10 MV 두 에너지 모두에서 비슷한 양의 증가 추세를 보이고 있었다. 또한 CHECKMATE$^{TM}$의 측정치와 절대선량 값의 오차율은 모두 0.34 이하로 나타나 두 측정값이 서로 유의관계가 있음을 보였다. CHECKMATE$^{TM}$ 측정값이 상향됨을 확인 후 이를 조정하기 위하여 절대선량 측정하였고, 오차범위 2~3%에 근접한 경우 출력 값을 교정한 횟수는 21EX는 4회, IX는 3회였다. 결 론: CHECKMATE$^{TM}$를 이용하여 2년 동안 일일 출력 변화를 측정하고 분석한 결과, 품질 관리 시 준수해야 하는 출력 값이 허용 오차 범위 2~3% 내에서 절대선량의 측정값들과 유의성이 있음을 보였다. 따라서 선형가속기의 보다 효율적이며 신뢰도가 있는 일일 출력 검증을 위하여, CHECKMATE$^{TM}$의 사용을 고려할 수 있다. 또한 각 기관에서의 선형가속기 출력 값의 변화 추세를 파악하고 각 출력 값 교정에 참고할 수 있는 좋은 기준이 될 수 있다고 사료된다.