DOI QR코드

DOI QR Code

Ceramic on Ceramic Total Hip Arthroplasty - The Results of a Minimum 10-Years Follow up -

세라믹-세라믹형 인공 고관절 전치환술 - 최소 10년 이상 추시 보고 -

  • Moon, Do-Hyun (Department of Orthopedic Surgery, Gil Medical Center, College of Medicine, Gachon University) ;
  • Choi, Jang-Seok (Department of Orthopedic Surgery, Gil Medical Center, College of Medicine, Gachon University) ;
  • Kim, Dong-Whan (Department of Orthopedic Surgery, Gil Medical Center, College of Medicine, Gachon University) ;
  • Kim, Kwang-Hee (Department of Orthopedic Surgery, Gil Medical Center, College of Medicine, Gachon University)
  • 문도현 (가천의과학대학교 길병원 정형외과학교실) ;
  • 최장석 (가천의과학대학교 길병원 정형외과학교실) ;
  • 김동환 (가천의과학대학교 길병원 정형외과학교실) ;
  • 김광희 (가천의과학대학교 길병원 정형외과학교실)
  • Published : 2011.06.30

Abstract

Purpose: We wanted to evaluate the clinical and radiographic results of ceramic on ceramic bearing cementless total hip arthroplasty with a minimum of 10-years follow up. Materials and Methods: This study was performed on 44 patients and 51 hips, and the patients were recruited among the patients who underwent ceramic-ceramic type cementless total hip arthroplasty from August 1998 to February 2000 and they were available for 10-years or longer follow up. The gender ratio was 28 males and 16 females, and the mean age at the time of surgery was 49 years (range: 26~77 years). Regarding the causality of disease, osteonecrosis was noted in 36 cases, degenerative arthritis was noted in 4 cases, acetabular dysplasia was noted in 4 cases and septic or tuberculous hip sequelae was noted in 7 cases. The clinical evaluation was performed according to the Harris hip score, and for the radiological evaluation of the femoral component, we assessed the degree of subsidence and the presence of periprosthetic osteolysis, endosteal new bone formation and cortical hypertrophy, and the final fixation state of the implant. In the cases with an acetabular cup osteoscleostic line in the vicinity of the acetabular cup, endosteal new bone formation, acetabular osteolysis and the vertical migration as well as the horizontal migration of the acetubular cup were assessed. Results: Regarding the clinical outcomes, the Harris hip score was improved from an average 59 points to 93 points at the final follow up. For the radiological results, osteolysis was not detected in the acetabular and femoral components, the subsidence of femoral stem was not beyond 2 mm on average and significant subsidence was not detected. In regard to complications, linear fracture during surgery was noted in 6 cases and postoperative hip dislocation was noted in 1 case. A ceramic head fracture occurred in one case and a ceramic liner fracture occurred in 2 cases. Conclusion: This type of articulation is a promising option at a minimum of 10-years follow up of 51 hips that underwent ceramic-ceramic bearing cementless total hip arthroplasty. However, our results demonstrated that surgeons should be aware of the potential risks of ceramic fracture when using ceramic-on-ceramic bearing surfaces. Additional longer-term follow-up is necessary.

목적: 세라믹-세라믹 관절면을 이용한 무시멘트 인공고관절 전치환술 후 10년 이상의 추시 관찰한 임상적, 방사선학적 결과를 보고하고자 한다. 대상 및 방법: 1998년 8월부터 2000년 2월까지 세라믹-세라믹 관절면을 이용한 무시멘트 인공 고관절 전치환술을 시행한 환자 중 10년 이상 추시 가능하였던 44명, 51예의 고관절을 대상으로 하였다. 44명의 환자 중 남자 28명, 여자 16명이었으며, 수술시 평균연령은 49세(26~77세)였다. 원인 질환으로는 대퇴 골두 무혈성 괴사 36예, 퇴행성관절염 4예, 고관절 이형성증 4예, 화농성 및 결핵성 고관절 후유증 7예였다. 임상적 평가는 Harris 고관절 점수에 근거하였으며, 방사선학적 평가는 대퇴삽입물의 경우 수직침강의 정도, 삽입물 주위의 골용해, 골내막 신생골의 형성, 피질골 비후의 유무, 삽입물의 최종 고정 상태를 관찰하였다. 비구컵의 경우 비구컵 주위의 경화선, 골내막 신생골 형성, 비구 골용해, 비구컵의 수직, 수평 이동을 관찰하였다. 결과: 임상 결과는 Harris 고관절 점수가 술 전 평균 59점에서 최종 추시 시 93점으로 향상되었다. 비구컵과 대퇴 삽입물 주위의 골 용해 소견은 없었고, 대퇴 스템의 침강은 평균 2 mm 이하로 의미 있는 침강소견을 보이는 경우는 없었다. 합병증으로 수술시 대퇴골 선상골절이 6예, 술 후 재발성 고관절 탈구가 1예, 세라믹 골두 파손이 1예, 라이너 파손이 2예에서 발생하였다. 결론: 세라믹-세라믹 관절면을 이용한 무시멘트 인공 고관절 전치환술을 시행 받은 51예의 10년 이상 추시결과 Harris 고관절 점수의 향상과 전례에서 골용해 소견 관찰되지 않았고 안정적인 골침습적 고정을 보이는 양호한 결과를 관찰할 수 있었다. 그러나 세라믹 골두의 파손이 1예, 라이너의 파손이 2예에서 발생하여 향후 보다 많은 예의 장기적인 추시 관찰이 필요할 것으로 사료된다.

Keywords

References

  1. Willert HG, Semlitsch M. Reactions of the articular capsule to wear products of artificial joint prostheses. J Biomed Mater Res. 1977;11:157-64. https://doi.org/10.1002/jbm.820110202
  2. Sedel L. Evolution of alumina-on-alumina implants: a review. Clin Orthop Relat Res. 2000;379:48-54.
  3. Nizard RS, Sedel L, Christel P, Meunier A, Soudry M, Witvoet J. Ten-year survivorship of cemented ceramicceramic total hip prosthesis. Clin Orthop Relat Res. 1992;282:53-63.
  4. Yoon TR, Rowe SM, Jung ST, Seon KJ, Maloney WJ. Osteolysis in association with a total hip arthroplasty with ceramic bearing surfaces. J Bone Joint Surg Am. 1998;80:1459-68.
  5. Han CD, Choi CJ, Yang IW, Lee DH. Total hip arthroplasty with ceramic-on-ceramic articulation - minimum 5-year follow-up results. J Korean Orthop Assoc. 2006;41:421-7.
  6. Bizot P, Larrouy M, Witvoet J, Sedel L, Nizard R. Pressfit metal-backed alumina sockets: a minimum 5-year followup study. Clin Orthop Relat Res. 2000;379:134-42.
  7. Garino JP. Modern ceramic-on-ceramic total hip systems in the United States: early results. Clin Orthop Relat Res. 2000;379:41-7.
  8. Murphy SB. Ceramic-ceramic bearings in THA: the new gold standard--in the affirmative. Orthopedics. 2002;25:933-4.
  9. Kim YM, Kim SR, Rhyu KH, Kim JK, Yoo JJ, Kim HJ. Short term results of ceramic-on-ceramic bearing Bicontact total hip arthroplasty. J Korean Hip Soc. 2001;13:1-7.
  10. Lee SH, Kwon KD, Cho SH, Shin SJ, Youm YS. Ceramic on ceramic in cementless total hip arthroplasty: preliminary reports with minimal 3-year follow up. J Korean Hip Soc. 2003;15:124-30.
  11. Yoo JJ ,Kim YM, Yoon KS, et al. Alumina-on-alumina THA in active patients younger than 40 years old-A 5-year minimum follow-up study. J Korean Orthop Assoc. 2006;41:404-12.
  12. Bands R, Pelker RR, Shine J, Bradburn H, Margolis R, Leach J. The noncemented porous-coated hip prosthesis. A three-year clinical follow-up study and roentgenographic analysis. Clin Orthop Relat Res. 1991;269:209-19.
  13. Callaghan JJ, Dysart SH, Savory CG. The uncemented porous-coated anatomic total hip prosthesis. Two-year results of a prospective consecutive series. J Bone Joint Surg Am. 1988;70:337-46.
  14. Gruen TA, McNeice GM, Amstutz HC. "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop Relat Res. 1979;141:17-27.
  15. Engh CA, Bobyn JD, Glassman AH. Porous-coated hip replacement. The factors governing bone ingrowth, stress shielding, and clinical results. J Bone Joint Surg Br. 1987;69:45-55.
  16. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop Relat Res. 1976;121:20-32.
  17. Mallory TH, Kraus TJ, Vaughn BK. Intraoperative femoral fractures associated with cementless total hip arthroplasty. Orthopedics. 1989;12:231-9.
  18. Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518-28.
  19. Boutin P. Total arthroplasty of the hip by fritted aluminum prosthesis. Experimental study and 1st clinical applications. Rev Chir Orthop Reparatrice Appar Mot. 1972;58:229-46.
  20. Mittelmeier T, Walter A. The influence of prosthesis design on wear and loosening phenomena. CRC Crit Rev Biocompat. 1987;3:19.
  21. Bierbaum BE, Nairus J, Kuesis D, Morrison JC, Ward D. Ceramic-on-ceramic bearings in total hip arthroplasty. Clin Orthop Relat Res. 2002;405:158-63.
  22. Pfaff HG. Ceramic component failure and the role of proof testing. Clin Orthop Relat Res. 2000;379:29-33.
  23. Clarke IC, Good V, Williams P, et al. Ultra-low wear rates for rigid-on-rigid bearing in total hip replacements. Proc Inst Mech Eng H. 2000;214:331-47. https://doi.org/10.1243/0954411001535381
  24. Lee YK, Ha YC, Yoo JJ, Koo KH, Yoon KS, Kim HJ. Alumina-on-alumina total hip arthroplasty: a concise follow-up, at a minimum of ten years, of a previous report. J Bone Joint Surg Am. 2010;92:1715-9. https://doi.org/10.2106/JBJS.I.01019
  25. Heros R, Willmann G. Ceramic in total hip arthroplasty: history, mechanical properties, clinical results and current state of the art. Seminars Arthroplasty. 1998;9:114-22.
  26. Sedel L, Nizard R, Bizot P, Meunier A. Perspective on a 20-year experience with ceramic-on-ceramic articulation in total hip replacement. Semin Arthroplasty. 1998;9:123-34.
  27. Sedel L, Kerboull L, Christel P, Meunier A, Witvoet J. Alumina-on-alumina hip replacement. Results and survivorship in young patients. J Bone Joint Surg Br. 1990;72:658-63.
  28. Koo KH, Ha YC, Jung WH, Kim SR, Yoo JJ, Kim HJ. Isolated fracture of the ceramic head after thirdgeneration alumina-on-alumina total hip arthroplasty. J Bone Joint Surg Am. 2008;90:329-36. https://doi.org/10.2106/JBJS.F.01489
  29. Ha YC, Kim SY, Kim HJ, Yoo JJ, Koo KH. Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty. Clin Orthop Relat Res. 2007;458:106-10.
  30. Park YS, Hwang SK, Choy WS, Kim YS, Moon YW, Lim SJ. Ceramic failure after total hip arthroplasty with an alumina-on-alumina bearing. J Bone Joint Surg Am. 2006;88:780-7 https://doi.org/10.2106/JBJS.E.00618
  31. Min BW, Song KS, Kang CH, Bae KC, Won YY, Lee KY. Delayed fracture of a ceramic insert with modern ceramic total hip replacement. J Arthroplasty. 2007;22:136-9. https://doi.org/10.1016/j.arth.2005.12.012
  32. Toni A, Traina F, Stea S, et al. Early diagnosis of ceramic liner fracture. Guidelines based on a twelve-year clinical experience. J Bone Joint Surg Am. 2006;88 Suppl:55-63. https://doi.org/10.2106/JBJS.F.00587
  33. Gallo J, Stewart T, Novotny R, Dusza J, Galusek D. Early fracture of a plasma cup ceramic liner: a case report and surface analysis. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2007;151:341-6. https://doi.org/10.5507/bp.2007.058
  34. Lee YK, Yoo JJ, Koo KH, Yoon KS, Kim HJ. Metal neck and liner impingement in ceramic bearing total hip arthroplasty. J Orthop Res. 2011:29:218-22. https://doi.org/10.1002/jor.21246
  35. Yang IH, Han CD, Lee JS. Cementless total hip arthroplasty using ceramic-on-ceramic oseonics system: minimum 8-year follow up results. J Korean Orthop Assoc. 2009;44:520-5. https://doi.org/10.4055/jkoa.2009.44.5.520

Cited by

  1. Liner Dissociation by Extrusion of the Acetabular Cup Fixation Screw after Total Hip Replacement : Two Cases Report vol.25, pp.3, 2011, https://doi.org/10.5371/hp.2013.25.3.220