DOI QR코드

DOI QR Code

The Acoustic Output Estimation for Therapeutic Ultrasound Equipment using Electro-Acoustic Radiation Conductance

전기-음향 방사컨덕턴스를 이용한 치료용 초음파 자극기의 음향출력 예측

  • Yun, Yong-Hyeon (Department of Electrical & Electronic Engineering, Graduate School of Yonsei University) ;
  • Jho, Moon-Jae (Division of Technology Services, Korea Research Institute of Standards and Science) ;
  • Kim, Yong-Tae (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Lee, Myoung-Ho (Department of Electrical & Electronic Engineering, Graduate School of Yonsei University)
  • 윤용현 (연세대학교 대학원 전기전자공학과) ;
  • 조문재 (한국표준과학연구원 성과확산부 표준보급센터) ;
  • 김용태 (한국표준과학연구원 기반표준본부 유동음향센터) ;
  • 이명호 (연세대학교 대학원 전기전자공학과)
  • Received : 2011.05.06
  • Accepted : 2011.06.29
  • Published : 2011.09.30

Abstract

To increase therapeutic efficiency and biological safety, it is important to precision control of acoustic output for therapeutic ultrasound equipment. In this paper, the electro-acoustic radiation conductance, one of electroacoustic characteristics of therapeutic ultrasound equipment, was measured by the radiation force balance method according to IEC 61161 standards and the acoustic output was estimated using the electro-acoustic radiation conductance. The estimation of acoustic output was conducted to continuous wave mode and pulse wave mode of duty cycle between 20% and 80%. The differences between prediction values and measurement results are within 5% of measurement uncertainty, which is a reasonably good agreement. The results show that acoustic output controlled by electro-acoustic radiation conductance was found to be an effective method.

Keywords

References

  1. C.R. Hill and G. ter Haar, "Ultrasound in non-ionizing radiation protection," WHO Regional Publications, European Series. No. 10, World Health Organisation, Copenhagen, 1981.
  2. H.F. Stewart, G.R. Harris, B.A. Herman, R.A. Robinson, M.E. Haran, G.R. McCall, G. Carless, and D. Rees, "Survey of use and performance of ultrasonic therapy equipment in Pinellas County, Florida," Phys. Ther., vol. 54, no. 7, pp. 707-715, 1974. https://doi.org/10.1093/ptj/54.7.707
  3. M.H. Repacholi and D.A. Benwell, "Using surveys of ultrasound therapy devices to draft performance standards," Health Phys., vol. 36, no. 6, pp. 679-686, 1979. https://doi.org/10.1097/00004032-197906000-00003
  4. S.D. Pye, and C. Milford, "The performance of ultrasound physiotherapy machines in Lothian Region, Scotland, 1992," Ultrasound Med. Biol., vol. 20, no. 4, pp. 347-359, 1994. https://doi.org/10.1016/0301-5629(94)90003-5
  5. S. Pye, "Ultrasound Therapy Equipment--Does it Perform?," Physiotherapy, vol. 82, no. 1, pp. 39-44, 1996. https://doi.org/10.1016/S0031-9406(05)66996-9
  6. S. Pye and B. Zeqiri, Guidelines for the testing and calibration of physiotherapy ultrasound machines, Institute of Physics and Engineering in Medicine Report No. 84., York, 2001.
  7. P.A. Artho, J.G. Thyne, B.P. Warring, C.D. Willis, J.M. Brismee, and N.S. Latman, "A calibration study of therapeutic ultrasound units," Phys. Ther., vol. 82, no. 3, pp. 257-263, 2002.
  8. R. Guirro, and S. C. Britshcy Dos Santos, "Evaluation of the acoustic intensity of new ultrasound therapy equipment," Ultrasonics, vol. 39, no. 8, pp. 553-557, 2002. https://doi.org/10.1016/S0041-624X(02)00251-2
  9. R.T. Hekkenberg, A. Richards, K. Beissner, B. Zeqiri, R.A. Bezemer, M. Hodnett, G. Prout, and C. Cantrall, "Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines," Ultrasound Med. Biol., vol. 32, no. 9, pp. 1423-1429, 2006. https://doi.org/10.1016/j.ultrasmedbio.2006.04.001
  10. IEC, "IEC 61161," Ultrasonics - Power measurement - Radiation force balances and performance requirements International Electrotechnical Commission, 2006.
  11. IEC, "IEC 61101," Ultrasonics - The absolute calibration of hydrophones using the planar scanning technique in the frequency range 0.5 MHz to 15 MHz, International Electrotechnical Commission, 1991.
  12. M.C. Ziskin and P.A. Lewin, Ultrasonic exposimetry, Boca Raton, FL: CRC Press, 1992.
  13. B. Fay and M. Rinker, "Determination of the absolute value of ultrasonic power by means of thermoacoustic sensors," Acustica, vol. 82, no. 2, pp. 274-279, 1996.
  14. K. Beissner, "Acoustic radiation pressure in the near field," J. Sound Vibr., vol. 93, no. 4, pp. 537-548, 1984. https://doi.org/10.1016/0022-460X(84)90422-X
  15. K. Beissner, "Primary measurement of ultrasonic power and dissemination of ultrasonic power reference values by means of standard transducers," Metrologia, vol. 36, no. 4, pp. 313-320, 1999. https://doi.org/10.1088/0026-1394/36/4/10
  16. S.E. Fick, "Ultrasound power measurement by pulsed radiation pressure," Metrologia, vol. 36, no. 4, pp. 351-356, 1999. https://doi.org/10.1088/0026-1394/36/4/15
  17. 조문재, 김용태, 윤용현, 정성수, "초음파 파워 측정 시스템구성 및 성능평가," 한국음향학회지, vol. 24, no. 8, pp. 431-440, 2005.
  18. B. Zeqiri and C.J. Bickley, "A new anechoic material for medical ultrasonic applications," Ultrasound Med. Biol., vol. 26, no. 1, pp. 481-485, 2000. https://doi.org/10.1016/S0301-5629(99)00147-7
  19. L.E. Kinsler, Fundamentals of acoustics, 3rd ed., New York: Wiley, 1982.
  20. J.M.M. Pinkerton, "The Absorption of Ultrasonic Waves in Liquids and its Relation to Molecular Constitution," Proc. Phys. Soc. B, vol. 62, no. 2, pp. 129-141, 1949. https://doi.org/10.1088/0370-1301/62/2/307
  21. 이충휘, 전기치료학, 서울: (주) 탑메디오피아, 2005.
  22. 윤용현, 조문재, 김용태, 이명호, "펄스 신호에 대한 초음파 파워 측정 시스템의 방사힘 응답 특성," in 2005년도 한국음향학회 춘계학술발표대회, 2005, pp. 193-194.