TA Study on Patient Exposure Dose Used the Phantom for Interventional Procedure

중재적 시술 시 팬텀을 이용한 환자의 피폭선량 분석

  • Kang, Byung-Sam (Department of Radiological Technology, Shingu University) ;
  • Dong, Kyung-Rae (Department of Radiological Technology, Gwangju Health College University)
  • Received : 2011.03.26
  • Accepted : 2011.04.27
  • Published : 2011.06.30

Abstract

Because interventional procedure operates looking at premier as real time when perate intervention enemy, by patient is revealed during suitableness time in radiation, side effect such as radiation injury of skin is apt to happen. It established by purpose of study that measure exposure dose that patient receives about these problem, and find solution for radiation injury and repletion method. In this study, we used Rando phantom of identical structure with the human body which becomes accomplished with 4 branch ingredient of the attempt and system equivalent material them and absorbed dose were measured by TLD. According to the laboratory, it shows that operations such as TFCA procedure or uterine myoma embolization are more dangerous than TACE procedure. If both operations are inspected during a short time, it is not affected in being bombed. However, it can lead to palliative agenesis or depilate, definitive agenesis only if operations are repeated more than three times. Dose distibution based on experiment, to reduce radiation exposure to patients result from reduction of scatter ray as we control field size of radiation and protection of side organs except for tumor. also we knew that we can protect patients form radiation exposure, if we increas SOD and decrease SID.