Abstract
Inveon PET is a recently developed preclinical PET system for small animal. This study was conducted to measure the performance of Inveon PET as recommended by the NEMA NU 4-2008. We measured the spatial resolution, the sensitivity, the scatter fraction and the NECR using a F-18 source. A 3.432 ns coincidence window was used. A $1\;mm^3$ sized F-18 point source was used for the measurement of spatial resolution within an energy window of 350~625 keV. PET acquisition was performed to obtain the spatial resolution from the center to the 5 cm offset toward the edge of the transverse FOV. Sensitivity, scatter fraction, and NECR were measured within an energy window of 350~750 keV. For measuring the sensitivity, a F-18 line source (length: 12.7 cm) was used with concentric 5 aluminum tubes. For the acquisition of the scatter fraction and the NECR, two NEMA scatter phantoms (rat: 50 mm in diameter, 150 mm in length; mouse: 25 mm in diameter, 70 mm in length) were used and the data for 14 half-lives (25.6 hr) was obtained using the F-18 line source (rat: 316 MBq, mouse: 206 MBq). The spatial resolution of the F-18 point source was 1.53, 1.50 and 2.33 mm in the radial, tangential and axial directions, respectively. The volumetric resolution was $5.43\;mm^3$ in the center. The absolute sensitivity was 6.61%. The peak NECR was 486 kcps @121 MBq (rat phantom), and 1056 kcps @128 MBq (mouse phantom). The values of the scatter fraction were 20.59% and 7.93% in the rat and mouse phantoms, respectively. The performances of the Inveon animal PET scanner were measured in this study. This scanner will be useful for animal imaging.
Inveon PET은 최근에 출시된 소동물 전용 PET 시스템이다. 이 연구에서는 Inveon PET 스캐너의 성능을 평가하기 위하여 공간 분해능, 민감도, 산란분획, 잡음등가계수(Noise equivalent count rate: NECR)를 측정하였다. 공간 분해능 측정은 에너지창 350~625 keV, 민감도, 산란분획, NECR 측정은 350~750 keV에서 수행하였고 동시계수창은 3.432 ns였다. 크기 $1\;mm^3$의 F-18 점 선원을 만들어 중심에서부터 5 cm 벗어난 위치까지 공간 분해능을 측정하였다. 민감도를 측정하기 위하여 스캐너의 축방향 길이와 동일한 길이 12.7 cm의 F-18 선 선원을 만들고 두께 2 mm의 알루미늄 관을 1개에서 5개까지 차례로 씌우며 절대 민감도를 계산하였다. 산란분획과 NECR 측정하기 위하여 두 가지 NEMA 산란 팬텀(랫(rat): 지름 50 mm, 길이 150 mm/마우스(mouse): 지름 25 mm, 길이 70 mm)을 이용하였고, F-18 선 선원(랫: 353 MBq, 마우스: 201 MBq)를 만들어 14반감기(25.6시간) 동안 데이터를 획득하였다. F-18의 중심에서 공간 분해능은 반경, 접선, 축 방향에서 각각 1.53, 1.50, 2.33 mm이고, 체적 공간 분해능은 $5.43\;mm^3$이었다. 절대민감도는 6.61%이었다. F-18 최대 NECR은 486 kcps @121 MBq (랫 팬텀), 1056 kcps @128 MBq (마우스 팬텀)이었다. 랫과 마우스의 산란분획은 각각 20.59%, 7.93%이었다. 이 연구에서 최신 소동물용 PET인 Inveon PET의 표준성능을 평가하였고 소동물 PET영상 획득에 유용함 을 보여주었다.