DOI QR코드

DOI QR Code

A Low-Power Portable ECG Touch Sensor with Two Dry Metal Contact Electrodes

  • Received : 2010.11.28
  • Published : 2010.12.31

Abstract

This paper describes the development of a low-power electrocardiogram (ECG) touch sensor intended for the use with two dry metal electrodes. An equivalent ECG extraction circuit model encountered in a ground-free two-electrode configuration is investigated for an optimal sensor read-out circuit design criteria. From the equivalent circuit model, (1) maximum sensor resolution is derived based on the electrode's background thermal noise, which originates from high electrode-skin contact impedance, together with the input referred noise of instrumentation amplifier (IA), (2) 60 Hz electrostatic coupling from mains and motion artifact are also considered to determine minimum requirement of common mode rejection ratio (CMRR) and input impedance of IA. A dedicated ECG read-out front end incorporating chopping scheme is introduced to provide an input referred circuit noise of 1.3 ${\mu}V_{rms}$ over 0.5 Hz ~ 200 Hz, CMRR of IA > 100 dB, sensor resolution of 7 bits, and dissipating only 36 ${\mu}W$. Together with 8 bits synchronous successive approximation register (SAR) ADC, the sensor IC chip is implemented in 0.18 ${\mu}m$ CMOS technology and integrated on a 5 cm $\times$ 8 cm PCB with two copper patterned electrodes. With the help of proposed touch sensor, ECG signal containing QRS complex and P, T waves are successfully extracted by simply touching the electrodes with two thumbs.

Keywords

References

  1. Fact sheet for obesity and cardiovascular disease: A report from the American Heart Association, 2009.
  2. D. Jabaudon, et.al, “Usefulness of ambulatory 7-day ECG monitoring for the detection of atrial fibrillation and flutter after acute stroke and transient ischemic attack,” Stroke, J.Amer. Heart Assoc., Vol.35, pp.1647-1651, May, 2004.
  3. E.Rodriguez-Cruz, Hypertension : Overview-eMedicine. Available : http://emedicine.medscape.com/article/889877-overview
  4. Task Force of the Eourpean Society of Cardiology and the North American Society of Pacing and Electrophysiology, “Heart Rate Variability: Standards of Measurement,” Physiological Interpretation, and Clinical Use, 93 1996.
  5. C.Jing, et al, “The Research on Emotion Recognition from ECG Signal,” International Conference on Information Technology and Computer Science, pp. 497-500, 2009.
  6. A.Haag, et al, “Emotion Recognition Using Bio-Sensors: First Steps Towards an Automatic System,” Proceedings of the Kloster Irsee Tutorial and Research workshop on Affective Dialogue Systems, pp.36-48, 2004.
  7. Rigas, et.al, “A User Independent, Biosignal Based, Emotion Recognition Method,” Lecture notes in Computer Science, No.4511, pp.314-318, 2007 https://doi.org/10.1007/978-3-540-73078-1_36
  8. J. Yoo, et.al, “A Wearable ECG Acquisition System With Compact Planar-Fashionable Circuit Board- Based Shirt,” IEEE Transactions on Information Technology in Biomedicine, Vol.13, pp.892-902, 2009. https://doi.org/10.1109/TITB.2009.2033053
  9. A. C-W. Wong, et al, “A 1V Micropower Systemon-Chip for Vital-Sign Monitoring in Wireless Body Sensor Networks,” ISSCC Dig. Tech. Papers, pp.138-139, Feb., 2008.
  10. E.M. Spinelli, et.al, “A Transconductance Driven-Right-Leg Circuit,” IEEE Transactions on Biomedical Engineering, Vol.47, pp.1466-1470, 1999. https://doi.org/10.1109/10.804574
  11. S.N. Pocock, “Earth-free patient monitoring, Part 1: principles, practice and hazards,” Biomed. Eng., Vol.7, pp.21-25, 1972.
  12. E.M. Spinelli, et.al, “Two-Electrode biopotential measurements: Power line interfere Analysis,” IEEE Transaction on biomedical engineering, Vol. 52, pp.1436-442, Aug. 2005. https://doi.org/10.1109/TBME.2005.851488
  13. D.E.Wood, et.al, “Comparative analysis of powr line interference between two-or three-electrode biopotential amplifiers,” Medical & Biological Engineering & Computing, Vol.33, pp.63-68, 1995. https://doi.org/10.1007/BF02522948
  14. N.V. Thakor, et.al, “Ground-free ECG Recording with Two Electrodes,” IEEE Transaction on biomedical engineering, Vol.BME-27, pp.699-704, Dec. 1980. https://doi.org/10.1109/TBME.1980.326595
  15. K. Hachisuka, et al, “Development and performance analysis of an intra-body communication device,” in Int. Solid-State Sens., Actuators, Microsyst. Conf., pp.1722-1725, 2003.
  16. George E. Bergey, et al, “Electrocardiogram recording with pasteless electrodes,” IEEE Trans. on Bio-medical Engineering, Vol.18, pp.206-211,1971. https://doi.org/10.1109/TBME.1971.4502833
  17. Long Yan, et al, “A Two-Electrode 2.88nJ/Conv. Biopotential Acquisition System for Portable Healthcare Device,” Proc. Of the IEEE Asian Solide-State Circuit Conference (A-SSCC), pp.329- 332, Nov., 2008.
  18. K.N.Leung and P.K.T.Mok, “A CMOS Voltage Reference based pm weighted ${\Delta}V_{GS}$ for CMOS Low-Dropout Linear Regulators,” IEEE Journal of Solid-State Circuits, Vol.38, pp.146-150, Jan., 2003. https://doi.org/10.1109/JSSC.2002.806265
  19. Refet Firat Yazicioglu, Chris Van Hoff, et.al, “A 60 uW 60 nV/srtHz Redout Front-End for Portable Biopotential Acquisition Systems,” IEEE Journal of Solid-State Circuits, Vol.42, pp.1100-1109, May., 2007. https://doi.org/10.1109/JSSC.2007.894804
  20. Christian C. ENZ and Gabor C. Temes, “Circuit Techniques for Reducing the Effects of Op-Amp Imperfections: Autozeroing, Correlated Double Sampling, and Chopper Stablization,” Proc. IEEE, Vol.84, pp.1584-1614, Nov., 1996. https://doi.org/10.1109/5.542410
  21. Datasheet of INA321, Simplified ECG Circuit for Medical Applications. Available : http://focus.ti.com/lit/ds/sbos168d/sbos168d.pdf

Cited by

  1. A Digital-Domain Calibration of Split-Capacitor DAC for a Differential SAR ADC Without Additional Analog Circuits vol.60, pp.11, 2013, https://doi.org/10.1109/TCSI.2013.2252475
  2. Low-Power Wearable ECG Monitoring System for Multiple-Patient Remote Monitoring vol.16, pp.13, 2016, https://doi.org/10.1109/JSEN.2016.2564995
  3. A low-power and miniaturized electrocardiograph data collection system with smart textile electrodes for monitoring of cardiac function vol.39, pp.4, 2016, https://doi.org/10.1007/s13246-016-0483-5
  4. A Study on Performance Evaluation for Electrocardiography Signal Measurement Electrode based on Conductive Fabric vol.50, pp.2, 2013, https://doi.org/10.5573/ieek.2013.50.2.210
  5. Fashionable wrist band using highly conductive fabric for electrocardiogram signal monitoring pp.1530-8057, 2018, https://doi.org/10.1177/1528083718779427