DOI QR코드

DOI QR Code

Microwave and RF Heating for Medical Application under Noninvasive Temperature Measurement Using Magnetic Resonance

  • Nikawa, Yoshio (Department of Health and Medical Engineering, School of Science and Engineering, Kokushikan University) ;
  • Ishikawa, Akira (Department of Health and Medical Engineering, School of Science and Engineering, Kokushikan University)
  • 투고 : 2010.10.01
  • 발행 : 2010.12.31

초록

Recent development of magnetic resonance imaging (MRI) equipment enables interventional radiology (IVR) as diagnosis and treatment under MRI usage. In this paper, a new methodology for magnetic resonance (MR) scanner to apply not only diagnostic equipment but for treatment one is discussed. The temperature measuring procedure under MR is to measure phase shift of $T_1$, which is the longitudinal relaxation time of proton, for the position inside a sample material with the application of pulsed RF for heating inside the sample as artificial dielectrics. The result shows the possibility to apply MR as temperature measuring equipment and as a heating equipment for applying such as hyperthermia heating modality.

키워드

참고문헌

  1. Hiroyuki Muranaka, Osamu Nakamura, Shinji Usui, Yoshitake Ueda, and Kaoru Morikawa, "Influence of implants on human body during MRI examinations: Fundamental experiment using metal balls," Japanese Society of Radiological Technology, vol. 61, no. 7, pp. 1015-1019, Jul. 2005.
  2. Hiroshi Sugimoto, "About defect report (MDR) to FDA concerning MRI," Japanese Society of Radiological Technology, vol. 61, no. 7, pp. 972-973, Jul. 2005. https://doi.org/10.6009/jjrt.KJ00003326898
  3. Toshiaki Miyaji, "Safety of MRI," Japanese Society of Radiological Technology, vol. 59, no. 12, pp. 1508-1514 Dec. 2003. https://doi.org/10.6009/jjrt.KJ00000921680
  4. "Hyperthermia manual," Japanese Society for Thermal Medicine Ed., 1999.
  5. Kenji Ino, Katsuhiko Hirai, Yoshiro Satake, Minoru Fukuda, Kazuo Mima, Haruo Sato, Kinji Abe, and Naoto Hayashi, "Preliminary study of IVR with open type MRI," Japanese Society of Radiological Technology, vol. 53, no. 7, pp. 962 Jul. 1997. https://doi.org/10.6009/jjrt.KJ00001355979
  6. "IVR Manual," IGAKU-SHOIN Ltd., Aug. 2002.
  7. Yoshio Nikawa, Masashi Nishioka, "Microwave diagnosis based on MRI imaging," GCMEA 2008/ MAJIC 1st, pp. 799-802, 2008.
  8. Michiko Dohi, Junta harada, Akiko Yoshihiro, Kazumi Komura, Takuji Mogami, and Tetsuhiko Takahashi, "MR temperature mapping 0.3T open type MR scanner," HITACHI MEDICAL Corporation, MEDIX, vol. 32 pp. 14-18 Sep. 2000.
  9. Yoshio Nikawa, "Microwave energy absorption in artificial dielectric material and metamaterial," Proceedings of GCMEA 2008/MAJIC 1st, pp. 263-266, 2008.
  10. Takayuki Takase, Yoshio Nikawa, "Optimization of element arrangement for metamaterial," GCMEA 2008/MAJIC 1st, pp. 819-822, 2008.
  11. Yoshio Nikawa, "Application of metamaterial in microwave energy absorption," 2008 Asia-Pacific Microwave Conference Proceedings, In CD, 4 pages, 2008.
  12. Yoshio Nikawa, "Thin and light applicator for localized microwave heating," Progress in Electromagnetics Research Symposium, p. 373, Jul. 2001.

피인용 문헌

  1. Study on Temperature Measurement Using MRI during Acupuncture and Moxibustion vol.100, pp.5, 2017, https://doi.org/10.1002/ecj.11958