DOI QR코드

DOI QR Code

Development of Auto-Masking Puretone Audiometer supporting Multiple Modes

다중모드 지원 자동차폐 순음청력검사 시스템 개발

  • Kim, Jin-Dong (Department of Biomedical Engineering, School of Medicine Pusan National University) ;
  • Shin, Bum-Joo (Department of Biomedical Engineering, College of Natural Resource and Life Science, Pusan National University) ;
  • Jeon, Gye-Rok (Department of Biomedical Engineering, School of Medicine Pusan National University) ;
  • Wang, Soo-Geun (Department of Otolaryngology, School of Medicine Pusan National University)
  • 김진동 (부산대학교 의학전문대학원 의공학교실) ;
  • 신범주 (부산대학교 바이오메디컬공학과) ;
  • 전계록 (부산대학교 의학전문대학원 의공학교실) ;
  • 왕수건 (부산대학교 의학전문대학원 이비인후과교실)
  • Published : 2009.06.30

Abstract

Puretone audiometer, which is a machine used for measuring the minimum hearing threshold, can be cost-effectively implemented using computer with sound card and software. In this paper, we describe a puretone audiometer which has been designed and implemented based on a general PC with sound card. It supports air conduction and bone conduction test taking with automatic masking. It also provides multiple modes consisted of self-test, auto-test and manual test mode. Such multiple modes makes it possible to use in various environments like as home and/or hospital. Through measure of waveform of output voltage and sound pressure, we verified that puretone audiometer of this paper properly operates.

최소가청역치를 측정하기 위한 기기인 순음청력검사 시스템은 사운드카드를 가진 컴퓨터에서 소프트웨어로 구현될 경우 가격 경쟁력을 가질 수 있다. 본 논문에서는 일반 PC를 기반으로 설계 및 구현된 순음청력검사 시스템을 기술한다. 개발된 PC 기반 순음청력검사 시스템은 기도 및 골도 검사시 지동 차폐 기능도 지원한다. 또한 자가 검사, 자동 검사 및 수동 검사 모드로 구성되는 다중 모드를 제공함으로써 집이나 병원 등 다양한 환경에서 사용될 수 있도록 설계되었다. 본 연구에서는 구현된 시스템의 출력 전압 파형과 출력 음압 측정을 통해 정상적으로 동작함을 검증하였다.

Keywords

References

  1. 한국산업안전공단, "순음청력 검사에 관한 지침" 한국산업안전공단, 2006.10.
  2. 김규상 외 3 인, "특수건강진단기관 청력검사기의 음향보정" 대한산업의학회지 제16권 제4호, pp381-389, 2004.
  3. 이동훈외 7 인, "지동차폐가 가능한 웹 기반하 디지털 기도순음청력기기의 개발", 대한이비인후과학회지, 제 50권 10 호, 2007.
  4. 신승원외 4 인, "음압 보정을 통한 이동형 청력 검사 시스템의 구현" 전기학회논문지 제56권 6호, 2007.
  5. British Society of Audiology, "Pure tone air and bone conduction threshold audiometry with and without masking and determination of uncomfortable loudness levels", BSA, Mar. 2004.
  6. G. D. Hillman and J. E. Lane, "Real-Time Determination of IIR Coefficients for Cascaded Butterworth Filters", International Conference on Acoustics, Speech and Signal Processing, pp.1353-1356, 1989. https://doi.org/10.1109/ICASSP.1989.266688
  7. Darkoman, "CWave - Simple C++ Class to Maunipulate Way Files", http://www.codeproject.com/KB/audio-video/CWave.aspx, 2008.
  8. Toby Opferman, "Playing .Wav files using the Windows Multi-Media Libray", http://www.codeproject.com/KB/audio-video/wavefiles.aspx, 2004.
  9. Carlos Antollini, "A set of ADO classes -version 2.2.0", http://www.codeproject.com/KB/dalabase/caaadoclass1.aspx, 2005
  10. David B. Thomas and at. el, "Gaussian Random Number Generators", ACM Computing Surveys, Vol. 39, No.4, 2007.
  11. Seymour Shlien and John Lodge, "Real-Time Generation of Narrow-Band Audio Noise", IEEE Signal Processing Letters Vol. 3 No.4 1996. https://doi.org/10.1109/97.489063
  12. E. Zwicker and H. Fastl, "Psychoaccoustics Fact and Models", Springer Verlag, 1990.
  13. John Tomarakos, "The Relationship of Dynamic Range to Data Word Size in Digital Audio Processing", http://www.audiodesignline.com/showArticIe.jhtmI?articIelD=192200610, 2002.

Cited by

  1. Auto ABLB Audiometry System Supporting One-to-many Model vol.24, pp.6, 2011, https://doi.org/10.4313/JKEM.2011.24.6.519