DOI QR코드

DOI QR Code

Measurement of transmitted vibration to stapes and tympanic membrane by DFMT's vibration in implantable middle ear hearing devices

중이 이식형 보청기에서 DFMT의 진동에 의한 등골 및 고막 방향으로 전달되는 진동력 측정

  • Lee, Myoung-Won (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Seong, Ki-Woong (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Lim, Hyung-Gyu (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Kim, Min-Woo (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Jung, Eui-Sung (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Lee, Jang-Woo (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Kim, Dong-Wook (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University) ;
  • Lee, Jyung-Hyun (Department of Biomedical Engineering, Kyungpook National University Hospital) ;
  • Lee, Sang-Heun (Otolaryngology, School of Medicine, Kyungpook National University) ;
  • Lee, Kyu-Yup (Otolaryngology, School of Medicine, Kyungpook National University) ;
  • Cho, Jin-Ho (Graduate School of Electrical Engineering and Computer Science, Kyungpook National University)
  • 이명원 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 성기웅 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 임형규 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 김민우 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 정의성 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 이장우 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 김동욱 (경북대학교 대학원 전자전기컴퓨터학부) ;
  • 이정현 (경북대학교병원 의공학과) ;
  • 이상흔 (경북대학교 의학전문대학원 이비인후과학교실) ;
  • 이규엽 (경북대학교 의학전문대학원 이비인후과학교실) ;
  • 조진호 (경북대학교병원 의공학과)
  • Published : 2009.07.31

Abstract

The implantable middle ear hearing devices(IMEHDs) have been developed to overcome the conventional hearing aid's problem(ringing effect caused by the acoustic feedback, cosmetic problem, etc.). In the IMEHDs, the vibrating transducer is a key component because its vibration enables to hear for hearing impaired people. The vibrating transducer is implanted on ossicular chain by surgical operation. The coupling status between implanted transducer and ossicular chain has an effect on delivering vibrating force from transducer to stapes. Noninvasive method is required to investigate the output characteristics of IMEHDs after implementation. Recently, emitted sound pressure measuring method of tympanic membrane is proposed to investigate the output characteristics of IMEHDs. However, the relationship between displacement of stapes and sound pressure by tympanic membrane was not cleared. In this paper, displacement of stapes and sound pressure by tympanic membrane were measured using the differential floating mass transducer(DFMT) that implanted on the ossicular chain of the human temporal bone and physical ear model. Through the experiments results, the relationship between displacement of stapes and sound pressure by tympanic membrane was investigated.

Keywords

References

  1. R. L. Goode, M. L. Rosenbaum, and A. J. Maniglia, 'The history and development of the hearing aid', Otlolaryngologic Clinics of North America, vol. 28, pp. 1-6, 1995
  2. A. Chasin, 'Current trends in implantable hearing aids', Trends in Amplification, vol. 2, no. 3, pp. 84-107, 1997 https://doi.org/10.1177/108471389700200302
  3. J. Suzuki, Middle Ear Implant: Implantable Hearing Aids, Karger Inc., Tokyo, 1988
  4. B. S. Song, T. Y. Jung, S. P. Chae, M. N. Kim, and J. H. Cho, 'Proposal and evaluation of vibration transducer with minimal magnetic field interference for use in IME system by in-vitro experiment', IEICE Transactions on Electronics, vol. E85-C, no.6, pp. 1374-1377, 2002
  5. 조진호, 송병섭, 김명남, 원철호, 박세광, 이상흔, '전자 트랜스듀서 방식에 의한 청각 보조용 이식형 인공중이 시스템 설계', 센서학회지, 제6권, 제6호, pp.466-475, 1997
  6. H. H. Kim and D. M. Barrs, 'Hearing aids: A review of what’s new', Otolaryngology-Head and Neck Surgery. vol. 134, no. 6, pp. 1043-1050, 2006 https://doi.org/10.1016/j.otohns.2006.03.010
  7. 조진호, 박일용, 임형규, 윤영호, 성기웅, 정의성, 한지훈, 김민규, 이상흔, '완전 이식형 인공중이의 국내 개발현황', 재활복지, 제10권, 제3호, 2006
  8. M. K. Kim, Y. H. Yoon, I. Y. Park, and J. H. Cho, 'Design of differential electromagnetic transducer for implantable middle ear hearing device using finite element method', Sensors and Actuators A, vol. 130-131, pp. 234-240, 2006 https://doi.org/10.1016/j.sna.2006.01.038
  9. M. K. Kim, I. Y. Park, B. S. Song, and J. H. Cho, 'Fabrication and optimal design of differential electromagnetic transducer for implantable middle ear hearing device', Biosensors and Bioelectronics, vol. 21, no. 11 pp. 2170-2175, 2006 https://doi.org/10.1016/j.bios.2005.10.003
  10. M. Winter, B. P. Weber, and T. Lenarz, 'The use of reverse transfer function (RTF) in the fitting procedure of implantable hearing devices', Proceedings of the 4th International Symposium on Electronic Implants in Otology, 2005
  11. 김동욱, 성기웅, 임형규, 김민우, 이명원, 정의성, 이장우, 이정현, 조진호, '완전 이식형 인공중이용 마이크로폰의 위치 결정을 위한 물리적 귀 모델에서의 음향 반향 현상 분석', 한국재활복지공학회논문지, 제2권, 제1호, pp. 27-33, 2008
  12. W. F. Decraemer, 'Measument, visualization and quantitative analysis of complete three-dimensional kinematical data sets of human and cat middle ear', Middle Ear Mechanics in Research and Otology, World Scientific, Singapore, pp. 3-10, 2004

Cited by

  1. Study on frequency response of implantable microphone and vibrating transducer for the gain compensation of implantable middle ear hearing aid vol.19, pp.5, 2010, https://doi.org/10.5369/JSST.2010.19.5.361