A LONG-TERM FIELD TEST OF A LARGE VOLUME IONIZATION CHAMBER BASED AREA RADIATION MONITORING SYSTEM DEVELOPED AT KAERI

  • Published : 2009.06.30

Abstract

An Area Radiation Monitoring System (ARMS) ionization chamber, which had an 11.8 L active volume, was fabricated and performance-tested at KAERI. Low leakage currents, linearities at low and high dose rates were achieved from performance tests. The correlation coefficients between the ionization currents and the dose rates are 1 at high dose rate and 0.99 at low dose rate. In this study, an integration-type ARMS ionization chamber was tested over a year for an evaluation of its long-term stability at a radioisotope (RI) repository of the Young-gwang nuclear power plant. The standard deviation of dose rate of 1 day data and over a 100-days mean value were 6.2 $\mu$R/h and 2.9 $\mu$R/h, respectively. The fabricated ARMS ionization chamber showed stable performance from the results of the long-term tests. Design and performance characteristics of the fabricated ionization chamber for the ARMS from performance-tests are also addressed.

Keywords

References

  1. Knoll GF. Radiation detection and measurement. 3th ed. New York; John Wiley and Sons, 1999:129-157
  2. Kenneth RK. The dosimetry of ionizing radiation. Vol.2, New York; Academic press, 1987: 169-243
  3. Burlin TE. A general theory of cavity ionization. Br. J. Radiol. 1966;39:727-734 https://doi.org/10.1259/0007-1285-39-466-727
  4. Bohm J. Saturation corrections for plane-parallel ionization chambers. Phys. Med. Biol. 1976;21:754-761 https://doi.org/10.1088/0031-9155/21/5/004
  5. Scott PB, Greening JR. The determination of saturation currents in free-air ionization chambers by extrapolation method. Phys. Med. Biol. 1963;8:51-59 https://doi.org/10.1088/0031-9155/8/1/304
  6. Weinhous MS, Meli A. Determining P$_{ion}$, the correction factor for recombination losses in an ionization chamber. Med. Phys. 1984;11:849-856 https://doi.org/10.1118/1.595574
  7. Piermattei A, Azario L, Arcovito G, Toni M. Ionrecombination correction factor ${k}^{sat}$ for spherical ion chambers irradiated by continuous photon beams. Phys. Med. Biol. 1996;41:1025-1035 https://doi.org/10.1088/0031-9155/41/6/006
  8. IAEA, Technical Report Series No. 277. 1987
  9. Keithley Model 6517A. Electrometer user's manual, http://www.keithley.com
  10. Decampo JA, Beck HL, Raft PD. High pressure argon ionization chamber systems for the measurement of environmental radiation exposure rates. New York, U. S. Atomic Energy Commission, 1972:1-69