DOI QR코드

DOI QR Code

사중극 질량 분석기의 이온소스 오염이 이온전류에 미치는 영향

The Effect of Contamination of Ion Source on Ionic Current of Quadrupole Mass Spectrometer

  • 이규찬 (한국표준과학연구원 진공센터) ;
  • 박창준 (한국표준과학연구원 진공센터) ;
  • 김진태 (한국표준과학연구원 진공센터) ;
  • 오은순 (충남대학교 물리학과) ;
  • 홍기성 (한국표준과학연구원 진공센터) ;
  • 홍승수 (한국표준과학연구원 진공센터) ;
  • 임인태 (한국표준과학연구원 진공센터) ;
  • 윤주영 (한국표준과학연구원 진공센터) ;
  • 강상우 (한국표준과학연구원 진공센터) ;
  • 신용현 (한국표준과학연구원 진공센터)
  • Lee, K.C. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Park, C.J. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Kim, J.T. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Oh, E.S. (Physics Department, Chungnam National University) ;
  • Hong, K.S. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Hong, S.S. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Lim, I.T. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Yun, J.Y. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Kang, S.W. (Vacuum center, Korea Research Institute of Standards and Science) ;
  • Shin, Y.H. (Vacuum center, Korea Research Institute of Standards and Science)
  • 발행 : 2009.05.30

초록

사중극 질량분석기(Quadrupole Mass Spectrometer, QMS) 이온전류의 안정성은 진공공정 가스를 모니터링 하는데 중요한 요소 중 하나이다. 진공챔버에 질소가스를 주입하여 압력을 일정하게 유지하면서 시간에 따른 이온전류의 변화를 모니터링 하였다. 진공챔버는 측정하기 전에 잡음신호를 줄이기 위해 ${\sim}3{\times}10^{-9}\;Torr$ 까지 배기하였고, 두개의 이온소스를 측정했다; 하나는 오염된 것으로 갈색 또는 검은색을 띄고 있고 다른 하나는 새 것이다. 질소 압력 $1{\times}10^{-5}\;Torr$에서, 오염된 이온소스의 이온 전류는 시간이 지남에 따라 더 빨리 감소했다. 대략 5.5 시간이 지난 후, 감소율은 새 것이 ${\sim}46%$이고 오염된 것은 ${\sim}84%$였다. 필라멘트 재질이 이온 전류감소에 미치는 영향을 관찰하기 위해서 텅스텐 선의 반을 산화이트륨($Y_2O_3$)으로 코팅하여 필라멘트를 제작하였다. 유사한 이온전류 감소현상이 재질이 다른 두 필라멘트에서 나타났는데 이것은 필라멘트 재질에 의한 온도의 변화 즉 baking 효과로는 이온전류 감소의 원인을 개선할 수 없다는 것을 의미한다. 전반적으로 이온전류의 감소율은 필라멘트 재질보다 이온소스의 오염과 더 밀접하게 관계되어 있다.

The long term stability of ion current of QMS has been one of key parameters for monitoring gas process in vacuum. The time dependence of ionic current was monitored while the pressure of nitrogen gas was kept at a fixed pressure by introducing the gas into vacuum chamber. The chamber was evacuated to ${\sim}3{\times}10^{-9}\;Torr$ to reduce background signals before the measurement. Two ion sources were tested; one had brownish or black color due to gas contamination and the other one was new, i.e. cleaner. At a nitrogen pressure of $1{\times}10^{-5}\;Torr$, the ionic currents measured by the contaminated ion source decreased faster with time. The decrease rate was respectively ${\sim}46%$ for cleaner one and ${\sim}84%$ for contaminated one after ${\sim}5.5%$ hours. In order to test the effect of filament material on the ion current decrease, we fabricated a tungsten(W) filament which consisted of two parts; one half was made of W and the other was coated with yttria. The similar decrease of ionic currents were shown for the two types of filaments, indicating that slight change of temperature of filament due to material difference i.e. baking effect could not improve the origin of ionic current decrease. Overall the decreasing rate of ionic current is more closely associated with contaminated ion source of QMS rather than its filament materials.

키워드

참고문헌

  1. 박종도, 한국 진공학회지 17(1), 1 (2008) https://doi.org/10.5757/JKVS.2008.17.1.001
  2. L. Lieszkovszky, A. R. Filippelli, and C. R. Tilford, J. Vac. Sci. Technol. A8, 3838 (1990) https://doi.org/10.1116/1.576458
  3. S. Watanabe, H. Oyama, S. Kato, and M. Aono, American Institute of Physics. 70, 1880 (1999)
  4. K. Jousten, Handbook of vacuum technology, (WILEY-VCH, 2008), p. 631-662
  5. 홍승수, 임인태, 김진태, 한국진공학회지 16(2), 91 (2007) https://doi.org/10.5757/JKVS.2007.16.2.091
  6. J. A. Basford, M. D. Boeckmann, R. E. Ellefson, A. R. Filippelli, D. H. Holkeboer, L. Lieszkovszky, and C. M. Stupak, J. Vac. Sci. Technol. A11, A22 (1993) https://doi.org/10.1116/1.4755937
  7. James E blessing, J. Vac. Sci. Technol. A25(1), 167 (2007) https://doi.org/10.1116/1.2364001
  8. James A. Basford, J. Vac. Sci. Technol. A9, 1991 (1991)
  9. H. Yoshida, Stability test for Quadrupole mass spectrometer(QMS) in NMIJ (2009), in private communication
  10. 박창준, 한국진공학회지 17(4), 262 (2008) https://doi.org/10.5757/JKVS.2008.17.4.262
  11. Fu Ming Mao and J. H. Leck, Vacuum. 37, 669 (1987) https://doi.org/10.1016/0042-207X(87)90053-4
  12. N. Takahashi, T. Hayashi, H. Akimichi, and Y. Tuzi, J. Vac. Sci. Technol. A19, 1688 (2001) https://doi.org/10.1116/1.1368666
  13. R. J Ferran and S. Boumsellek, J. Vac. Sci.Technol. A14, 1258 (1996) https://doi.org/10.1116/1.579938