Quantitative Analysis for Digital Intraoral Sensor

구강 내 촬영용 센서의 정량적 평가

  • Jung, Jae-Ho (Department of Electronic Engineering, Kwang Woon University) ;
  • Jung, Jae-Eun (Department of Digital Management and Information, Nambu University) ;
  • Kang, Hee-Doo (Department of Electronic Engineering, Kwang Woon University) ;
  • Ra, Keuk-Hwan (Department of Electronic Engineering, Kwang Woon University)
  • 정재호 (광운대학교 전자공학과) ;
  • 정재은 (남부대학교 디지털경영정보학과) ;
  • 강희두 (광운대학교 전자공학과) ;
  • 나극환 (광운대학교 전자공학과)
  • Published : 2009.11.30

Abstract

Carried out an experiment for quantitative evaluation about intraoral sensor among oral dental imaging equipments in PACS environment. For evaluation, testing environment & evaluation criteria are established that refer to the 'IEC 62220-1-2' and experimented with set up the standard radiation penetration that correspond with RQA-3(IEC 61267). Results of experiments using the image J, derived the ESF(edge spread function), LSF (line spread function) and calculated the MTF(modulation transfer function) finally. As a result, the MTF that 0.1, 0.2sec are 10% about 10 lp/mm and 0.32 sec is the 10% about 9 lp/mm in level represents the value of the MTF. Change of scan condition in dental environmental, according to the MTF value taken note that no changes can be seen enough. However, the dimensions of each other size $1(1200{\times}1600)$, size $2(1440{\times}1920)$ intraoral sensors for 3 lp/mm, respectively 40%, 90% of the note might have been the difference between the value of MTF, in accordance with standard sensors might note differences could be observed.