Inter-comparison of temperature measurement capability using standard platinum resistance thermometers

표준백금저항온도계를 이용한 온도측정능력 상호비교

  • Gam, K.S. (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Kang, C.S. (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Lee, Y.J. (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Lee, K.B. (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Kim, Y.G. (Division of Physical Metrology, Korea Research Institute of Standards and Science) ;
  • Park, S.N. (Division of Physical Metrology, Korea Research Institute of Standards and Science)
  • Published : 2009.01.31

Abstract

Temperature measurement capability was inter-compared using the transfer standard platinum resistance thermometers(SPRT) among four laboratories of KRISS. The transfer SPRTs were primarily calibrated at the triple point of water and Ga melting point, then used at inter-comparison experiment. Temperature difference of calibration value between temperature laboratory and length laboratory at $20^{\circ}C$ was -0.7 mK and +2.4 mK at density laboratory. Temperature measured near $20^{\circ}C$, $25^{\circ}C$ and $30^{\circ}C$ at fluid flow laboratory was deviated by $34.2{\sim}80.4\;mK$ from the calibration values of the transfer SPRT. Ga melting points was inter-compared among three laboratories, and the difference of Ga melting points against the standard Ga melting point of temperature laboratory were $0.03{\sim}0.54\;mK$ at length laboratory and 0.02 mK at density laboratory.

Keywords

References

  1. H. Preston-Thomas, 'The international temperature scale of 1990 (ITS-90)', Metrologia, vol. 27, pp. 3-10, 1990 https://doi.org/10.1088/0026-1394/27/1/002
  2. 안종찬, 감기술, 이영희, 김선곤, 윤병수, '국가교정 검사기관 측정능력 평가(온도분야)', KRISS-98-097- IR, KASTO98-014, KRISS, 1998
  3. 감기술, 유성호, 김성민, 이인식, '백금저항온도센서 를 이용한 국가교정기관의 교정능력 평가 및 실험방 법', 센서학회지, 제14권, 제4호, pp. 231-236, 2005 https://doi.org/10.5369/JSST.2005.14.4.231
  4. Xumo Li, D.Farley, Deming Chen, Mingjian Zhao, and M.Hirst, 'Mini metal-cased fixed-point cells', Proc. of 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, pp. 777-782, Berlin, Germany, 2001
  5. G. F. Strouse, 'Small fixed-point cells for use in dry well block calibrators', Proc. of 8th International Symposium on Temperature and Thermal Measurements in Industry and Science, pp. 783-788, Berlin, Germany, 2001

Cited by

  1. Development of a multi channel measurement system for the cellular respiration measurement vol.19, pp.1, 2010, https://doi.org/10.5369/JSST.2010.19.1.036