The Research for Scannogram

Scannogram, Whole Spine image 획득에 관한 연구

  • Published : 2008.05.30

Abstract

I.objective The composition method with acquiring 2${\sim}$5 image not only lack perfection in compensating the geometrical distortion but could also cause inaccurate connection problem of compensating the geometrical distortion but could also cause inaccurate connection problem of continuing bone structure due to the enlargement ratio difference of ruller and bone, density difference in image composition process. This paper studies those mentioned problems through the experiments. II. Object and method Experiment 1 After attaching 3 step wedge(Pb) on detector and 2 ruller on upper, lower part of the step wedge, we have exposed them to create density difference between 1st image and 2nd image, then examined if there were any errors. Experiment 2 1st 100cm ruller was attached on the center of detector, and by escalating the distance between 2nd 100cm ruller and detector 5cm, 10cm, 15cm respectively, we investigated if there any errors caused by enlargement ratio. Experiment 3 The ruller was placed on detector, a joint photographing was performed through spot photographing after fully operating the electric field photographing to include hip joint ${\sim}$ ankle joint part and we have compared the values from two methods. III. Result : A horizontal axis error was caused when photopraphing is not poerated in equal density since the image shifted horizontally in accordance with the difference of Pb step wedge 1 due to the density difference, as the result of Epperiment 1.

Keywords