A Compatibility Assessment and Verification of Suitable to DICOM of PACS DATA CD : Current Situation Investigation of Korea

PACS DATA CD의 호환성 평가 및 DICOM 적합성에 대한 검증을 통한 기준 제시

  • 정재호 (경희의료원 의학정보센터 PACS 팀) ;
  • 성동욱 (경희대학교 의과대학 영상의학과 교실) ;
  • 박범진 (경희의료원 의학정보센터 PACS 팀) ;
  • 손기경 (경희의료원 의학정보센터 PACS 팀) ;
  • 강희두 (경희의료원 의학정보센터 PACS 팀)
  • Published : 2008.05.30

Abstract

Purpose To analyze the input and output error of data CD which records the image information and the problems of the server of the compatibility. And to report a compatibility assessment and verification of suitable to DICOM of PACS data CD with investigation of current situation of Korea METHOD AND MATERIALS Date CD of each 8 vendors in 30 hospitals was analyzed. We grasped a main verification element existence of a generation compatibility of data CD. The items of element are media identification, DICOM compression, DICOM viewer send, specified object information modify, auto-run, DICOM content type, etc, and give 1 point for each item. We divided the assessment about an each item into 5 levels. Verification about. DICOM conformance by using DICOM validation tool kit is shown to be classified pass or fail according to error occurrence of tag valus. Classify the prequency of tag occurrence as the item. RESULTS The average point of date CD compatibility is 8 point (very good), lowest is 5 point (6.6%), and highest is 10 point (23%_. Most high occurrence frequency's distribution is 7 point (36.6%). As a result of verification about DICOM conformance, PASS in 8 occurrence frequency's distribution is 7 point (36.6%). As a result of verification about DICOM maximum length numbers (14 items), DICOM error of modality (10 items), discord of pixel data length (6 items). etc.