Measurements of Two-dimensional Gratings Using a Metrological Atomic Force Microscope with Uncertainty Evaluation

  • Kim, Jong-Ahn (Length/Time Group, Korea Research Institute Standards and Science) ;
  • Kim, Jae-Wan (Length/Time Group, Korea Research Institute Standards and Science) ;
  • Kang, Chu-Shik (Length/Time Group, Korea Research Institute Standards and Science) ;
  • Eom, Tae-Bong (Length/Time Group, Korea Research Institute Standards and Science)
  • 발행 : 2008.04.01

초록

The pitch and orthogonality of two-dimensional (2-D) gratings were measured using a metrological atomic force microscope (MAFM), and the measurement uncertainty was analyzed. Gratings are typical standard devices for the calibration of precision microscopes, Since the magnification and orthogonality in two perpendicular axes of microscopes can be calibrated simultaneously using 2-D gratings, it is important to certify the pitch and orthogonality of such gratings accurately for nanometrology. In the measurement of 2-D gratings, the MAFM can be used effectively for its nanometric resolution and uncertainty, but a new measurement scheme is required to overcome limitations such as thermal drift and slow scan speed. Two types of 2-D gratings with nominal pitches of 300 and 1000 nm were measured using line scans to determine the pitch measurement in each direction. The expanded uncertainties (k = 2) of the measured pitch values were less than 0.2 and 0.4 nm for each specimen, and the measured orthogonality values were less than $0.09^{\circ}$ and $0.05^{\circ}$, respectively. The experimental results measured using the MAFM and optical diffractometer agreed closely within the expanded uncertainty of the MAFM. We also propose an additional scheme for measuring 2-D gratings to increase the accuracy of calculated peak positions, which will be the subject of future study.

키워드

참고문헌

  1. Teague, E. C., "Nanometrology," Proc. AIP Conference on Scanned Probe Microscopy, Vol. 241, pp. 371-407, 1991
  2. Carneiro, K., "The need for metrology in nanotechnology," Danish Institute of Fundamental Metrology, pp. 1-5, 2001
  3. Postek, M. T., Vladar, A. E., Jones, S. N. and Keery, W. J., "Interlaboratory study on the lithographically produced scanning electron microscope magnification standard prototype," Journal of Research of the National Institute of Standards and Technology, Vol. 98, No. 4, pp. 447-467, 1993 https://doi.org/10.6028/jres.098.033
  4. Nakayama, Y., Okazaki, S. and Sugimoto, A., "Proposal for a new submicron dimension reference for an e-beam metrology system," Journal of Vacuum Science and Technology B, Vol. 6, No. 6, pp. 1930-1933, 1988 https://doi.org/10.1116/1.584135
  5. Meli, F., Thalman, R. and Blattner, P., "High precision pitch calibration of gratings using laser diffractometry,"First International Conference and General Meeting of the Euspen Society for Precision Engineering and Nanometrology, pp. 252-255, 1999
  6. Kim, J. A., Kim, J. W., Park, B. C., Kang, C. S. and Eom, T. B., "Measurement of grating pitch standards using optical diffractometry and uncertainty analysis," Journal of the Korean Society for Precision Engineering, Vol. 23, No. 8, pp. 72-79, 2004
  7. Meli, F. and Thalmann, R., "Long-range AFM profiler used for accurate pitch measurements," Measurement Science and Technology, Vol. 9, No. 7, pp. 1087-1092, 1998 https://doi.org/10.1088/0957-0233/9/7/014
  8. Misumi, I., Gonda, S., Kurosawa, T. and Takamasu, K., "Uncertainty in pitch measurements of one-dimensional grating standards using a nanometrological atomic force microscope," Measurement Science and Technology, Vol. 14, No. 4, pp. 463-471, 2003 https://doi.org/10.1088/0957-0233/14/4/309
  9. Kim, J. A., Kim,J. W., Park, B. C., Eom, T. B. and Hong, J. W., "Development of a metrological atomic force microscope for the length measurements of nanometer range," Journal of the Korean Society for Precision Engineering, Vol. 21, No. 11, pp. 75-82, 2004
  10. Kim, J. A., Kim,J. W., Park, B. C., Eom, T. B. and Kang, C. S., "Pitch measurement of one-dimensional gratings using a metrological atomic force microscope and uncertainty evaluation," Journal of the Korean Society for Precision Engineering, Vol. 22, No. 4, pp. 84-91, 2005
  11. BIPM, IEC, IFCC, ISO, IUPAC, IUPAP and OIML, "Guide to the Expression of Uncertainty in Measurement," ISO, 1993