A Non-invasive Measurement of Abdominal Pressure on Ambulatory Urodynamics Study Using Surface Electromyography

휴대용 요역동학 검사 시 근전도 신호를 이용한 복압측정 방법

  • Published : 2008.11.25

Abstract

Conventional rectal catheters which are used for the measurement of abdominal pressure can cause erroneous result affecting detrusor pressure, and the catheter itself is not comfortable to the patients. To reduce these problems, we invented a new method for measuring abdominal pressure in non invasive manner using surface electromyography (EMG) signals of the rectus abdominis muscle. Our results showed that the correlation coefficient and root mean square error (RMSE) between the measured abdominal pressures by the conventional rectal catheters and the estimated values by our proposed algorithm were $0.79{\pm}0.06$ and $0.10{\pm}0.07$, respectively. These findings suggest that the surface EMG of rectus abdominis muscle might be used indirectly for more convenient measurement of abdominal pressure on ambulatory urodynamic study.

복압 측정을 위한 기존의 카데터 방법은 배뇨압에 영향을 줄 수 있는 많은 오차를 유발할 수 있으며 환자에게 불편함을 야기한다. 이러한 문제점을 해결하기 위해, 본 연구에서는 표면 근전도를 이용하여 비침습적으로 복압을 측정할 수 있는 새로운 방법을 개발하였다. 연구 결과, 기존 직장 카테터를 이용하여 측정된 복압과 복압측정 알고리즘을 이용하여 추정된 복압 간의 상관계수는 $0.79{\pm}0.06$이었으며, 두 신호간의 오차는 $0.10{\pm}0.07$이었다. 본 연구에서 개발된 복압추정 알고리즘을 이용하면 휴대용 요역동학 검사 시 편리하게 간접적으로 복압을 측정할 수 있을 것으로 생각된다.

Keywords

References

  1. W. Schafer, P. Abrams, L. Liao, A. Mattiasson, F. Pesce, A. Spangberg, A. M. Sterling, N. R. Zinner and P. Van Kerrebroeck, 'Good urodynamic practices: Uroflowmetry, filling cystometry and pressure-flow studies', Neurourol. Urodyn., Vol. 21, pp. 261-274, 2002 https://doi.org/10.1002/nau.10066
  2. P. C. Ridings, G. L. Bloomfield, C. R. Blocher, 'Cardiopulmonary effects of raised intra- abdominal pressure before and after intravascular volume expansion', J. Trauma, Vol. 39, pp. 1071-1075, 1995 https://doi.org/10.1097/00005373-199512000-00010
  3. J. H. Seo, Y. C. Jeong, C. G. Song, M. H. Ko, S. H. Park, 'Change of intra-abdominal pressure on daily activities in patients with spinal cord injury', J. Korean Continence Society, Vol. 6, pp. 54-61, 2002 https://doi.org/10.5213/jkcs.2002.6.2.54
  4. A. MacDonald, P. J.Paterson, J. N. Baxter, I. G. Finlay, 'Relationship between intra-abdominal pressure and intra-rectal pressure in the proctometrogram', Br. J. Surg., Vol. 80, pp. 1070-1071, 1993 https://doi.org/10.1002/bjs.1800800852
  5. M. L. N. G. Malbrain, 'Different technique to measure intra-abdominal pressure(IAP)', Intensive Care Med., Vol. 30, pp. 357-371, 2004 https://doi.org/10.1007/s00134-003-2107-2
  6. E. W. Van Doorn, K. Anders, V. Khullar, S. Kulsenh-Hanssen, F. Pesce, A. Robertson, D. Rosario, W. Schafer, 'Standardisation of ambulatory urodynamic monitoring : Report of the standardisation sub-committee of the international continence society for ambulatory urodynamic studies', Neurourol. Urodyn. Vol. 19, pp. 113-125, 2000 https://doi.org/10.1002/(SICI)1520-6777(2000)19:2<113::AID-NAU2>3.0.CO;2-#
  7. P. Neumann and V. Gill, 'Pelvic floor and abdominal muscle interaction: EMG activity and intra-abdominal pressure', Int. Urogynecol. J., Vol. 13, pp. 125-132, 2002 https://doi.org/10.1007/s001920200027
  8. F. Demaria, R. Porcher, S. S. Ismael, G. Amarenco, X. Fritel, P. Madelenat and J. L. Benifla, 'Using intercostals muscle EMG to quantify maternal expulsive efforts during vaginal delivery : A pilot study', Neurourol. Urodyn., Vol. 23, pp. 675-678, 2004 https://doi.org/10.1002/nau.20050
  9. G. Amarenco, S. S. Ismael, D. Lagauche, P. Raibaut, P. Rene-Corail, N. Wolff, P. Thoumie and F. Haab, 'Couch anal reflex: Strict relationship between intravesical pressure and pelvic floor muscle electromyography activity during cough', Urol., Vol. 173, pp. 149-152, 2005 https://doi.org/10.1097/01.ju.0000147305.00443.df
  10. J. Finsterer, 'EMG-interference pattern analysis', J. Electromyogr. Kinesiol., Vol. 11, pp. 231-246, 2001 https://doi.org/10.1016/S1050-6411(01)00006-2
  11. W. S. Cobb, J. M. Burns, K. W. Kercher, B. D. Matthews, H. J. Norton and T. H. Heniford, 'Normal intra-abdominal pressure in healthy adults', J.Surg. Res., Vol. 129, pp. 231-235, 2005 https://doi.org/10.1016/j.jss.2005.06.015
  12. E. Katsavrias. E. Primetis, N.Karandreas, 'A new electromyographic technique and normal motor unit action potential values', Clin. Neurophysiol., Vol. 116, pp. 2528-2532, 2005 https://doi.org/10.1016/j.clinph.2005.07.020
  13. J. H. Seo, K. M. Kim, M. H. Ko, S. H. Park, N. G. Kim, 'Motor unit action potentials of the rectus abdominis muscle according to changes of abdominal pressure', J. Korean Acad. Rehab. Med., Vol. 29, pp. 457-461, 2005
  14. C. Nordander, J. Willner, G. A. Hansson, B. Larsson, J. Unge, L. Granquist, S. Skerfving, 'Influence of the subcutaneous fat layer, as measured by ultrasound, skinfold calipers and BMI, on the EMG amplitude', Eur. J. Appl. Physiol., Vol. 89, pp. 514-519, 2003 https://doi.org/10.1007/s00421-003-0819-1