Investigation of the Regression Analysis Method for a Quantitative Evaluation of Implant Crestal Bone Stresses

회귀분석법에 의한 임플란트 경부골 응력의 정량적 분석에 대한 연구

  • Kim, Woo-Shik (Department of Prosthodontics, School of Dentistry, Kyungpook National University) ;
  • Jo, Kwang-Hun (Department of Prosthodontics, School of Dentistry, Kyungpook National University) ;
  • Lee, Kyu-Bok (Department of Prosthodontics, School of Dentistry, Kyungpook National University)
  • 김우식 (경북대학교 치의학전문대학원 치과보철학교실) ;
  • 조광헌 (경북대학교 치의학전문대학원 치과보철학교실) ;
  • 이규복 (경북대학교 치의학전문대학원 치과보철학교실)
  • Received : 2008.06.21
  • Accepted : 2008.09.25
  • Published : 2008.09.30

Abstract

In this study, the regression analysis method was tested for the estimation of peak stress at stress concentration area in the cervical bone. Submerge type EZ plus implant (Megagen. Daegu, Korea), 4.1 mm in cervical diameter and 9.6 mm in endosseous length, were axisymmetrically modelled together with surrounding alveolar bone of which the width was 10 mm. Vertical force of 100 N was applied to a head of crown above 8.5 mm from the outer surface of the cortical bone. Four different mesh models were composed of differently sized elements in vicinity of sharp corners, and they include 6 stress monitoring points that are located in the same geometrical points regardless of the differences in the meshes. Primary consideration was given to the stresses in the cortical bone surrounding the implant neck. The results showed that virtually all the stresses were concentrated in the cortical bone regardless of mesh designs. The peak stresses were successfully calculated by a regression analysis in a stable manner, as far as the mesh is designed to represent the acute gradient of stresses near the sharp corner.

경부골 최대응력을 안정적으로 산출하기 위해서, 경부 디자인 형상이 비교적 단순한 메가젠 임플란트 모델을 대상으로, 유한요소 모델링을 통하여 경부골의 응력분포를 계산하였다. 그리고 산출된 응력분포를 회귀분석법으로 통계 처리하여 응력 집중점의 응력을 정량화 하였다. 결과는 모든 모델에서 경부골 응력은 치밀골 외면에 집중되었으며, 그 범위는 0.5mm 이내였다. 회귀분석법을 활용하기 위해서는 mesh 정밀도가 높아야 했으며, 이 경우 특이점의 최대 응력을 안정적으로 산출 할 수 있었다.

Keywords

References

  1. Jung ES, Jo KH, Lee CH. A finite element stress analysis of the bone around implant following cervical bone resorption. KAID 2003;22:38-47
  2. Kitamura E, Stegaroiu R, Nomura S, Miyakawa O. Influence of marginal bone resorption on stress around an implant -a three-dimensional finite element analysis. J Oral Rehabil 2005;32:279-86 https://doi.org/10.1111/j.1365-2842.2004.01413.x
  3. Borchers L, Reichart P. Three-dimensional stress distribution around a dental implant at different stages of interface development. J Dental Res 1983;62:155-9 https://doi.org/10.1177/00220345830620021401
  4. Papavasiliou G, Kamposiora P, Bayne SC. Three-dimensional finite element analysis of stress-distribution around single tooth implants as afunction of bony support, prosthesis type, and loading during function. J Prosth Dent 1996;67(6):633-40
  5. Kalandiia AI. Remarks on the singularity of elastic solutions near corners. J Appl Math Mech 1969;33:127-31 https://doi.org/10.1016/0021-8928(69)90121-X
  6. Murakami S, Liu Y. Mesh-Dependence in Local Approach to Creep Fracture. Int J Damage Mech 1995;4:230-50 https://doi.org/10.1177/105678959500400303
  7. Bozkaya D, Muftu S, Muftu A. Evaluation of load transfer characteristics of five different implants in compact bone at different load levels by finite elements analysis. J Prosth Dent 2004;92:523-30 https://doi.org/10.1016/j.prosdent.2004.07.024
  8. NISA II / DISPLAY III User's Manual, Engineering Mechanics Research Corporation (EMRC), IL, U.S.A
  9. Boyer RR. Aerospace Applications of Beta Titanium Alloys (Overview). JMet. 1994:20-3
  10. Jang YJ. Influence of the crestal bone width and the diameter of implant on the bone stresses: A Finite Element Approach, MD thesis, Department of Dentistry, Kyungpook National Univ 2006
  11. Chung JM. Finite element analysis of the peri-implant bone stress influenced by cervical module configuration of endoosseous implant: A Finite Element Approach, MD thesis, Department of Dentistry, Kyungpook Nat Univ 2006
  12. Sevimay M, Turhan F, Kilicarslan MA, Eskitascioglu G. Three-dimensional finite element analysis of the effect of different bone quality on stress distribution in an implant-supported crown. J Prosth Dent 2005;93:227-34 https://doi.org/10.1016/j.prosdent.2004.12.019
  13. Natali AN, Pavan PG, Ruggero AL. Analysis of bone-implant interaction phenomena by using a numerical approach. Clin Oral Implants Res 2006;17:67-74 https://doi.org/10.1111/j.1600-0501.2005.01162.x
  14. Petrie CS, Williams JL. Comparative evaluation of implant designs: influence of diameter, length, and taper on strains in the alveolar crest. A three-dimensional finite-element analysis. Clin Oral Implants Res 2005;16:486-94 https://doi.org/10.1111/j.1600-0501.2005.01132.x
  15. Matsushita Y, Kitoh M, Mizuta K, Ikeda H, Suetsugu T. Two-dimensional FEM analysis of hydroxyapatite implants: diameter effects on stress distribution. J Oral Implantol 1990;16:6-11
  16. Tada S. Stegaroiu R, Kitamura E, Miyakawa O, Kusakari H. Influence of implant design and bone quality on stress/strain distribution in bone around implants: a 3-dimensional finite element analysis. Int J Oral Maxillofac Implants 2003;18:357-68
  17. Kitagawa T, Tanimoto Y, Nemoto K, Aida M. Influence of cortical bone quality on stress distribution in bone around dental implant. Dent Mater J 2005;24:219-24 https://doi.org/10.4012/dmj.24.219
  18. Chun HJ, Cheong SY, Han JH, Heo SJ, Chung JP, Choi YC, Rhyu IC, Kim MH. Effects of Design Parameters of Osseointegrated Implant on Stress Distribution in Jaw Bone. Proceedings of the 22nd Annual EMBS International Conference, July 23-28, 2000, Chicago IL. USA
  19. Holmgren, E.P., Seckinger, R.J., Kilgren, L.M. Mante, F. Evaluating parameters of osseointegrated dental implants using finite element analysis - a two-dimensional comparative study examining the effects of implant diameter, implant shape, and load direction. J Oral Implantol 1998;24:80-8 https://doi.org/10.1563/1548-1336(1998)024<0080:EPOODI>2.3.CO;2
  20. 임종원, 김왕식, 김일규, 손충렬, 변효인. 교정용 미니스크류 식립 시 스크류의 길이, 직경 및 피질골 두께에 따른 응력 분포에 관한 3차원 유한요소법적 연구. 대한치과교정학회지 2003;33:11-20