Improvement of Convection by Pulsatile Blood Pump in High Flux Hemodialysis System

고 유량 혈액 투석 시스템에서 박동류 펌프에 대한 대류 향상에 관한 연구

  • Choi, D.S. (Department of Biomedical Engineering, College of Medicine, Seoul National University) ;
  • Um, K.M. (Department of Mechanics & Biomedical Engineering, Kangwon National University) ;
  • Lee, J.C. (Institute of Medical and Biological Engineering, Medical Research Center, Seoul National University) ;
  • Lee, S.R. (Korea Artificial Organ Center, Korea University) ;
  • Mun, C.H. (Korea Artificial Organ Center, Korea University) ;
  • Choi, H. (Korea Artificial Organ Center, Korea University) ;
  • Min, B.G. (BHK. CO., Ltd.) ;
  • Kim, H.C. (Department of Biomedical Engineering, College of Medicine, Seoul National University)
  • 최대석 (서울대학교 의과대학 의공학 교실) ;
  • 임기무 (강원대학교 기계의용공학과) ;
  • 이정찬 (서울대학교 의과대학 의학연구원) ;
  • 이사람 (한국인공장기센터, 고려대학교) ;
  • 문초혜 (한국인공장기센터, 고려대학교) ;
  • 최혁 (한국인공장기센터, 고려대학교) ;
  • 민병구 ((주)BHK) ;
  • 김희찬 (서울대학교 의과대학 의공학 교실)
  • Published : 2008.10.31

Abstract

High-flux dialysis treatment removes various toxins via diffusion as well as convection, which is induced by ultrafiltration and backfiltration. In this study, in vitro (Using the distilled water and the bovine's blood) comparison test was performed to determine whether utilization of a high flux dialyzer paired with different pumps would increase the efficiency of convection. At the same blood flow rates, a pulsatile pump and a roller pump were employed to propel the distilled water and bovine whole blood to a high flux dialyzer. Pressures at the dialyzer inlet and outlet in the blood circuit and in the dialysate circuit were measured, respectively. From these data, we calculated the transmembrane pressure and predicted the ultrafiltration and backfiltration rates developed by both pumps. Using the bovine's blood experiment, ultrafiltration and backfiltration rates were 1.6 times higher with the pulsatile pump than with the roller pump. We conclude that utilization of a pulsatile pump in high flux hemodialysis treatments increases ultrafiltration volume, compared with a roller pump under conditions of the same blood flow rate.

Keywords

References

  1. John F. MAHER, Replacement of Renal Function by Dialysis, Holland, Nederland : Kluwer Academic Publishers, 2004, pp 300- 346
  2. Juan P. Bosch, Jay H. Stein, Hemodialysis(High-Efficiency treatments), New York, USA: Churchill Livingstone Inc.,1993, pp 105-133
  3. Young-Suk Yoon, "Principle of Hemodialysis," The Korean Society of Nephrology., vol. 7, no.3, pp. 1-7, 1988
  4. TM. Runge, JC. Briceno, ME. Sheller, et al,"Hemodialysis: evidence of enhanced molecular clearance and ultrafiltration volume by using pulsatile flow," Int J Artif Organs., vol. 16, no. 9, pp. 645-52, 1993 https://doi.org/10.1177/039139889301600904
  5. K.M. Lim, J.Y. Park, J.C. Lee, et al, "Quantitative Analysis of Pulsatile Flow Contribution to Ultrafiltration," Artificial Organs, accepted 18 Apr 2008
  6. K. Lee, J.H. Jeong, C.H. Mun, et al, "The effect of pulsatile flow on ultrafiltration : In-vitro study and comparison with roller pump," J. Biomed. Eng. Res., vol. 26, no. 5, pp. 337-341, 2005
  7. C. Ronco, PM. Ghezzi, A. Brendolan, et al, "The haemodialysis system: basic mechanisms of water and solute transport in extracorporeal renal replacement therapies," Nephrology Dialysis Transplantation., vol. 13, no. 13, pp.3-9, 1998
  8. C. Ronco, G. Orlandini, A. Brendolan, et al, "Enhancement of convective transport by internal filtration in a modified experimental hemodialyzer: technical note," Kidney Int ., vol. 54, no. 3, pp. 979-85, 1998 https://doi.org/10.1046/j.1523-1755.1998.00049.x
  9. M. Mineshima, I. Ishimori, K. Ishida, et al. "Effects of internal filtration on the solute removal efficiency of a dialyzer," ASAIO Journal., vol. 46, no. 4, pp.456-460, 2000 https://doi.org/10.1097/00002480-200007000-00018
  10. J.C. Lee, J. Chung, Y.R. Rho, et al, "Clinical applications and recent progress of therapeutic artificial organ using movingactuator mechanisms," International Journal of Applied Electromagnetics and Mechanics., vol.27, no. 3, pp. 205-214, 2008
  11. M. Ruperez, C. Sanchez, C. Garcia, et al. "Continuous venovenous renal replacement therapy using a pulsatile blood pump". Pediatr Nephrol., vol. 18, pp. 29-32, 2003 https://doi.org/10.1007/s00467-002-0991-2
  12. M. Ruperez, JL. Herce, C. Sanchez, et al. "Comparison of a tubular pulsatile pump and a volumetric pump for continuous venovenous renal replacement therapy in a pediatric animal model." Asaio J., vol. 51, pp. 372-375, 2005 https://doi.org/10.1097/01.mat.0000169112.40062.0e
  13. JL. Herce, M. Ruperez, C. Sanchez, et al. "Continuous venovenous renal replacement therapy with a pulsatile tubular blood pump: analysis of efficacy parameters." Artif Organs., vol. 30, pp. 64-69, 2006 https://doi.org/10.1111/j.1525-1594.2006.00181.x
  14. S. Troyanov, J. Cardinal, D. Geadah, et al. "Solute clearances during continuous venovenous haemofiltration at various ultrafiltration flow rates using Multiflow-100 and HF1000 filters." Nephrology Dialysis Transplantation., vol. 18, pp. 961-966, 2003 https://doi.org/10.1093/ndt/gfg055
  15. Z. Ricci, C. Ronco, A. Bachetoni, et al. "Solute removal during continuous renal replacement therapy in critically ill patients: convection versus diffusion." Critical Care., vol. 10, pp. 1-7, 2006
  16. S. Brunet, M. Leblanc, D. Parent, et al. "Diffusive and convective solute clearances during continuous renal replacement therapy at various dialysate and ultrafiltration flow rates." American Journal of Kidney Diseases., vol. 34, pp. 486-492, 1999 https://doi.org/10.1016/S0272-6386(99)70076-4