SURFACE CHANGE AND FIT OF TIN-COATED ABUTMENT SCREW AFTER REPEATED CLOSING AND OPENING

TiN 코팅된 지대주 나사의 반복 착탈 후 표면 변화와 적합도에 관한 연구

  • Kim, Jong-Nam (Department of Dentistry, Graduate School of Chosun University) ;
  • Chung, Chae-Heon (Department of Dentistry, Graduate School of Chosun University) ;
  • Kim, Hee-Jung (Department of Dentistry, Graduate School of Chosun University)
  • 김종남 (조선대학교 치과대학 보철학교실) ;
  • 정재헌 (조선대학교 치과대학 보철학교실) ;
  • 김희중 (조선대학교 치과대학 보철학교실)
  • Published : 2007.02.28

Abstract

Statement of problem: A few dry lubricants were applied to abutment screws for the improvement of joint stability. Purpose: The purpose of this study was to evaluate the surface change and fit of TiN-Coated abutment screw through the examination of tested screws in the field emission scanning electron microscope(FE-SEM;Netherland, Phillips co., model:XL 30 SFEG) after repeated closing and opening. Materials and method: Titanium(3i/implant Innovations Inc, USA) and Gold-Tite abutment screws(3i/implant Innovations Inc, USA) were selected for Group A and C respectively. TiN coated titanium abutment screws were also divided into two groups, Group B and D. Abutment screws of each group and the fit of abutment screw/implant fixture/abutment were observed on FE-SEM after repeated closing and opening test respectively. Results: 1. The abutment screws of TiN coated groups(Group B and D) showed more remarkable wear resistance in the threads of the screw than those of the other group(Group A and C). 2. There were more severe wear and defect of TiN coating in Group D tightened to 32 Ncm than Group B to 20 Ncm. 3. There was no difference in the fit of abutment screw/fixture/abutment among four groups, Group A, Group B, Group C and Group D. Conclusion: Under the conditions of this study, it is suggested that TiN coating of abutment screw be clinically acceptable and be expected to reduce the risk of abutment screw loosening. TiN coating of abutment screw showed good resistance against wear and the adequate fit of abutment screw/implant fixture/abutment.

Keywords

References

  1. Taylor. Prosthodontic problems and limitations associated with osseointegration. J Prosthet Dent 1998;79:74-78 https://doi.org/10.1016/S0022-3913(98)70197-0
  2. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent 1993;21:203-208 https://doi.org/10.1016/0300-5712(93)90127-C
  3. Goodacre CJ, Bernal G, Rungcharassaeng K, Kan JY. Clinical complications with implants and implant prostheses. J Prosthet Dent 2003;90:121-132 https://doi.org/10.1016/S0022-3913(03)00212-9
  4. Binon PP, Sutter F, Beaty K, Brunski J, Gulbransen H, Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants (supplement) 1994;9:8-63
  5. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofacimplants 1995;10:529-536
  6. Motosh N. Development of design charts for bolts preloaded up to the plastic range. J Eng Ind 1976;98:849-851 https://doi.org/10.1115/1.3439041
  7. Cavazos E, Bell FA. Preventing loosening of implant abutment screws. J Prosthet Dent 1996;75:566-569 https://doi.org/10.1016/S0022-3913(96)90464-3
  8. Martin WC, Woody RD, Miller BM, Miller AW. Implant abutment screw rorations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32 https://doi.org/10.1067/mpr.2001.116230
  9. Drago CJ. A clinical study of the efficacy of gold-tite square abutment screws in cement- retained implant restorations. Int J Oral Maxillofac Implants 2003;18:273-278
  10. Kim SK, Lee JB, Koak JY, Heo SJ, Lee KR, Cho LR, Lee SS. An abutment screw loosening study of a Diamond Like Carboncoated CP titanium implant. J Oral Rehabil 2005;32:346-50 https://doi.org/10.1111/j.1365-2842.2004.01475.x
  11. Choi JU, Jeong CM, Jeon YC, Lim JS, Jeong HC, Eom TG. Influence of Tungsten Carbide/Carbon coating on the preload of implant abutment screws. J Korean Acad Prosthodont 2006;44:229-242
  12. Mezger PR, Creugers NH. Titanium nitride coatings in clinical dentistry. J Dent 1992;20:342-344 https://doi.org/10.1016/0300-5712(92)90021-4
  13. Scarano A, Piattelli M, Vrespa G, Petrone G, Iezzi G, Piattelli A. Bone healing around titanium and titanium nitridecoated dental implants with three surfaces: an experimental study in rats. Clin Implant Dent Relat Res 2003;5:103-111 https://doi.org/10.1111/j.1708-8208.2003.tb00191.x
  14. Scarano A, Piattelli M, Vrespa G, Caputi S, Piattelli A. Bacterial adhesion on titanium nitride-coated and uncoated implants: an in vivo human study. J Oral Implantol 2003;29:80-85 https://doi.org/10.1563/1548-1336(2003)029<0080:BAOTNA>2.3.CO;2
  15. Weiss EI, Kozak D, Gross MD Effect of repeated closures on opening torque values in seven abutment-implant systems. J Prosthet Dent 2000;84:194-9 https://doi.org/10.1067/mpr.2000.108069
  16. Bickford JH. An introduction to the design and behavior of bolted joints. 3rd ed. New York: Marcel Dekker 528-42, 1995
  17. Shigley JE. Mechanical Engineering Design. 3rd Edition. New York: McGraw Hill, 1987:227-52
  18. Byrne D, Jacobs S, O'connell B, Houston F, Claffey N. Preloads generated with repeated tightening in three types of screws used indental implant assemblies. J Prosthodont 2006;15:164-71 https://doi.org/10.1111/j.1532-849X.2006.00096.x
  19. Kim HJ, Choe HC, Kim SG, Chung CH. Effect of TiN coating of abutment screw on detorque force. Int J Oral Maxillofac Implants 2006 under review
  20. Abokowitz S, Burke JJ, Hiltz RH. Titanium in industry. New York(NY): Van Nostrand Co Inc; 1955
  21. Koo CI, Chung CH, Choe HC. Effect of surface coating on the screw release of dental implant screw. J Korean Acad Prosthodont 2004;42:210-226
  22. Milic M, Milosavljevic M, Bibic N, Nenadovic T. Mechanical Properties of TiN coatings. Thin Solid Films 1985;126:319- 323 https://doi.org/10.1016/0040-6090(85)90327-X
  23. Binon PP. Evaluation of matching accuracy and consistency of selected implant, standard abutments and laboratory analog. Int J Prosthodont 1995;8:162-178
  24. Binon PP. Evaluation of three slip fit hexagonal implants. Implant Dent 1996; 5:253-248
  25. Binon PP. Implants and components: Entering the new millennium. Int J Oral Maxillofac Implants 2000;5:76-94
  26. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont 1996;9:149-152
  27. Binon PP, McHugh MJ. The effect of eliminating implant/ abutment rotational misfit on screw joint stability. Int J Prosthodont 1996;9:511-519
  28. Binon PP, Sutter F, Brunski J, Gulbransen H, Weiner R. The role of screws in implant systems. Int J Oral Maxillofac Implants 1994;9:48-62
  29. Byren D, Houston F, Cleary R. Claffey N. The fit of cast and premachined implant abutments. J Prosthet Dent 1998;80:184- 192 https://doi.org/10.1016/S0022-3913(98)70108-8
  30. Carr AB, Brunski JB, Hurley E. Effect of fabrication, finishing, and polishing procedures on preload in prostheses using conventional gold and plastic cylinders. Int J Oral Maxillofac Implants 1996;11:589-598
  31. Carr AB, Brunski JB. Hurley E. The effect of finishing and polishing techniques in implant cylinder preload;9th Annual Meeting of the Academy of Osseointegration (Abstract). Int J Oral Maxilofac Implants 1995;10:123-124