Abstract
Measurement uncertainty assessment is very important in measurement and calibration. RRL provides antenna calibration services for EMI test. Reliability of EMI test depends on accurate antenna calibration. Antenna calibration results have to be accompanied with measurement uncertainty for its better reliability. In the late of 2005, CISPR issued the CISPR/A/644/C which describes the antenna calibration and measurement uncertainty. In this paper, on the basis of CISPR/A/644/C, we provide the measurement uncertainty values for dipole antenna calibration at the Calibration Test Site(CALTS) of Icheon. The antenna calibration method is 3-antenna height-scanning-averaging method, which measures the free-space antenna factor. We also considered all uncertainty sources that can affect measurement results during calibration.
측정 분야나 교정 분야에서 측정 불확도를 보고하는 것은 매우 중요하다. 정보통신부 전파연구소에서는 EMI 측정 시험소를 위하여 안테나 교정 서비스를 제공하고 있는데, 안테나 교정의 정확성은 측정 불확도 평가로 집약되어, EMI 시험의 신뢰성을 판가름하는 시금석이 된다. EMI에 관한 국제기구인 CISPR에서는 2005년 말에 안테나 교정에 대한 최초의 표준 초안인 CISPR/A/644/C를 내놓았다. 이 초안은 안테나교정 및 관련된 측정 불확도를 포함하고 있다. 본 논문에서는 이 문서를 근간으로 전파연구소 이천의 안테나 교정시험장에서 EMI용 다이폴안테나를 교정할 때의 측정 불확도를 산출하였다. 전파연구소의 안테나 교정 방법은 자유 공간 안테나 인자를 측정하는 3-안테나의 height scanning averaging 법으로서, 본 방법에 의하여 안테나가 교정될 때 측정 결과에 영향을 미칠 수 있는 모든 불확도 요인을 고려하였다.