DOI QR코드

DOI QR Code

Removal of Cerenkov Light in Fiber-optic Radiation Sensor Using Optical Filters

광학 필터를 이용한 광섬유 방사선 센서의 체렌코프 빛 제거

  • Jang, Kyoung-Won (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Lee, Bong-Soo (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Cho, Dong-Hyun (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Kim, Hyung-Shik (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Yi, Jeong-Han (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Lee, Jeong-Whan (School of Biomedical Engineering, College of Biomedical & Health, Konkuk University) ;
  • Kim, Sin (Department of Nuclear and Energy Engineering, College of Engineering, Cheju National University) ;
  • Cho, Hyo-Sung (Department of Radiation Science, Institute of Health Science, Yonsei University)
  • 장경원 (건국대학교 의료생명대학 의학공학부) ;
  • 이봉수 (건국대학교 의료생명대학 의학공학부) ;
  • 조동현 (건국대학교 의료생명대학 의학공학부) ;
  • 김형식 (건국대학교 의료생명대학 의학공학부) ;
  • 이정한 (건국대학교 의료생명대학 의학공학부) ;
  • 이정환 (건국대학교 의료생명대학 의학공학부) ;
  • 김신 (제주대학교 공과대학 에너지공학과) ;
  • 조효성 (연세대학교 보건대학 방사선학과)
  • Published : 2006.04.01

Abstract

In this study, a miniature fiber-optic radiation sensor has been developed using a water-equivalent organic scintillator for electron beam therapy dosimetry. The intensity of Cerenkov light is measured and characterized as a function of the incident angle of the electron beam from a LINAC. Also, a subtraction method using a background optical fiber without a scintillator and an optical discrimination method using optical filters are investigated to remove Cerenkov light, which could cause problems or limit the accuracy for detecting a fluorescent light signal in a fiber-optic radiation sensor.

본 연구에서는 유기 섬광체와 플라스틱 광섬유를 이용하여 치료용 전자선의 계측을 위한 광섬유 방사선 센서를 제작하였다. 또한, 선형가속기에서 발생되는 고 에너지 전자선에 의해 광섬유 방사선 센서를 이용한 전자선 계측에 있어 방해요소로 존재하는 체렌코프 빛을 감법 및 광학 필터링을 이용하여 제거하였고 두 가지방법들을 비교, 분석하였다.

Keywords

References

  1. B. Lee, W. Y. Choi, and J. K. Walker, 'Polymerpolymer miscibility study for plastic gradient index optical fiber,' Polymer Eng. and Sci, vol. 40, no. 9, pp. 1996-1999, 2000 https://doi.org/10.1002/pen.11331
  2. E. Takada, A. Kimura, Y. Hosono, H. Takahashi, and M. Nakazawa, 'Radiation distribution sensor with optical fibers for high radiation fields,' J. of Nucl. Sci. and Tech., vol. 36, no. 8, pp. 641-645, 1999 https://doi.org/10.3327/jnst.36.641
  3. B. M. Rogina, and B. Vojnovic, 'Application of optical fiber sensors for radiation dosimetry,' Radiation Measurements, vol. 26, no. 4, pp. 599-602, 1996 https://doi.org/10.1016/1350-4487(96)00032-7
  4. A. S Beddar, T. J. Kinsella, A. Ikhlef, and C. H. Sibata, 'A miniature Scintillator-Fiberoptic-PMT detector system for the dosimetry of small fields in stereotactic radiosurgery' IEEE Transactions on Nuclear Science, vol. 48, no. 3, pp. 924-928, 2001 https://doi.org/10.1109/23.940133
  5. A. S. Bedder, T. R. Mackie, and F. H. Attix, 'Cerenkov light generated in optical fibers and other light pipes irradiated by electron beams,' Phys. Med. Biol, vol. 37, pp. 925-935, 1992 https://doi.org/10.1088/0031-9155/37/4/007
  6. S. F. de Boer, A. S. Beddar and J. A. Rawlison. 'Optical filtering and spectral measurements of radiationinduced light in plastic scintillation dosimetry' Phys. Med. Biol, vol. 38 pp. 945-958, 1993 https://doi.org/10.1088/0031-9155/38/7/005
  7. Knoll, Glenn F. Knoll, Radiation detection and measurement - 3rd ed.(John Wiley & Sons, New York, USA, 1999), pp. 711-713